Literature DB >> 12507102

Impact of the model C and Automatic Positioning System on gamma knife radiosurgery: an evaluation in vestibular schwannomas.

Jean Régis1, Motohiro Hayashi, Denis Porcheron, Christine Delsanti, Xavier Muracciole, Jean Claude Peragut.   

Abstract

OBJECT: The technical advances associated with the model C gamma knife include a robotized system enabling automatic positioning of the stereotactic coordinates. The purpose of this study was to analyze the clinical impact of this technical modification.
METHODS: The authors studied a sample of patients with vestibular schwannoma (VS). This sample included three groups treated using gamma knife radiosurgery. Group I comprised 21 patients with VS treated just before the installation of the Automatic Positioning System (APS). Group II included patients in Group I with new dose plans created using the APS (in other words, simulated dose plans). Group III consisted of a control group of 20 patients matched for tumor grade with the previous group and treated recently with the APS. Treatment times were calculated after correcting the time for each shot according to the age of the sources after reloading. The treatment times, including total time, irradiation time, and duration of the neurosurgical procedure, were analyzed. In addition, dose planning including number of isocenters, number of different collimators, malfunctions, and the conformity and selectivity indices were recorded. The trend was to reduce the mean number of collimator runs from 7.9 to 1.2 and to increase the mean number of shots from 7.9 to 15.6, mostly by using the 4-mm collimator exclusively. The APS-related conformity and selectivity were improved from 95 to 97% and from 78 to 84%, respectively. The total treatment time was reduced by 53%, and time required to interact with the patient in the room was considerably reduced (75%), giving the neurosurgeon greater freedom to perform other tasks during the treatment period. The reduction of the time spent by the neurosurgeon at work in the room was 84%. The total radiation time was increased by 54%.
CONCLUSIONS: The preliminary results of this study indicate that the robotization of the gamma knife is likely a major advance in radiosurgery.

Entities:  

Mesh:

Year:  2002        PMID: 12507102     DOI: 10.3171/jns.2002.97.supplement

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  6 in total

1.  Diagrams for the Leksell Gamma Knife 4C automatic positioning system accessible shot coordinates.

Authors:  Yoichi Watanabe
Journal:  Radiol Phys Technol       Date:  2012-11-18

Review 2.  Stereotactic radiosurgery for pituitary adenomas: a review of the literature.

Authors:  Edward R Laws; Jason P Sheehan; Jonas M Sheehan; Jay Jagnathan; John A Jane; Rod Oskouian
Journal:  J Neurooncol       Date:  2004 Aug-Sep       Impact factor: 4.130

3.  Modern Gamma Knife radiosurgery of vestibular schwannomas: treatment concept, volumetric tumor response, and functional results.

Authors:  Samuel M Lipski; Motohiro Hayashi; Mikhail Chernov; Marc Levivier; Yoshikazu Okada
Journal:  Neurosurg Rev       Date:  2014-12-19       Impact factor: 3.042

4.  Impact of decaying dose rate in gamma knife radiosurgery: in vitro study on 9L rat gliosarcoma cells.

Authors:  Ajay Niranjan; Glenn Gobbel; Josef Novotny; Jagdish Bhatnagar; Wendy Fellows; L Dade Lunsford
Journal:  J Radiosurg SBRT       Date:  2012

5.  First year experience with newly developed Leksell Gamma Knife Perfexion.

Authors:  Jagdish P Bhatnagar; Josef Novotny; Ajay Niranjan; Douglas Kondziolka; John Flickinger; Dade Lunsford; M Saiful Huq
Journal:  J Med Phys       Date:  2009-07

Review 6.  Neurosurgical robotics: a review of brain and spine applications.

Authors:  Chris S Karas; E Antonio Chiocca
Journal:  J Robot Surg       Date:  2007-02-27
  6 in total