PURPOSE: The aim of this study was to evaluate the thrombogenicity of different peripheral stent types in a standardized in vitro model with fresh human whole blood.
MATERIALS AND METHODS: Different stents (N = 77; n = 7 of each of 11 types) were implanted in polyvinyl chloride tubing loops and filled with donor blood samples. After 120 minutes of blood circulation, the thrombin-antithrombin III complex (TAT) levels, beta-thromboglobulin (beta-TG) levels, and platelet counts were assessed.
RESULTS: After 2 hours, significant differences were seen. TAT values (+/- SD) with the investigated stents were 31 micro g/mL +/- 20 (control, no stent), 328 micro g/mL +/- 206 (Saxx stent, peripheral medium CrNi31 L), 651 micro g/mL +/- 760 (Palmaz Corinthian Stent, 316 L stainless steel, electropolished), 1,609 micro g/mL +/- 1,264 (Palmaz Corinthian Stent, 316 L stainless steel, not electropolished), 810 micro g/mL +/- 578 (Palmaz Schatz long medium stent), 569 micro g/mL +/- 347 (Smart Nitinol stent), 1,037 micro g/mL +/- 577 (Megalink peripheral stent), 543 micro g/mL +/- 487 (peripheral stent, electropolished), 1,674 micro g/mL +/- 2,057 (peripheral stent, not electropolished), 3,128 micro g/mL +/- 1,812 (SelfX Nitinol stent, polished), 5,897 micro g/mL +/- 2,380 (SelfX Nitinol stent, unpolished), and 1,458 micro g/mL +/- 887 (bridge stent). The platelet count (x1,000/ micro L +/- SD) was 218 +/- 35 (control, no stent), 188 +/- 22 (Saxx stent), 187 +/- 20 (Palmaz Corinthian stent, electropolished), 135 +/- 37 (Palmaz Corinthian stent, not electropolished), 170 +/- 24 (Palmaz Schatz stent), 180 +/- 36 (Smart Nitinol stent), 159 +/- 26 (Megalink peripheral stent), 173 +/- 17 (peripheral stent, electropolished), 133 +/- 51 (peripheral stent, not electropolished), 123 +/- 37 (SelfX Nitinol stent, polished), 52 +/- 27 (SelfX Nitinol stent, unpolished), and 130 +/- 31 (bridge stent).
CONCLUSION: This standardized study showed a wide range of platelet activation after stent implantation. Electropolishing clearly reduced the thrombogenicity of the stents.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12397125     DOI: 10.1016/s1051-0443(07)61868-5

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  15 in total

1.  Effects of micro-magnetic field at the surface of 316L and NiTi alloy on blood compatibility.

Authors:  Qiang Liu; Xiao Nong Cheng; Huang Xia Fei
Journal:  Med Biol Eng Comput       Date:  2010-10-09       Impact factor: 2.602

2.  Preparation and characterization of rapamycin-loaded PLGA coating stent.

Authors:  C J Pan; J J Tang; Y J Weng; J Wang; N Huang
Journal:  J Mater Sci Mater Med       Date:  2007-07-03       Impact factor: 3.896

3.  Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings.

Authors:  Martin Heuschmid; Benjamin Wiesinger; Gunnar Tepe; Oliver Luz; Andreas F Kopp; Claus D Claussen; Stephan H Duda
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

Review 4.  In Vitro models for thrombogenicity testing of blood-recirculating medical devices.

Authors:  Deepika N Sarode; Shuvo Roy
Journal:  Expert Rev Med Devices       Date:  2019-06-10       Impact factor: 3.166

5.  In vitro hemocompatibility of thin film nitinol in stenotic flow conditions.

Authors:  C P Kealey; S A Whelan; Y J Chun; C H Soojung; A W Tulloch; K P Mohanchandra; D Di Carlo; D S Levi; G P Carman; D A Rigberg
Journal:  Biomaterials       Date:  2010-12       Impact factor: 12.479

6.  Assessment of Epicardial Fat Volume With Threshold-Based 3-Dimensional Segmentation in CT: Comparison With the 2-Dimensional Short Axis-Based Method.

Authors:  Mi Jung Park; Jung Im Jung; Yong Seog Oh; Ho-Joong Youn
Journal:  Korean Circ J       Date:  2010-07-26       Impact factor: 3.243

7.  The role of stents in the treatment of congenital heart disease: Current status and future perspectives.

Authors:  Bjoern Peters; Peter Ewert; Felix Berger
Journal:  Ann Pediatr Cardiol       Date:  2009-01

8.  Effects of antiplatelet agents and other factors on neointimal proliferation in iliac artery stenting: intravascular ultrasound analysis.

Authors:  Hisao Kumakura; Hiroyoshi Kanai; Yoshihiro Araki; Akira Koizumi; Shu Kasama; Hiroyuki Sumino; Shuichi Ichikawa
Journal:  Ann Vasc Dis       Date:  2009-12-14

9.  PDMS content affects in vitro hemocompatibility of synthetic vascular grafts.

Authors:  Dario Spiller; Paola Losi; Enrica Briganti; Silverio Sbrana; Silvia Kull; Ilaria Martinelli; Giorgio Soldani
Journal:  J Mater Sci Mater Med       Date:  2007-02-01       Impact factor: 4.727

10.  Oligonucleotide and Parylene Surface Coating of Polystyrene and ePTFE for Improved Endothelial Cell Attachment and Hemocompatibility.

Authors:  Martina Schleicher; Jan Hansmann; Bentsian Elkin; Petra J Kluger; Simone Liebscher; Agnes J T Huber; Olaf Fritze; Christine Schille; Michaela Müller; Katja Schenke-Layland; Martina Seifert; Heike Walles; Hans-Peter Wendel; Ulrich A Stock
Journal:  Int J Biomater       Date:  2012-03-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.