| Literature DB >> 1189628 |
Abstract
Various biological factors affecting breath-alcohol analysis were studied experimentally. End-expiratory temperatures in 55 healthy subjects were found to range from 32.41 to 35.69 degrees C with a mean of 34.53 degrees C. Forced vital capacity in the same subjects ranged from 1825 to 6550 ml with a mean of 4038 ml, and maximum exhalation after normal inhalation ranged from 1180 to 4550 ml with a mean of 2730 ml. It was found that 65-70% of available breath must be discarded before the alveolar plateau is reached during expiration. End-expiratory (alveolar) carbon dioxide in 155 healthy subjects was 3.5-8.3% by volume (mean = 6.52). After oral alcohol intake, retained mouth-alcohol in 8 subjects had disappeared after 11 minutes without subsequent water-rinsing of the mouth, and after 8 minutes with rinsing. Water condensation in plastic mouthpieces/saliva traps during breath sampling yielded mean weight gains of 13.0, 8.6, and 4.6 mg., respectively, at initial mouthpiece temperatures of 3 degrees C, 22.5 degrees C, and 34.7 degrees C, respectively.Entities:
Mesh:
Substances:
Year: 1975 PMID: 1189628 DOI: 10.1007/bf00200555
Source DB: PubMed Journal: Z Rechtsmed ISSN: 0044-3433