Multislice CT angiography of the iliac arteries in the presence of various stents: in vitro evaluation of artifacts and lumen visibility.

D Maintz1, R Fischbach, K U Juergens, T Allkemper, J Wessling, W Heindel.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate the imaging characteristics of various iliac artery stents and stent-grafts in a multislice, computed tomography angiography (MSCTA) phantom study.
METHODS: Twelve types of stents and three stent-grafts made of different materials (steel, nitinol, tantalum, cobalt-based alloy) were implanted in plastic tubes with an 8-mm inner diameter filled with iodinated contrast medium diluted to 200 HU. To evaluate the influence of scan parameters on artifacts, the Palmaz stent (as one example) was scanned with a four-slice scanner (Siemens VolumeZoom) with different detector collimations and pitches. All other stents were examined with a collimation of 4 x 1 mm and a table feed of 4 mm/rotation. Axial images and multiplanar reformations were evaluated regarding artifact size, lumen visibility, and intraluminal attenuation values.
RESULTS: Higher pitch values caused more pronounced artifacts. Image quality and delineation of the stent struts improved with thinner detector collimation. The size of the stent-related artifacts and the visibility of the stent lumen depended on the underlying stent material and design. Pronounced artifacts, resulting in an insufficient delineation of the stent lumen, were caused by the Strecker tantalum stent and the Zenith stent. Moderate artifacts were caused by the Wallgraft, Passager, Palmaz P 395 and P 424, Bridge, Perflex, and ZA stents. Some artificial lumen narrowing but good lumen delineation was seen with the Strecker nitinol, Symphony, Memotherm, SMART, Corinthian, and Wallstent stents.
CONCLUSIONS: Scanning parameters, stent material, and stent design influence lumen assessability and image quality in MSCTA. Detailed lumen assessment is impossible in the Strecker tantalum and Zenith stents and limited in the Wallgraft, Passager, Palmaz P 395 and PS 424, Bridge, Perflex, and ZA-stent stent-grafts but feasible in all other examined products.

Entities:  

Mesh:

Year:  2001        PMID: 11753140     DOI: 10.1097/00004424-200112000-00004

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  7 in total

1.  Assessment of coronary artery stents using 16-slice MDCT angiography: evaluation of a dedicated reconstruction kernel and a noise reduction filter.

Authors:  Harald Seifarth; Rainer Raupach; Stefan Schaller; Eva Maria Fallenberg; Thomas Flohr; Walter Heindel; Roman Fischbach; David Maintz
Journal:  Eur Radiol       Date:  2005-02-12       Impact factor: 5.315

Review 2.  Transitioning from 16-slice to 64-slice multidetector computed tomography for the assessment of coronary artery disease: are we really making progress?

Authors:  Razi Khan; Sapna Rawal; Mark J Eisenberg
Journal:  Can J Cardiol       Date:  2009-09       Impact factor: 5.223

3.  Comparison of 16-slice computed tomography with conventional angiography to evaluate coronary artery stent patency.

Authors:  Irmak Durur-Subasi; Mecit Kantarci; Afak Durur-Karakaya; Adnan Okur
Journal:  Eurasian J Med       Date:  2009-04

4.  Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings.

Authors:  Martin Heuschmid; Benjamin Wiesinger; Gunnar Tepe; Oliver Luz; Andreas F Kopp; Claus D Claussen; Stephan H Duda
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

5.  Multidetector-row CT Angiography of Lower Extremities: Usefulness in the Diagnosis of and Intervention for Peripheral Arterial Disease.

Authors:  Kazuhiro Matsumoto; Masahiro Jinzaki; Kozo Sato; Yutaka Tanami; Seishi Nakatsuka; Subaru Hashimoto; Sachio Kuribayashi
Journal:  Ann Vasc Dis       Date:  2010-12-25

6.  Peripheral artery stent visualization and in-stent stenosis analysis in 16-row computed tomography: an in-vitro evaluation.

Authors:  Christopher Herzog; Christine Grebe; Andreas Mahnken; Joern O Balzer; Martin G Mack; Stefan Zangos; Hanns Ackermann; Stefan Schaller; Tobias Seifert; Bernd Ohnesorge; Thomas J Vogl
Journal:  Eur Radiol       Date:  2005-07-14       Impact factor: 5.315

7.  Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types.

Authors:  Michael Köhler; Matthias C Burg; Alexander C Bunck; Walter Heindel; Harald Seifarth; David Maintz
Journal:  Radiol Res Pract       Date:  2011-07-18
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.