Reducing loss of image quality because of the attenuation artifact in uncorrected PET whole-body images.

Johan Nuyts1, Sigrid Stroobants, Patrick Dupont, Stefaan Vleugels, Patrick Flamen, Luc Mortelmans.   

Abstract

UNLABELLED: In whole-body PET, it is not unusual to shorten the study time by omitting the transmission scan and to ignore attenuation during reconstruction. If a transmission scan is available, many centers reconstruct the images with, but also without, attenuation correction. Although ignoring attenuation leads to an artifact in the reconstructed images, these images still provide valuable diagnostic information in oncologic applications. Several authors have reported that the attenuation artifact may actually increase the tumor-to-background ratio. In this study, we analyzed the causes of the artifact and proposed a new algorithm to reduce the adverse effects on visual image quality.
METHODS: We analyzed the causes of the attenuation artifact mathematically and numerically, and we examined its effect on tumor-to-background ratio and on signal-to-noise ratio. In addition, we showed that the attenuation artifact may lead to loss of image detail in conventional maximum-likelihood expectation maximization (MLEM) reconstruction. A new maximum-likelihood algorithm allowing negative reconstruction values (NEG-ML) was derived to reduce this loss.
RESULTS: The attenuation artifact consists of 2 components. The first component is the well-known scaling effect: The apparent activity is reduced because attenuation decreases the fraction of detected photons. The second component is a relatively smooth negative contribution that is added to attenuated regions surrounded by activity. The second component tends to increase the tumor-to-background ratio. However, a simulation experiment shows that this increase in signal may be entirely offset by an increase in noise. The negative contribution can interfere with the nonnegativity constraint of the MLEM algorithm, leading to loss of image detail in regions of high attenuation. The new NEG-ML algorithm avoids the problem by allowing negative pixel values. The algorithm is similar to MLEM in the suppression of the streak artifact but provides more anatomic information. In our department, it is in routine clinical use for reconstruction of PET whole-body images without attenuation correction.
CONCLUSION: Ignoring attenuation may increase the tumor-to-background ratio, but this increase does not imply improved tumor detection. The NEG-ML algorithm reduces the adverse effect of the attenuation artifact on visual image quality.

Entities:  

Mesh:

Year:  2002        PMID: 12163632

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  9 in total

1.  Location-known-exactly human-observer ROC studies of attenuation and other corrections for SPECT lung imaging.

Authors:  Andre Lehovich; Howard C Gifford; Michael A King
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2006

2.  Comparison of HRRT and HR+ scanners for quantitative (R)-[11C]verapamil, [11C]raclopride and [11C]flumazenil brain studies.

Authors:  Floris H P van Velden; Syahir M Mansor; Daniëlle M E van Assema; Bart N M van Berckel; Femke E Froklage; Shaonan Wang; Robert C Schuit; Marie-Claude Asselin; Adriaan A Lammertsma; Ronald Boellaard; Marc C Huisman
Journal:  Mol Imaging Biol       Date:  2015-02       Impact factor: 3.488

3.  Bias reduction for low-statistics PET: maximum likelihood reconstruction with a modified Poisson distribution.

Authors:  Katrien Van Slambrouck; Simon Stute; Claude Comtat; Merence Sibomana; Floris H P van Velden; Ronald Boellaard; Johan Nuyts
Journal:  IEEE Trans Med Imaging       Date:  2014-08-14       Impact factor: 10.048

4.  Evaluation of bias and variance in low-count OSEM list mode reconstruction.

Authors:  Y Jian; B Planeta; R E Carson
Journal:  Phys Med Biol       Date:  2014-12-05       Impact factor: 3.609

5.  Direct 4D parametric imaging for linearized models of reversibly binding PET tracers using generalized AB-EM reconstruction.

Authors:  Arman Rahmim; Yun Zhou; Jing Tang; Lijun Lu; Vesna Sossi; Dean F Wong
Journal:  Phys Med Biol       Date:  2012-01-18       Impact factor: 3.609

6.  An evaluation of iterative reconstruction strategies based on mediastinal lesion detection using hybrid Ga-67 SPECT images.

Authors:  Nicholas F Pereira; Howard C Gifford; P Hendrik Pretorius; Mark Smyczynski; Robert Licho; Peter Schneider; Troy Farncombe; Michael A King
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

7.  The Influence of Photon Attenuation on Tumor-to-Background and Signal-to-Noise Ratios for SPECT Imaging.

Authors:  Edward J Soares; Michael A King; Charles L Byrne; Howard C Gifford; Andre Lehovich
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007

8.  A study to improve the image quality in low-dose computed tomography (SPECT) using filtration.

Authors:  Sc Kheruka; Uc Naithani; Ak Maurya; Nk Painuly; Lm Aggarwal; S Gambhir
Journal:  Indian J Nucl Med       Date:  2011-01

Review 9.  Radioembolization and the Dynamic Role of (90)Y PET/CT.

Authors:  Alexander S Pasciak; Austin C Bourgeois; J Mark McKinney; Ted T Chang; Dustin R Osborne; Shelley N Acuff; Yong C Bradley
Journal:  Front Oncol       Date:  2014-02-27       Impact factor: 6.244

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.