Literature DB >> 12211674

The performance of two hydroxyapatite-coated acetabular cups compared with Charnley cups. From the Norwegian Arthroplasty Register.

L I Havelin1, B Espehaug, L B Engesaeter.   

Abstract

We have compared the survival of two hydroxyapatite (HA)-coated cups, 1208 Atoll hemispheric and 2641 Tropic threaded, with cemented Charnley all-polyethylene cups (16 021) using the Cox regression model. The Tropic cup used in combination with an alumina ceramic femoral head, had good results, similar to those of the Charnley cup. When used in combination with a stainless-steel head, however, the risk of revision beyond four years was increased 3.4 times for the Tropic cup compared with the Charnley cup (p < 0.001). Over the same period, the Atoll cup had an increased risk of revision of 3.8 times when used with the alumina heads (p < 0.001) and an increased risk of 6.1 times when used with stainless-steel heads (p < 0.001). Revision because of wear and osteolysis was more common with both types of HA-coated cup than with the Charnley cup. The rate of revision of the Atoll cup because of aseptic loosening was also increased. We found that HA-coated cups did not perform better than the Charnley cup.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12211674     DOI: 10.1302/0301-620x.84b6.12492

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  22 in total

1.  Excellent results with cementless total hip arthroplasty and alumina-on-alumina pairing: minimum ten-year follow-up.

Authors:  Alexander M Kress; Rainer Schmidt; Ulrich Holzwarth; Raimund Forst; Lutz A Mueller
Journal:  Int Orthop       Date:  2010-11-16       Impact factor: 3.075

2.  Head material influences survival of a cemented total hip prosthesis in the Norwegian Arthroplasty Register.

Authors:  Thomas Kadar; Eva Dybvik; Geir Hallan; Ove Furnes; Leif Ivar Havelin
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

3.  Long-term follow-up (11 years plus) results of JRI (Furlong) total hip arthroplasty in young patients: cause for concern regarding acetabular cup?

Authors:  D K Sharma; Stuart Brooks
Journal:  Int Orthop       Date:  2006-06-08       Impact factor: 3.075

4.  Long-term results of the Wagner conical screw cup.

Authors:  Alexander Schuh; Annemarie Schraml; Gerd Hohenberger
Journal:  Int Orthop       Date:  2008-06-26       Impact factor: 3.075

Review 5.  Total hip arthroplasty: areview of advances, advantages and limitations.

Authors:  Xin-Wei Liu; Ying Zi; Liang-Bi Xiang; Yu Wang
Journal:  Int J Clin Exp Med       Date:  2015-01-15

6.  Does a titanium-coated polyethylene press-fit cup give reliable midterm results?

Authors:  Nikki Hooper; Harry Sargeant; Chris Frampton; Gary Hooper
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

7.  Cementless total hip arthroplasty for primary osteoarthritis in patients aged 55 years and older.

Authors:  Keijo T Mäkelä; Antti Eskelinen; Pekka Paavolainen; Pekka Pulkkinen; Ville Remes
Journal:  Acta Orthop       Date:  2010-02       Impact factor: 3.717

8.  No differences in outcomes between cemented and uncemented acetabular components after 12-14 years: results from a randomized controlled trial comparing Duraloc with Charnley cups.

Authors:  Kristian Bjørgul; Wendy M Novicoff; S T Andersen; K Brevig; F Thu; M Wiig; O Ahlund
Journal:  J Orthop Traumatol       Date:  2010-03-03

9.  Long-term results of the threaded Weill cup in primary total hip arthroplasty: a 15-20-year follow-up study.

Authors:  Michael Clarius; Alexander W Jung; Patric Raiss; Marcus R Streit; Christian Merle; Peter R Aldinger
Journal:  Int Orthop       Date:  2009-07-24       Impact factor: 3.075

10.  The Nordic Arthroplasty Register Association: a unique collaboration between 3 national hip arthroplasty registries with 280,201 THRs.

Authors:  Leif I Havelin; Anne M Fenstad; Roger Salomonsson; Frank Mehnert; Ove Furnes; Søren Overgaard; Alma B Pedersen; Peter Herberts; Johan Kärrholm; Göran Garellick
Journal:  Acta Orthop       Date:  2009-08       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.