| Literature DB >> 11885679 |
Timothy A Ritter1, Thomas R Shrout, Rick Tutwiler, K Kirk Shung.
Abstract
Ultrasound imaging at frequencies above 20 MHz is capable of achieving improved resolution in clinical applications requiring limited penetration depth. High frequency arrays that allow real-time imaging are desired for these applications but are not yet currently available. In this work, a method for fabricating fine-scale 2-2 composites suitable for 30-MHz linear array transducers was successfully demonstrated. High thickness coupling, low mechanical loss, and moderate electrical loss were achieved. This piezo-composite was incorporated into a 30-MHz array that included acoustic matching, an elevation focusing lens, electrical matching, and an air-filled kerf between elements. Bandwidths near 60%, 15-dB insertion loss, and crosstalk less than -30 dB were measured. Images of both a phantom and an ex vivo human eye were acquired using a synthetic aperture reconstruction method, resulting in measured lateral and axial resolutions of approximately 100 microm.Entities:
Mesh:
Year: 2002 PMID: 11885679 DOI: 10.1109/58.985706
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725