Literature DB >> 11846941

Comparison of frameless stereotactic systems: accuracy, precision, and applications.

E A Benardete1, M A Leonard, H L Weiner.   

Abstract

OBJECTIVE: Frameless stereotactic systems have become an integral part of neurosurgical practice. At our center, we recently introduced for clinical use a small, portable, frameless stereotactic system, namely the Cygnus PFS system (Compass International, Rochester, MN). The purpose of this study was to compare the accuracy of the Cygnus PFS system with that of two larger systems that are also currently in use at our institution, i.e., the SMN system (Zeiss, Oberkochen, Germany) and the ISG viewing wand (ISG Technologies, Toronto, Canada). These systems represent three kinds of frameless stereotactic technologies that are commercially available. Each system uses a different method of spatial localization, i.e., mechanical linkage (ISG system), magnetic field digitization (Cygnus system), or optical technology (SMN system).
METHODS: Using a stereotactic "phantom," we measured the accuracies of all three systems with identical data sets. The errors in localization in three-dimensional space for nine targets were calculated by using 10 magnetic resonance imaging data sets. The precision of each system was also calculated.
RESULTS: With this experimental protocol, the Cygnus system attained a mean accuracy of 1.90 +/- 0.7 mm, the ISG viewing wand system a mean accuracy of 1.67 +/-0.43 mm, and the SMN microscope a mean accuracy of 2.61 +/- 0.99 mm. The precision values were not significantly different among the systems.
CONCLUSION: We observed only small differences in accuracy and precision among these three systems. We briefly review the advantages and disadvantages of each system and note that other factors, such as portability, ease of use, and microscope integration, should influence the selection of a frameless stereotactic system.

Mesh:

Year:  2001        PMID: 11846941     DOI: 10.1097/00006123-200112000-00020

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  6 in total

1.  Smart stylet: the development and use of a bedside external ventricular drain image-guidance system.

Authors:  Vaibhav Patil; Rajiv Gupta; Raúl San José Estépar; Ronilda Lacson; Arnold Cheung; Judith M Wong; A John Popp; Alexandra Golby; Christopher Ogilvy; Kirby G Vosburgh
Journal:  Stereotact Funct Neurosurg       Date:  2015-01-31       Impact factor: 1.875

2.  Comparative study of application accuracy of two frameless neuronavigation systems: experimental error assessment quantifying registration methods and clinically influencing factors.

Authors:  Dimitrios Paraskevopoulos; Andreas Unterberg; Roland Metzner; Jens Dreyhaupt; Georg Eggers; Christian Rainer Wirtz
Journal:  Neurosurg Rev       Date:  2011-01-19       Impact factor: 3.042

Review 3.  Advances in local ablation of malignant liver lesions.

Authors:  Robert M Eisele
Journal:  World J Gastroenterol       Date:  2016-04-21       Impact factor: 5.742

4.  Utility of Image Guidance in the Localization of Disappearing Colorectal Liver Metastases.

Authors:  Linda M Pak; Johan Gagnière; Peter J Allen; Vinod P Balachandran; Michael I D'Angelica; Ronald P DeMatteo; William R Jarnagin; Michael I Miga; Amber L Simpson; T Peter Kingham
Journal:  J Gastrointest Surg       Date:  2019-01-24       Impact factor: 3.452

5.  [Stereotaxy and robotics for ablation - toy or tool?].

Authors:  R Bale; G Widmann; W Jaschke
Journal:  Radiologe       Date:  2012-01       Impact factor: 0.635

6.  Stereotactic Neuro-Navigation Phantom Designs: A Systematic Review.

Authors:  Marko Švaco; Ivan Stiperski; Domagoj Dlaka; Filip Šuligoj; Bojan Jerbić; Darko Chudy; Marina Raguž
Journal:  Front Neurorobot       Date:  2020-10-23       Impact factor: 2.650

  6 in total