The biomechanics of wire fixation in the Ilizarov system.

M M Mullins1, A W Davidson, David Goodier, M Barry.   

Abstract

The purpose of this study was to establish the optimal fixation of tensioned wires to the frame construct in the Ilizarov system. The usual torque to which the fixation bolts were tightened in clinical practice was established by serial testing of orthopaedic surgeons' work in our unit. The force required to produce wire slippage from the different types of wire fixation bolts, tightened to a predetermined torque, was measured using a testing rig. Analysis of the usual torque to which bolts were tightened in clinical practice, revealed values in the range of 5-10Nm. The load required to cause failure of the tensioned wire varied considerably depending on the bolt configuration used and the applied torque. In clinical practice, wires are tensioned using a dynamometer to a variety of loads, depending on the clinical situation, up to 1275N. In applying multiple wires across a single ring, as is normal practice, these loads may be increased still further. The total load transmitted by single wires in weightbearing may be up to 2000N. Utilising the results of our work has enabled us to choose, where practically possible, the optimal bolt configuration for wire fixation. In addition, this work has revealed that in order to withstand the loads seen in clinical practice, wire fixation bolts should be tightened to at least 10Nm. Post-operatively, we now tighten all our wire fixation bolts to 10Nm, using a calibrated torque wrench.

Entities:  

Mesh:

Year:  2003        PMID: 12565025     DOI: 10.1016/s0020-1383(02)00262-0

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  14 in total

1.  Analytical modelling of Kirschner wires in Ilizarov circular external fixator as pretensioned slender beams.

Authors:  A R Zamani; S O Oyadiji
Journal:  J R Soc Interface       Date:  2009-03-06       Impact factor: 4.118

2.  Towards a solution of the wires' slippage problem of the Ilizarov external fixator.

Authors:  C Bairaktari; G Athanassiou; E Panagiotopoulos; D Deligianni
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-08-20

3.  Mechanobiology in the management of mobile atrophic and oligotrophic tibial nonunions.

Authors:  Nando Ferreira; Leonard Charles Marais; Colleen Aldous
Journal:  J Orthop       Date:  2015-11-01

Review 4.  [Complex reconstruction in Charcot arthropathy using the Ilizarov ring fixator].

Authors:  S H Wirth; N Espinosa; M Berli; L Jankauskas
Journal:  Orthopade       Date:  2015-01       Impact factor: 1.087

5.  A treatment for large defects of the tibia caused by infected nonunion: Ilizarov method with bone segment extension.

Authors:  K Xu; X Fu; Y-M Li; C-G Wang; Z-J Li
Journal:  Ir J Med Sci       Date:  2013-10-29       Impact factor: 1.568

6.  Direct and indirect loading of the Ilizarov external fixator: the effect on the interfragmentary movements and compressive loads.

Authors:  Jan Gessmann; Hinnerk Baecker; Birger Jettkant; Gert Muhr; Dominik Seybold
Journal:  Strategies Trauma Limb Reconstr       Date:  2011-02-11

7.  The use of twin-ring Ilizarov external fixator constructs: application and biomechanical proof-of principle with possible clinical indications.

Authors:  Theodoros B Grivas; Evangelos A Magnissalis
Journal:  J Orthop Surg Res       Date:  2011-08-11       Impact factor: 2.359

8.  The influence of a weight-bearing platform on the mechanical behavior of two Ilizarov ring fixators: tensioned wires vs. half-pins.

Authors:  Jan Gessmann; Mustafa Citak; Birger Jettkant; Thomas A Schildhauer; Dominik Seybold
Journal:  J Orthop Surg Res       Date:  2011-12-12       Impact factor: 2.359

9.  Heterologous mesenchymal stem cells successfully treat femoral pseudarthrosis in rats.

Authors:  Manoel Luiz Ferreira; Paulo Cesar Silva; Lucas Henrique Alvarez Silva; Danielle Cabral Bonfim; Lucas Cristo Conilho Macedo Müller; Christiano Costa Espósito; Alberto Schanaider
Journal:  J Transl Med       Date:  2012-03-20       Impact factor: 5.531

10.  The effect of HIV infection on the incidence and severity of circular external fixator pin track sepsis: a retrospective comparative study of 229 patients.

Authors:  Nando Ferreira; Leonard Charles Marais
Journal:  Strategies Trauma Limb Reconstr       Date:  2014-07-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.