Literature DB >> 12188475

Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head.

M R Norton1, R Yarlagadda, G H Anderson.   

Abstract

We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at five years. All hips which failed did so because of aseptic loosening with progressive osteolysis or radiolucencies. We therefore recommend early and regular review of all patients with this combination of implants and early revision surgery in order to avoid massive bone loss.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12188475     DOI: 10.1302/0301-620x.84b5.12679

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  8 in total

Review 1.  Manufacturers affect clinical results of THA with zirconia heads: a systematic review.

Authors:  Hiroyuki Yoshitomi; Satoru Shikata; Hiromu Ito; Takeo Nakayama; Takashi Nakamura
Journal:  Clin Orthop Relat Res       Date:  2009-01-31       Impact factor: 4.176

2.  The John Charnley Award: Highly crosslinked polyethylene in total hip arthroplasty decreases long-term wear: a double-blind randomized trial.

Authors:  Siôn Glyn-Jones; Geraint E R Thomas; Patrick Garfjeld-Roberts; Roger Gundle; Adrian Taylor; Peter McLardy-Smith; David W Murray
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

3.  Second-generation highly cross-linked X3™ polyethylene wear: a preliminary radiostereometric analysis study.

Authors:  David G Campbell; John R Field; Stuart A Callary
Journal:  Clin Orthop Relat Res       Date:  2010-02-12       Impact factor: 4.176

4.  The Elite-Plus stem migrates more than the flanged Charnley stem.

Authors:  Thord von Schewelov; Lennart Sanzén; Jack Besjakov; Ake Carlsson
Journal:  Acta Orthop       Date:  2010-06       Impact factor: 3.717

5.  Analysis of polyethylene wear in plain radiographs.

Authors:  Maiken Stilling; Kjeld Søballe; Niels Trolle Andersen; Kristian Larsen; Ole Rahbek
Journal:  Acta Orthop       Date:  2009-12       Impact factor: 3.717

Review 6.  Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty.

Authors:  Vasileios I Sakellariou; Peter Sculco; Lazaros Poultsides; Timothy Wright; Thomas P Sculco
Journal:  HSS J       Date:  2013-08-10

7.  The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.

Authors:  Maiken Stilling; Kristian Larsen; Niels T Andersen; Kjeld Søballe; Søren Kold; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

Review 8.  Adherence of hip and knee arthroplasty studies to RSA standardization guidelines. A systematic review.

Authors:  Rami Madanat; Tatu J Mäkinen; Hannu T Aro; Charles Bragdon; Henrik Malchau
Journal:  Acta Orthop       Date:  2014-06-23       Impact factor: 3.717

  8 in total