Literature DB >> 12413442

Myocardial contrast echocardiography yields best accuracy using quantitative analysis of digital data from pulse inversion technique: comparison with second harmonic imaging and harmonic power Doppler during simultaneous dipyridamole stress SPECT studies.

H von Bibra1, D Bone, U Niklasson, L Eurenius, A Hansen.   

Abstract

AIMS: This prospective study assesses the (1) feasibility of quantifying ultrasound myocardial perfusion studies based on the densitometric analysis of digital data and the (2) comparison of pulse inversion, second harmonic and harmonic power Doppler modalities with SPECT. METHODS AND
RESULTS: Twenty-three patients with suspected ischaemic heart disease had i.v. injections of Tc-Sestamibi and Optison during a dipyridamole stress test for echocardiography in pulse inversion, second harmonic and harmonic power Doppler mode. Analysis was (a) visual by scoring and (b) quantitative by densitometry of digital data for background subtracted myocardial opacification (a.u.) and normalized contrast effect (%). In the nine control patients, myocardial opacification at stress was greater (P< or =0.002) than in the pathologic group (5. +/- 3.3 vs 2.6 +/- 2.5 a.u. in pulse inversion, 5.4 +/- 2.1 vs 2.4 +/- 1.8 in second harmonic and 7.1 +/- 3.7 vs 4.9 +/- 3.7 a.u. in harmonic power Doppler). In the pathologic group, normalized contrast effect decreased significantly during stress (23.7 +/- 18.8 to 11.3 +/- 10.8%, P<0.003) only in pulse inversion. Kappa values for patient based diagnostic agreement with SPECT were 0.75 by pulse inversion, 0.62 by second harmonic and 0.52 by harmonic power Doppler for quantitative analysis, and 0.51, 0.37 and 0.35 respectively, for visual assessment.
CONCLUSION: Myocardial contrast echocardiography should be analysed using densitometry of digital data. The new technique pulse inversion demonstrates best agreement with SPECT data. Copyright 2002 The European Society of Cardiology

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12413442

Source DB:  PubMed          Journal:  Eur J Echocardiogr        ISSN: 1532-2114


  5 in total

1.  Augmented metabolic control improves myocardial diastolic function and perfusion in patients with non-insulin dependent diabetes.

Authors:  H von Bibra; A Hansen; V Dounis; T Bystedt; K Malmberg; L Rydén
Journal:  Heart       Date:  2004-12       Impact factor: 5.994

Review 2.  Characterization of the rdar morphotype, a multicellular behaviour in Enterobacteriaceae.

Authors:  U Römling
Journal:  Cell Mol Life Sci       Date:  2005-06       Impact factor: 9.261

3.  [Echocardiography : Important diagnostic pillar in cardiology].

Authors:  R Erbel
Journal:  Herz       Date:  2017-05       Impact factor: 1.443

4.  Global myocardial perfusion and diastolic function are impaired to a similar extent in patients with type 2 diabetes mellitus and in patients with coronary artery disease--evaluation by contrast echocardiography and pulsed tissue Doppler.

Authors:  V Dounis; T Siegmund; A Hansen; J Jensen; P-M Schumm-Draeger; H von Bibra
Journal:  Diabetologia       Date:  2006-09-21       Impact factor: 10.122

5.  Quantification of resting myocardial blood flow velocity in normal humans using real-time contrast echocardiography. A feasibility study.

Authors:  Siri Malm; Sigmund Frigstad; Frode Helland; Kjetil Oye; Stig Slordahl; Terje Skjarpe
Journal:  Cardiovasc Ultrasound       Date:  2005-06-16       Impact factor: 2.062

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.