Literature DB >> 11810678

Elimination of Nyquist ghosting caused by read-out to phase-encode gradient cross-terms in EPI.

Stuart M Grieve1, Andrew M Blamire, Peter Styles.   

Abstract

Echo-planar imaging (EPI) commonly suffers from ghosting artifacts caused by zero- or first-order phase differences between the odd and even echoes that constitute an EPI dataset. Small-bore imaging systems with shielded gradients may suffer significantly from cross-term eddy currents due to the high degree of manufacturing precision required in such systems compared to larger whole-body coils. A Nyquist ghost caused by cross-term eddy current contributions from the read-out to the phase-encode axis was identified in a small-bore system and characterized using a modified EPI experiment and a computer simulation. The artifact was corrected for using both a postprocessing approach and compensation blips along the phase-encode axis. Correction using compensation blips proved to be a more effective strategy to reduce this artifact than the postprocessing method used. Copyright 2002 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2002        PMID: 11810678     DOI: 10.1002/mrm.10055

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  10 in total

1.  A correction method for streak artifacts in gradient-echo EPI using spin-echo EPI reference data.

Authors:  Jun-Young Chung; Yeji Han; Zang-Hee Cho; Hyunwook Park
Journal:  MAGMA       Date:  2011-11-10       Impact factor: 2.310

2.  MODEL-BASED IMAGE RECONSTRUCTION FOR MRI.

Authors:  Jeffrey A Fessler
Journal:  IEEE Signal Process Mag       Date:  2010-07-01       Impact factor: 12.551

3.  Two-dimensional phase cycled reconstruction for inherent correction of echo-planar imaging Nyquist artifacts.

Authors:  Nan-Kuei Chen; Alexandru V Avram; Allen W Song
Journal:  Magn Reson Med       Date:  2011-03-28       Impact factor: 4.668

4.  Navigator-Free EPI Ghost Correction With Structured Low-Rank Matrix Models: New Theory and Methods.

Authors:  Rodrigo A Lobos; Tae Hyung Kim; W Scott Hoge; Justin P Haldar
Journal:  IEEE Trans Med Imaging       Date:  2018-04-02       Impact factor: 10.048

5.  Model-based iterative reconstruction for single-shot EPI at 7T.

Authors:  Uten Yarach; Myung-Ho In; Itthi Chatnuntawech; Berkin Bilgic; Frank Godenschweger; Hendrik Mattern; Alessandro Sciarra; Oliver Speck
Journal:  Magn Reson Med       Date:  2017-02-10       Impact factor: 4.668

6.  Joint correction of Nyquist artifact and minuscule motion-induced aliasing artifact in interleaved diffusion weighted EPI data using a composite two-dimensional phase correction procedure.

Authors:  Hing-Chiu Chang; Nan-Kuei Chen
Journal:  Magn Reson Imaging       Date:  2016-04-22       Impact factor: 2.546

Review 7.  Diffusion-weighted imaging in head and neck cancers.

Authors:  Sanjeev Chawla; Sungheon Kim; Sumei Wang; Harish Poptani
Journal:  Future Oncol       Date:  2009-09       Impact factor: 3.404

8.  Robust autocalibrated structured low-rank EPI ghost correction.

Authors:  Rodrigo A Lobos; W Scott Hoge; Ahsan Javed; Congyu Liao; Kawin Setsompop; Krishna S Nayak; Justin P Haldar
Journal:  Magn Reson Med       Date:  2020-12-17       Impact factor: 3.737

9.  Characterization of Nyquist ghost in EPI-fMRI acquisition sequences implemented on two clinical 1.5 T MR scanner systems: effect of readout bandwidth and echo spacing.

Authors:  Marco Giannelli; Stefano Diciotti; Carlo Tessa; Mario Mascalchi
Journal:  J Appl Clin Med Phys       Date:  2010-07-12       Impact factor: 2.102

10.  Quantitative Non-Gaussian Intravoxel Incoherent Motion Diffusion-Weighted Imaging Metrics and Surgical Pathology for Stratifying Tumor Aggressiveness in Papillary Thyroid Carcinomas.

Authors:  David Aramburu Núñez; Yonggang Lu; Ramesh Paudyal; Vaios Hatzoglou; Andre L Moreira; Jung Hun Oh; Hilda E Stambuk; Yousef Mazaheri; Mithat Gonen; Ronald A Ghossein; Ashok R Shaha; R Michael Tuttle; Amita Shukla-Dave
Journal:  Tomography       Date:  2019-03