Abstract

Bone cement with reduced amount of monomer and low curing temperature may improve implant fixation due to reduced toxicity. We analyzed the mechanical, chemical and thermal properties of such a cement (Cemex Rx) using Palacos R as control. The in vivo performance of the 2 cements was also evaluated in a prospective randomized study of 47 hips, where either of the cement types was used to fixate Lubinus SP2 prostheses with the stem made of titanium alloy. Cemex Rx had a reduced tensile strength, probably because this cement was manually mixed, as recommended by the manufacturer. A standardized labor tory test showed lower curing temperature for Cemex, but measurements at 37 degrees and with prechilled Palacos R and Cemex Rx, as in clinical work, showed no difference. In the clinical study radiostereometric measurements of cup and stem migration showed similar values in the 2 groups up to 5 years after the operation. The cement mantle was stable in both groups, but the stems migrated similarly inside the cement mantle regardless of the type of cement used. Proximal wear was low (0.04-0.05 mm/year) and tended to be lower in the Cemex group (p = 0.02). Aluminum and vanadium levels in serum increased 5 years after the operation, but no difference was noted between the 2 groups. Collagen markers (PICP, ICTP) showed similar increases in bone turnover 6 weeks and 6 months after operation in both groups.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11817871     DOI: 10.1080/000164701317268987

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  9 in total

1.  Length of clinically proven cemented hip stems: state of the art or subject to improvement?

Authors:  Moussa Hamadouche; Alexander Jahnke; Caroline Scemama; Bernd Alexander Ishaque; Markus Rickert; Luc Kerboull; Eike Jakubowitz
Journal:  Int Orthop       Date:  2014-09-20       Impact factor: 3.075

2.  Reduced femoral component subsidence with improved impaction grafting at revision hip arthroplasty.

Authors:  D W Howie; S A Callary; M A McGee; N C Russell; L B Solomon
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

3.  Less wear with aluminium-oxide heads than cobalt-chrome heads with ultra high molecular weight cemented polyethylene cups: a ten-year follow-up with radiostereometry.

Authors:  Jon Dahl; Per Söderlund; Bo Nivbrant; Lars Nordsletten; Stephan M Röhrl
Journal:  Int Orthop       Date:  2011-08-26       Impact factor: 3.075

4.  Conversion to hypertrophic vertebral pseudarthrosis following percutaneous vertebroplasty.

Authors:  Satoshi Kawaguchi; Keiko Horigome; Hideki Yajima; Takashi Oda; Yuichiro Kii; Mitsunori Yoshimoto; Tsuneo Takebayashi; Toshihiko Yamashita
Journal:  Eur Spine J       Date:  2010-02-04       Impact factor: 3.134

Review 5.  Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies.

Authors:  Paul van der Voort; Bart G Pijls; Marc J Nieuwenhuijse; Jorrit Jasper; Marta Fiocco; Josepha W M Plevier; Saskia Middeldorp; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2015       Impact factor: 3.717

Review 6.  Adherence of hip and knee arthroplasty studies to RSA standardization guidelines. A systematic review.

Authors:  Rami Madanat; Tatu J Mäkinen; Hannu T Aro; Charles Bragdon; Henrik Malchau
Journal:  Acta Orthop       Date:  2014-06-23       Impact factor: 3.717

7.  Is the French Paradox cementing philosophy superior to the standard cementing? A randomized controlled radiostereometric trial and comparative analysis.

Authors:  Kirsti Sevaldsen; Otto Schnell Husby; Øystein Bjerkestrand Lian; Kamel Mohamed Farran; Vigdis Schnell Husby
Journal:  Bone Joint J       Date:  2022-01       Impact factor: 5.082

8.  10-year results of a new low-monomer cement: follow-up of a randomized RSA study.

Authors:  Per Söderlund; Jon Dahl; Stephan Röhrl; Bo Nivbrant; Kjell G Nilsson
Journal:  Acta Orthop       Date:  2012-11-01       Impact factor: 3.717

9.  The effect of high dose antibiotic impregnated cement on rate of surgical site infection after hip hemiarthroplasty for fractured neck of femur: a protocol for a double-blind quasi randomised controlled trial.

Authors:  Andrew P Sprowson; Cyrus D Jensen; Sanjay Gupta; Nick Parsons; Aradhyula N Murty; Simon M G Jones; Dominic Inman; Mike R Reed
Journal:  BMC Musculoskelet Disord       Date:  2013-12-17       Impact factor: 2.362

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.