Literature DB >> 12408301

Clinical implementation of AAPM TG61 protocol for kilovoltage x-ray beam dosimetry.

Sua Yoo1, Daniel Grimm, Ron Zhu, Paul Jursinic, Francisco Lopez, Jason Rownd, Michael Gillin.   

Abstract

Historically, there have been a variety of dosimetry protocols used for kilovoltage x-ray therapy beams with a set of conversion factors and correction factors taken from different references. Corresponding to the continued installation and use of kilovoltage machines, the American Association of Physicists in Medicine (AAPM) presented a unified protocol developed by Task Group 61 (TG61). TG61 determines the absorbed dose to water with an ionization chamber calibrated in air in terms of air kerma (Nk). TG61 presents both an in-air method and an in-phantom method. In this work we only examine the TG61 in-air method. Our traditional dosimetry procedure, which is based upon NCRP Report 69 and on material found in standard medical physics texts, has been compared to the TG61. A variety of kilovoltage beam energies were examined with a set of various field sizes and source to surface distances. TG61 published updated data for the mass absorption coefficient ratios, backscatter factors, and the average energy per ion pair factor. The following conclusions have been reached: (1) Our traditional procedures and the TG61 protocol for in-air measurements are equivalent. (2) The conversion and correction factors used in TG61 are different by up to 4.5% compared to the old factors that we have used. (3) The application of the TG61 factors can result in up to 5% differences in the determination of the absorbed dose.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12408301     DOI: 10.1118/1.1508376

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Application of MOSFET detectors for dosimetry in small animal radiography using short exposure times.

Authors:  Ming De Lin; Greta Toncheva; Giao Nguyen; Sangroh Kim; Colin Anderson-Evans; G Allan Johnson; Terry T Yoshizumi
Journal:  Radiat Res       Date:  2008-08       Impact factor: 2.841

2.  A novel technique for image-guided local heart irradiation in the rat.

Authors:  Sunil Sharma; Eduardo G Moros; Marjan Boerma; Vijayalakshmi Sridharan; Eun Young Han; Richard Clarkson; Martin Hauer-Jensen; Peter M Corry
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

3.  How low can you go? A CBCT dose reduction study.

Authors:  Arthur J Olch; Parham Alaei
Journal:  J Appl Clin Med Phys       Date:  2021-01-15       Impact factor: 2.102

4.  Longitudinal in-vivo quantification of tumour microvascular heterogeneity by optical coherence angiography in pre-clinical radiation therapy.

Authors:  Nader Allam; W Jeffrey Zabel; Valentin Demidov; Blake Jones; Costel Flueraru; Edward Taylor; I Alex Vitkin
Journal:  Sci Rep       Date:  2022-04-12       Impact factor: 4.379

5.  Performance characteristics of mobile MOSFET dosimeter for kilovoltage X-rays used in image guided radiotherapy.

Authors:  A Sathish Kumar; I Rabi Raja Singh; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2015 Jul-Sep
  5 in total