Literature DB >> 11885003

Real-time-MR guidance for placement of a self-made fully MR-compatible atrial septal occluder: in vitro test.

A Buecker1, E Spuentrup, R Grabitz, F Freudenthal, T Schaeffter, J J van Vaals, W Günther.   

Abstract

PURPOSE: This in vitro study investigated the feasibility to visualize the placement of three different atrial septal occluder systems using real-time MR control.
METHODS: The experiments were performed on an interventional 1.5 T high field whole body system. Real-time MR imaging was achieved by radial or spiral k-space filling in conjunction with the sliding window reconstruction technique yielding an imaging speed of 15 frames per second. The CardioSeal, Amplatzer Septal Occluder and a specially designed MR-compatible closure device were tested in a water bath. A punctured plastic wall served as model for the atrial septal defect.
RESULTS: The delivery systems of the CardioSeal and Amplatzer Occluder were ferromagnetic and caused substantial artifacts, making the device placement impossible, even if the magnetic forces would have been acceptable. The self-made prototype caused only minor susceptibility artifacts allowing its visualization on the MR images. The MR imaging techniques applied enabled real-time control of the occluder including steering through the artificial septal foramen and visualization of the occluder deployment.
CONCLUSION: Real-time MR imaging allows for guidance and placement of an MR-compatible septal occluder in vitro suggesting the feasibility to perform atrial septal occlusion under MR-guidance in vivo as well.

Entities:  

Mesh:

Year:  2002        PMID: 11885003     DOI: 10.1055/s-2002-20601

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  2 in total

1.  Evaluation of MR imaging compatibility of a new intracranial pressure monitoring device.

Authors:  Jens O Heidenreich; Rüdiger Stendel; Mario Brock; Karl-Jürgen Wolf; Andreas M Schilling
Journal:  AJNR Am J Neuroradiol       Date:  2005-02       Impact factor: 3.825

2.  Atrial septal defects type II: noninvasive evaluation of patients before implantation of an Amplatzer Septal Occluder and on follow-up by magnetic resonance imaging compared with TEE and invasive measurement.

Authors:  Christoph Weber; Michael Weber; Okan Ekinci; Thomas Neumann; Anja Deetjen; Andreas Rolf; Gerhard Adam; Christian W Hamm; Thorsten Dill
Journal:  Eur Radiol       Date:  2008-06-21       Impact factor: 5.315

  2 in total