MR imaging of extracapsular silicone from breast implants: diagnostic pitfalls.

Wendie A Berg1, Thuy Khanh Nguyen, Michael S Middleton, Mary Scott Soo, Gene Pennello, S Lori Brown.   

Abstract

OBJECTIVE: We sought to identify pitfalls in recognition of extracapsular silicone on MR imaging.
MATERIALS AND METHODS: Three experienced observers reviewed MR images from 359 women with current (n = 320), prior (n = 15), or both current and prior (n = 24) silicone gel implants. Axial and sagittal fast spin-echo T2-weighted images with water suppression, axial inversion-recovery T2-weighted images with water suppression, and axial T2-weighted images with silicone suppression were obtained in a dedicated phased array breast coil on a 1.5-T magnet. Images were reviewed again when only one observer saw extracapsular silicone, and reasons for disagreement were recorded.
RESULTS: Rupture was identified in 265 women (77%) with current silicone implants and 378 (55%) of 687 implants. Observers agreed in describing extracapsular silicone in 85 (12%) of 687 breasts with current silicone gel implants, of which 81 (95%) showed definite evidence of rupture on MR imaging. One observer reported extracapsular silicone in another 79 breasts. Confusion over contour deformity due to weakening versus breach of the capsule accounted for 33 (42%) of 79 disagreements. Another 20 (25%) of the 79 disagreements were attributed to poor conspicuity of extracapsular silicone on fast spin-echo T2-weighted images combined with intermittent observer failure to review inversion-recovery images. Subtlety of findings (n = 17, 22%) and technical issues (n = 9, 11%) with failed water suppression of pleural effusion or cysts and ghosting artifacts accounted for remaining disagreements.
CONCLUSION: Extracapsular rupture is usually manifest as local spread of silicone in the breast and is not well-depicted on fast spin-echo T2-weighted images. Water-suppressed inversion-recovery T2-weighted images are often needed to identify extracapsular silicone. Distinction of the bulge in the fibrous capsule from herniation through the capsule remains problematic.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11804919     DOI: 10.2214/ajr.178.2.1780465

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  11 in total

1.  Role of US in evaluating breast implant integrity.

Authors:  Michele Telegrafo; Marco Moschetta
Journal:  J Ultrasound       Date:  2015-06-23

2.  Indications for breast magnetic resonance imaging. Consensus document "Attualità in senologia", Florence 2007.

Authors:  F Sardanelli; G M Giuseppetti; G Canavese; L Cataliotti; S Corcione; E Cossu; M Federico; L Marotti; L Martincich; P Panizza; F Podo; M Rosselli Del Turco; C Zuiani; C Alfano; M Bazzocchi; P Belli; S Bianchi; A Cilotti; M Calabrese; L Carbonaro; L Cortesi; C Di Maggio; A Del Maschio; A Esseridou; A Fausto; M Gennaro; R Girometti; R Ienzi; A Luini; S Manoukian; S Morassutt; D Morrone; J Nori; A Orlacchio; F Pane; P Panzarola; R Ponzone; G Simonetti; P Torricelli; G Valeri
Journal:  Radiol Med       Date:  2008-10-16       Impact factor: 3.469

3.  MRI screening for silicone breast implant rupture: accuracy, inter- and intraobserver variability using explantation results as reference standard.

Authors:  M C Maijers; F B Niessen; J F H Veldhuizen; M J P F Ritt; R A Manoliu
Journal:  Eur Radiol       Date:  2014-03-06       Impact factor: 5.315

Review 4.  Imaging findings of mammary and systemic silicone deposition secondary to breast implants.

Authors:  Naziya Samreen; Katrina N Glazebrook; Asha Bhatt; Sudhakar K Venkatesh; Brendan P McMenomy; Anupam Chandra; Shuai Leng; Kalie E Adler; Cynthia H McCollough
Journal:  Br J Radiol       Date:  2018-05-10       Impact factor: 3.039

5.  Silicone Granuloma Associated with Pectoral Muscle Involvement after Ruptured Breast Implant: a Novel case report.

Authors:  Türkan İkizceli; Gökçe Gülşen; İlker Akın
Journal:  Eur J Breast Health       Date:  2018-01-01

6.  MRI in the evaluation of facial dermal fillers in normal and complicated cases.

Authors:  Marco Di Girolamo; Mauro Mattei; Alberto Signore; Francesca Romana Grippaudo
Journal:  Eur Radiol       Date:  2014-12-05       Impact factor: 5.315

7.  Imaging of breast implants-a pictorial review.

Authors:  Sergi Juanpere; Elsa Perez; Oscar Huc; Naiara Motos; Josep Pont; Salvador Pedraza
Journal:  Insights Imaging       Date:  2011-08-07

8.  Removing silicone artifacts in diffusion-weighted breast MRI by means of shift-resolved spatiotemporally encoding.

Authors:  Eddy Solomon; Noam Nissan; Rita Schmidt; Edna Furman-Haran; Uriel Ben-Aharon; Lucio Frydman
Journal:  Magn Reson Med       Date:  2015-06-22       Impact factor: 4.668

9.  Poly implant breast implants (PIP) and the rupture risk in asymptomatic patients: a warning for greater clinician suspicion in assessment and counselling.

Authors:  Senthurun Mylvaganam; Rebecca Taylor; Steven Thrush
Journal:  J Surg Case Rep       Date:  2013-02-06

10.  Breast MRI: guidelines from the European Society of Breast Imaging.

Authors:  R M Mann; C K Kuhl; K Kinkel; C Boetes
Journal:  Eur Radiol       Date:  2008-04-04       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.