Evaluation of a reagent-strip method for glucose in whole blood, as compared with a hexokinase method.

T C Stewart.   

Abstract

The hexokinase method (I) for glucose in biological fluids, proposed by the FDA as a potential national standard method, is compared with the Eyetone/Dextrostix/Control method (II) for measurement of glucose in whole blood. I used a new, two-step calibration procedure based on a 1.30 g/liter standard for the II system. The correlation coefficient for results by the two methods was 0.99 for 142 assays in duplicate. Recovery values were near 100% (I, range 100.0 to 106.2%; II, range 100.8 to 105.1%), and the coefficients of variation were respectively +/- 4.1% and +/- 4.5%. Considering the close correlation between the two methods, together with the simplicity, speed, and convenience of II, I conclude that it has great clinical utility, particularly in emergency situations. The precision and accuracy of II depends greatly on proper calibration and control procedures and on adequate training of the operator.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 1245062

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  5 in total

1.  Dextrostix-Eyetone in the insulin hypoglycaemia test.

Authors:  M A Preece; R G Newall
Journal:  Br Med J       Date:  1977-07-16

2.  A proposal for detecting and managing gestational diabetes by coordinating existing services.

Authors:  L Macupa
Journal:  Public Health Rep       Date:  1986 Jan-Feb       Impact factor: 2.792

3.  Performance evaluation of reflectance meter for glucose determination by two different reagent strips.

Authors:  D Spotti; C Rocco; O Carandente
Journal:  Acta Diabetol Lat       Date:  1985 Apr-Jun

4.  Prevalence of diabetes mellitus in elderly men aged 65 to 84 years in eastern and western Finland.

Authors:  J Tuomilehto; A Nissinen; S L Kivelä; J Pekkanen; E Kaarsalo; E Wolf; A Aro; S Punsar; M J Karvonen
Journal:  Diabetologia       Date:  1986-09       Impact factor: 10.122

5.  Assessment of a reflectance photometer in a veterinary laboratory.

Authors:  C J Belford; J H Lumsden
Journal:  Can Vet J       Date:  1984-06       Impact factor: 1.008

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.