A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use.

Loredana Baboi1, Fabien Subtil2, Claude Guérin3.   

Abstract

BACKGROUND: Turbine-powered ventilators are not only designed for long-term ventilation at home but also for hospital use. It is important to verify their capabilities in delivering fraction of oxygen in air (FIO2) and tidal volume (VT).
METHODS: We assessed the FIO2 accuracy and the VT delivery in four home care ventilators (HCV) on the bench. The four HCV were Astral 150, Elisée 150, Monnal T50 and Trilogy 200 HCV, which were connected to a lung model (ASL 5000). For assessing FIO2 accuracy, lung model was set to mimic an obstructive lung and HCV were set in volume controlled mode (VC). They supplied with air, 3 or 15 L/min oxygen and FIO2 was measured by using a ventilator tester (Citrex H4TM). For the VT accuracy, the lung model was set in a way to mimic three adult configurations (normal, obstructive, or restrictive respiratory disorder) and one pediatric configuration. Each HCV was set in VC. Two VT (300 and 500 mL) in adult lung configuration and one 50 mL VT in pediatric lung configuration, at two positive end expiratory pressures 5 and 10 cmH2O, were tested. VT accuracy was measured as volume error (the relative difference between set and measured VT). Statistical analysis was performed by suing one-factor ANOVA with a Bonferroni correction for multiple tests.
RESULTS: For Astral 150, Elisée 150, Monnal T50 and Trilogy 200, FIO2 averaged 99.2%, 93.7%, 86.3%, and 62.1%, respectively, at 15 L/min oxygen supplementation rate (P<0.001). Volume error was 0.5%±0%, -38%±0%, -9%±0%, -29%±0% and -36%±0% for pediatric lung condition (P<0.001). In adult lung configurations, Monnal T50 systematically over delivered VT and Trilogy 150 was sensitive to lung configuration when VT was set to 300 mL at either positive end-expiratory pressure (PEEP).
CONCLUSIONS: HCV are different in terms of FIO2 efficiency and VT delivery.

Entities:  

Keywords:  Home care ventilators (HCV); fraction of oxygen in air (FIO2); intermediate care; life support; tidal volume (VT)

Year:  2016        PMID: 28149559      PMCID: PMC5227209          DOI: 10.21037/jtd.2016.12.64

Source DB:  PubMed          Journal:  J Thorac Dis        ISSN: 2072-1439            Impact factor:   2.895


  26 in total

Review 1.  Which ventilators and modes can be used to deliver noninvasive ventilation?

Authors:  Robert L Chatburn
Journal:  Respir Care       Date:  2009-01       Impact factor: 2.258

2.  A trial of intraoperative low-tidal-volume ventilation in abdominal surgery.

Authors:  Emmanuel Futier; Jean-Michel Constantin; Catherine Paugam-Burtz; Julien Pascal; Mathilde Eurin; Arthur Neuschwander; Emmanuel Marret; Marc Beaussier; Christophe Gutton; Jean-Yves Lefrant; Bernard Allaouchiche; Daniel Verzilli; Marc Leone; Audrey De Jong; Jean-Etienne Bazin; Bruno Pereira; Samir Jaber
Journal:  N Engl J Med       Date:  2013-08-01       Impact factor: 91.245

3.  Patient-ventilator interaction and inspiratory effort during pressure support ventilation in patients with different pathologies.

Authors:  S Nava; C Bruschi; C Fracchia; A Braschi; F Rubini
Journal:  Eur Respir J       Date:  1997-01       Impact factor: 16.671

4.  Respiratory muscle activity related to flow and lung volume in preterm infants compared with term infants.

Authors:  Gerard J Hutten; Leo A van Eykern; Philipp Latzin; Cindy Thamrin; Wim M van Aalderen; Urs Frey
Journal:  Pediatr Res       Date:  2010-10       Impact factor: 3.756

5.  The appropriate setting of noninvasive pressure support ventilation in stable COPD patients.

Authors:  M Vitacca; S Nava; M Confalonieri; L Bianchi; R Porta; E Clini; N Ambrosino
Journal:  Chest       Date:  2000-11       Impact factor: 9.410

6.  Effect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator.

Authors:  Steven Zhou; Robert L Chatburn
Journal:  Respir Care       Date:  2014-08       Impact factor: 2.258

7.  Evaluation of a volume targeted NIV device: bench evaluation of the Breathe Technologies non-invasive open ventilation system (NIOV™).

Authors:  Thomas Blakeman; Richard Branson
Journal:  COPD       Date:  2014-09       Impact factor: 2.409

Review 8.  Ventilation with lower tidal volumes for critically ill patients without the acute respiratory distress syndrome: a systematic translational review and meta-analysis.

Authors:  Ary Serpa Neto; Liselotte Nagtzaam; Marcus J Schultz
Journal:  Curr Opin Crit Care       Date:  2014-02       Impact factor: 3.687

9.  Temperature of gas delivered from ventilators.

Authors:  Yusuke Chikata; Mutsuo Onodera; Hideaki Imanaka; Masaji Nishimura
Journal:  J Intensive Care       Date:  2013-11-08

10.  Leaks can dramatically decrease FiO2 on home ventilators: a bench study.

Authors:  Philippe Goutorbe; Erwan Daranda; Yves Asencio; Pierre Esnault; Bertrand Prunet; Julien Bordes; Bruno Palmier; Eric Meaudre
Journal:  BMC Res Notes       Date:  2013-07-21
View more
  1 in total

1.  Easier access to mechanical ventilation worldwide: an urgent need for low income countries, especially in face of the growing COVID-19 crisis.

Authors:  Claude Guérin; Patrick Lévy
Journal:  Eur Respir J       Date:  2020-06-04       Impact factor: 16.671

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.