Percutaneous aortic valve replacement: an experimental study. I. Studies on implantation.

Georg Lutter1, Daniela Kuklinski, Georg Berg, Patrick Von Samson, Juergen Martin, Michael Handke, Peter Uhrmeister, Friedhelm Beyersdorf.   

Abstract

OBJECTIVE: The purpose of this preliminary study was to devise a new surgical procedure for minimally invasive aortic valve implantation with a transluminal technique.
METHODS: The new collapsible heart valve was prepared by mounting a porcine aortic valve, taken from a freshly slaughtered pig, into a self-expandable nitinol stent by means of a suture technique. The outer diameter of the valved stent ranged from 15 to 23 mm, and the length ranged from 21 to 28 mm. Before implantation in vivo, these valved stents were tested in an in vitro circulatory system. Only in vitro-tested valved stents with a pressure gradient of less than 7 mm Hg and regurgitation of I degrees or less were used for transluminal aortic valve implantation in vivo. Six of these valved stents were implanted in the descending aorta and 8 in the ascending aorta of anesthetized pigs. The catheter delivery system (22F) was extraperitoneally inserted through the left iliac artery or the infrarenal aorta. Measurements for transvalvular gradient, valvular opening and closure, blood-flow characteristics, regurgitation, and macroscopic analysis were performed at baseline and after the observation period (164 +/- 48 minutes).
RESULTS: This preliminary study contained 14 animals. One animal died of ventricular fibrillation. Technical failure occurred in 2 pigs as a result of stent twisting. At the end of the observation period, the 11 successfully implanted valved stents demonstrated low transvalvular gradients (mean end-systolic Deltarho(max) of 5.4 +/- 3.3 mm Hg for the descending aorta group, 5.4 +/- 1.2 mm Hg for the supracoronary group, and 5.4 +/- 1.1 mm Hg for the subcoronary group), which did not differ from their in vitro gradients. Two-dimensional echocardiography demonstrated complete valvular closure and opening in 5 of 5 cases. Angiography indicated only a physiologic jet of regurgitation (0 degrees ) in 8 animals and mild (I degrees ) regurgitation in 3 animals. Color Doppler ultrasonography indicated no regurgitation in 5 of 5 cases and minor paravalvular leakage in 1 case.
CONCLUSION: Aortic valved stents can be successfully implanted without thoracotomy by using a transluminal catheter technique. Long-term function of the valves remains to be established.

Entities:  

Mesh:

Year:  2002        PMID: 11986605     DOI: 10.1067/mtc.2002.121157

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  17 in total

Review 1.  Which strategy for a protein crystallization project?

Authors:  C E Kundrot
Journal:  Cell Mol Life Sci       Date:  2004-03       Impact factor: 9.261

2.  Surgical approaches to vascular access for large-caliber devices in preclinical research models.

Authors:  Noah Barka; Nancy Rakow; Linnea Lentz; Michael Kopcak; Kent Wika; Ana Menk; Mike Green
Journal:  J Am Assoc Lab Anim Sci       Date:  2010-07       Impact factor: 1.232

3.  Significant differences in the material properties between aged human and porcine aortic tissues.

Authors:  Caitlin Martin; Thuy Pham; Wei Sun
Journal:  Eur J Cardiothorac Surg       Date:  2010-12-21       Impact factor: 4.191

4.  First human case of retrograde transcatheter implantation of an aortic valve prosthesis.

Authors:  David Paniagua; José A Condado; José Besso; Manuel Vélez; Bruno Burger; Salvatore Bibbo; Douglas Cedeno; Harry Acquatella; Carlos Mejia; Eduardo Induni; R David Fish
Journal:  Tex Heart Inst J       Date:  2005

5.  Transarterial aortic valve replacement with a self expanding stent in pigs.

Authors:  M Ferrari; H R Figulla; M Schlosser; I Tenner; I Frerichs; C Damm; V Guyenot; G S Werner; G Hellige
Journal:  Heart       Date:  2004-11       Impact factor: 5.994

Review 6.  Transcatheter aortic valve implantation and mitral valve repair: state of the art.

Authors:  C E Ruiz
Journal:  Pediatr Cardiol       Date:  2005 May-Jun       Impact factor: 1.655

7.  Temporary support with Tandemheart pVAD during percutaneous aortic valve replacement in an animal model: rationale and methodology.

Authors:  Biswajit Kar; Reynolds M Delgado; Andrew B Civitello; Pranav Loyalka; David Paniagua; R David Fish; Matthew D Forrester; Daniel Moser; Branislav Radovancevic
Journal:  Tex Heart Inst J       Date:  2005

8.  Transapical aortic valve replacement under real-time magnetic resonance imaging guidance: experimental results with balloon-expandable and self-expanding stents.

Authors:  Keith A Horvath; Dumitru Mazilu; Ozgur Kocaturk; Ming Li
Journal:  Eur J Cardiothorac Surg       Date:  2010-10-22       Impact factor: 4.191

9.  Aortic valve replacement vs. transcatheter aortic valve implantation: Patient selection.

Authors:  Jian Ye; Jia Lin Soon; John Webb
Journal:  Ann Cardiothorac Surg       Date:  2012-07

Review 10.  History of percutaneous aortic valve prosthesis.

Authors:  Henning Rud Andersen
Journal:  Herz       Date:  2009-08       Impact factor: 1.443

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.