Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressure.

Masahiro Kikuya1, Kenichi Chonan, Yutaka Imai, Eiji Goto, Masao Ishii.   

Abstract

OBJECT: Self-measurement of blood pressure (BP) might offer some advantages in diagnosis and therapeutic evaluation and in patient management of hypertension. Recently, wrist-cuff devices for self-measurement of BP have gained more than one-third of the world market share. In the present study, we validated wrist-cuff devices and compared the results between wrist- and arm-cuff devices. The factors affecting the accuracy of wrist-cuff devices were also studied.
METHOD: The research group to assess the validity of automated blood pressure measuring device consisted of 13 institutes in Japan, which validated two wrist-cuff devices (WC-1 and WC-2) and two arm-cuff devices (AC-1 and AC-2). They used a crossover method, where the comparison was done between auscultation, by two observers by means of a double stethoscope on one arm and the device on the opposite arm or wrist.
RESULTS: There was good inter-observer agreement for the auscultation method in each institute (systolic blood pressure (SBP), -0.1 +/- 2.8 mmHg; diastolic blood pressure (DBP), -0.1 +/- 2.6 mmHg, n = 498). The mean difference between auscultation and the device was minimal both in arm-cuff devices (mean difference for AC-1, 2.2/1.9 mmHg, n = 97 and for AC-2, 5.1/2.9 mmHg, n = 136, SBP/DBP) and wrist-cuff devices (mean difference for WC-1, -2.1/1.2 mmHg, n = 173 mmHg and for WC-2, -2.3/-5.6 mmHg, n = 92). The standard deviation of the difference (SDD) in wrist-cuff devices, however (SDD for WC-1, 9.7/7.3 mmHg and for WC-2, 10.2/8.6 mmHg), was larger than that of the arm-cuff devices (SDD for AC-1, 5.6/6.6 mmHg and for AC-2, 6.3/5.1 mmHg). Grading of AC-1 and AC-2 based on criteria of British Hypertension Society was A/A and B/A, respectively, while that of WC-1 and WC-2 was C/B and D/B, respectively. Using the same validation protocol, the results of validation for one device were divergent in each institute. In wrist-cuff devices, the BP value obtained in palmar flexion was significantly higher and that obtained in palmar dorsiflexion was significantly lower than that in palmar extension. In some cases, finger plethysmogram did not disappear during maximum inflation of the wrist-cuff (congruent with 250 mmHg), even in palmar extension and especially in palmar flexion, suggesting incomplete obstruction of radial and/or ulnar arteries during inflation.
CONCLUSION: The results suggest that wrist-cuff devices in the present form are inadequate for self-measurement of blood pressure and, thus, are inadequate for general use or clinical and practical use. However, there is much possibility in wrist-cuff device and the accuracy and reliability of wrist-cuff device are warranted by an improvement of technology.

Entities:  

Mesh:

Year:  2002        PMID: 11910297     DOI: 10.1097/00004872-200204000-00019

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  18 in total

1.  From mercury sphygmomanometer to electric device on blood pressure measurement: correspondence of Minamata Convention on Mercury.

Authors:  Kei Asayama; Takayoshi Ohkubo; Satoshi Hoshide; Kazuomi Kario; Yusuke Ohya; Hiromi Rakugi; Satoshi Umemura
Journal:  Hypertens Res       Date:  2016-01-28       Impact factor: 3.872

Review 2.  Self-measurement of blood pressure at home in the management of hypertension.

Authors:  Hilde Celis; Elly Den Hond; Jan A Staessen
Journal:  Clin Med Res       Date:  2005-02

3.  The impact of arm position and pulse pressure on the validation of a wrist-cuff blood pressure measurement device in a high risk population.

Authors:  Ali Reza Khoshdel; Shane Carney; Alastair Gillies
Journal:  Int J Gen Med       Date:  2010-04-08

4.  Italian society of hypertension guidelines for conventional and automated blood pressure measurement in the office, at home and over 24 hours.

Authors:  Gianfranco Parati; Stefano Omboni; Paolo Palatini; Damiano Rizzoni; Grzegorz Bilo; Mariaconsuelo Valentini; Enrico Agabiti Rosei; Giuseppe Mancia
Journal:  High Blood Press Cardiovasc Prev       Date:  2013-01-22

5.  Wrist devices with both accuracy and feasibility, new option to measure nocturnal blood pressure?

Authors:  Ran Heo; Jinho Shin
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-04-13       Impact factor: 3.738

6.  Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010.

Authors:  Hakuo Takahashi; Masamichi Yoshika; Toyohiko Yokoi
Journal:  Vasc Health Risk Manag       Date:  2013-05-28

7.  Reliability of automated blood pressure devices used by hypertensive patients.

Authors:  William C W Wong; Ivan K L Shiu; Thomas M T Hwong; James A Dickinson
Journal:  J R Soc Med       Date:  2005-03       Impact factor: 18.000

8.  Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage.

Authors:  Sakir Uen; Rolf Fimmers; Miriam Brieger; Georg Nickenig; Thomas Mengden
Journal:  BMC Cardiovasc Disord       Date:  2009-05-27       Impact factor: 2.298

9.  Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-02-25       Impact factor: 3.738

10.  Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T.

Authors:  Mitsuo Kuwabara; Kanako Harada; Yukiko Hishiki; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-01-04       Impact factor: 3.738