Comparison of Goldmann applanation tonometry with the Tonopen for measuring intraocular pressure in a population-based glaucoma survey in rural West Bengal.

Manas Bandyopadhyay1, Amitabha Raychaudhuri, Saibendu K Lahiri, Ellen C Schwartz, Mark Myatt, Gordon J Johnson.   

Abstract

AIM: This study was undertaken to compare the measurement of intraocular pressure (IOP) in a population-based glaucoma survey by the Goldmann Applanation Tonometer (GAT) and the Tonopen with respect to IOP readings, the time required to measure IOP and subjects' preference for the two instruments. The purpose of the study was to see which of the two instruments could be more suitable in a rural survey, particularly house-to-house, in trying to find cases of advanced glaucoma.
METHOD: The study was conducted on a total of 203 subjects as a part of a population-based glaucoma survey. The IOP of one randomly selected eye of each subject was measured, first with the GAT and subsequently with the Tonopen. The time taken for each procedure was noted and subjects' preference was ascertained.
RESULTS: The mean of the paired differences between the readings of the two instruments (Tonopen minus GAT) was 1 mm Hg with a standard deviation of 2.28. The Tonopen readings were within +/-3 mm Hg of the GAT readings in 86.2% of eyes. In 7.4% of the eyes the paired differences were > or =+/-5 mm Hg. Measurement of IOP by the Tonopen was much quicker, the mean time difference being 38.7 sec. Subjects' preference for the Tonopen was higher.
CONCLUSION: The overall agreement between the measurements of the two instruments was good but a small percentage of large difference (> or = +/-5 mm Hg in 7.4%) may be of concern in a population-based survey. The time to measure IOP and the subjects' preference were strongly in favour of the Tonopen.

Entities:  

Mesh:

Year:  2002        PMID: 12045888     DOI: 10.1076/opep.9.3.215.1518

Source DB:  PubMed          Journal:  Ophthalmic Epidemiol        ISSN: 0928-6586            Impact factor:   1.648


  11 in total

1.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

2.  Evaluation of the effectiveness of ophthalmic assistants as screeners for glaucoma in North India.

Authors:  S K Sinha; N Astbury
Journal:  Eye (Lond)       Date:  2011-07-01       Impact factor: 3.775

3.  Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication.

Authors:  Shikha Gupta; Gautam Sinha; Reetika Sharma; Bhagabat Nayak; Bharat Patil; Bibhuti Kashyap; Abdul Shameer; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

4.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

5.  Determinants and heritability of intraocular pressure and cup-to-disc ratio in a defined older population.

Authors:  Ta C Chang; Nathan G Congdon; Robert Wojciechowski; Beatriz Muñoz; Donna Gilbert; Ping Chen; David S Friedman; Sheila K West
Journal:  Ophthalmology       Date:  2005-07       Impact factor: 12.079

6.  Long term effect of latanoprost on intraocular pressure in normal tension glaucoma.

Authors:  A Ang; M A Reddy; L Shepstone; D C Broadway
Journal:  Br J Ophthalmol       Date:  2004-05       Impact factor: 4.638

7.  Point-of-care monitoring of perioperative intraocular pressure using portable tonometry in a patient with Posner-Schlossman syndrome: a case report.

Authors:  Sung-Hoon Kim; Jin-Ho Rhim; Young-Jin Moon; Jihion Yu; Jong-Yeon Park; Ashish Bangaari
Journal:  Korean J Anesthesiol       Date:  2014-03-28

8.  Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL.

Authors:  Ihsan Yilmaz; Cigdem Altan; Ebru Demet Aygit; Cengiz Alagoz; Okkes Baz; Sibel Ahmet; Semih Urvasizoglu; Dilek Yasa; Ahmet Demirok
Journal:  Clin Ophthalmol       Date:  2014-06-07

Review 9.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

10.  Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria.

Authors:  Chizoba Onochie; Obiekwe Okoye; Adunola Ogunro; Tade Aribaba; Kunle Hassan; Adeola Onakoya
Journal:  Med Princ Pract       Date:  2016-08-09       Impact factor: 1.927

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.