Literature DB >> 11686743

Experiences with the European guidelines on quality criteria for radiographic images in Tanzania.

W E Muhogora1, A M Nyanda, R R Kazema.   

Abstract

Objective assessment of the quality of radiographic images is practically a difficult task and protocols that address this problem are few. In 1996, the European union published nearly objective image quality criteria to unify the practices in Europe. However, experience with these criteria in countries of lower health care levels is little documented. As a case study in Tanzania, we present the general performance of European guidelines in some Tanzanian hospitals to a total of 200 radiographs obtained from some common x-ray examinations. The results show that more than 70% of chest (PA), lumbar spine (AP), and pelvis AP radiographs passed the quality criteria, while the performance of lumbar spine LAT x-ray examinations was about 50% and therefore less satisfactory. The corresponding mean entrance dose to the patient for specified x-ray techniques was of range 0.08-0.56 mGy, 3.1-7.7 mGy, 2.53-5.4 mGy, and 4.0-16.78 mGy for chest PA, lumbar spine AP, pelvis AP, and lumbar spine LAT x-ray examinations, respectively. Although a good number of observers were not well familiar to the guidelines, the quality criteria have been found useful and their adoption in the country recommended. The need to provide relevant education and training to staff in the radiology departments is of utmost importance.

Entities:  

Mesh:

Year:  2001        PMID: 11686743      PMCID: PMC5726009          DOI: 10.1120/jacmp.v2i4.2601

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Despite a small but increasing hazard of diagnostic x‐rays to human beings, , studies aimed at achieving low patient doses with sufficient image quality have continued to be the area of research interest. The relationship between the quality of radiographic image and dose to patient is known to depend on the performance characteristics of the x‐ray equipment, patient shape and size, type of image receptors, radiographic techniques, viewing conditions as well as staff experiences. , Protocols that objectively establish this relationship, are however, few worldwide and inexistent in many countries such as Tanzania. Recently, the European union has published a set of nearly objective guidelines for good radiographic techniques and the corresponding level of the image quality. The guidelines have proved to be a useful tool to unify the practices in Europe. In attempts to address the problem on dose reduction without affecting the patient care, a research on radiation protection in diagnostic radiology was initiated in Tanzania. The hospitals under study were MMC (Muhimbili Medical Center), KCMC (Kilimanjaro Christian Medical Center), BMC (Bugando Medical Center), MCH (Mbeya Consultant Hospital), and ARH (Arusha Regional Hospital). A study on film rejects rate analysis revealed varying radiographic techniques and skills (positioning, centering, tube voltage, and tube current‐time product selection), as well as patient influences as mainly responsible to more than 90% of waste film. The average magnitude of waste film per week for a given x‐ray examination was as summarized in Table I.
Table I

The average number of waste film per week for a given examination.

HospitalNumber of x‐ray roomsNumber of waste film per week
Chest APLumbar spine APLumbar spine LATPelvis AP
MMC55482563
KCMC323523
BMC313389
MCH1203514
ARH13332
The average number of waste film per week for a given examination. Patient dose assessment identified four methods that could reduce unnecessary dose to patients. The measures included lowering tube current‐time product, increasing tube voltage, increasing filtration, increasing the image receptor speed, and improving the processing conditions. During this study, up to 50% of radiographs were rated poor based on European guidelines on quality criteria for radiographic images. Following this magnitude of poor radiographs, consultancy with radiologists in attendance of radiographers was made on possible strategies to reduce the waste film level. Chief amongst the strategies were the use of 400‐speed class of film‐screen combination instead of the usual 200‐speed class, improvement and maintenance of film processing conditions, use of exposure charts, and adherence to the recommended practices of good imaging. Later on, the lowering of tube current‐time product and increasing the speed of film‐screen combination and optimizing film processing conditions were practically demonstrated to be the most effective dose reduction measures in the country. Recently, a follow‐up study on the effectiveness of dose reduction measures in general clinical conditions as a reliable indicator for optimization has been reported, where patient doses were reduced up to 77%. During this follow‐up study, the quality of radiographic images obtained after the implementation of dose reduction measures was again evaluated using the European quality criteria and the results are hereby presented. The fact that no similar study has been done in the country, the foreseen research results could provide first experience on the subject in the country and additional experience elsewhere.

METHODS

The study was conducted at the same hospitals using the same x‐ray equipment and film processors as during previous studies. , , , Four x‐ray projections namely chest (PA), pelvis (AP), lumbar spine (AP), and lateral (LAT) were employed for the intended study. Ten adult patients of weight ranging from 60 to 80 kg and thickness between 17 and 25 cm were selected for each of the four x‐ray projections at each hospital giving a collective sample of 200 patients. The employed x‐ray equipment had their tube voltage and tube current‐time product settings verified by noninvasive x‐ray test device, Victoreen model . The film processing conditions at each hospital were also optimized and maintained accordingly. This implies that the study was conducted to the x‐ray equipment whose performance was sufficiently known. Details on x‐ray equipment employed are summarized in Table II. During x‐ray examination, the requirements and technical parameters of good imaging were closely followed as recommended. The mean ESD ranges and the corresponding summary of selected techniques at each hospital are presented in Table III. The quality of the obtained radiographs per x‐ray projection was tested for compliance with the European guidelines on quality criteria for radiographic images by five well‐experienced radiologists. Three observers (including one of the authors, R.R.K.) are consultant radiologists with academic qualification at a postgraduate level in radiology and a working experience of over ten years. The other two observers have postgraduate diploma in radiology and have worked for three years under a consultant radiologist and over four years independently. According to the European guidelines, the image criteria refer to characteristic features of imagined anatomic structures of each radiograph with a specific degree of visibility. The observers evaluated the image quality of all ten radiographs of each x‐ray projection (according to the basis indicated in Table IV) against all anatomical structures (Tables V–VIII). Along with image quality assessment using the European quality criteria, optical densities of a sample of radiographs were also measured using a RMI densitometer, serial number 211–2176F to establish typical values for achieved optimized conditions.
Table II

Type of x‐ray equipment and film processors at hospitals.

HospitalX‐ray machineProcessorFocus to film distance (FFD)HVL at 80 kVp (mm Al)
MMCPhillips, model Emerald 125automatic, Kodak RPX Omat model M6B150 cm: chest PA 100 cm: rest x‐ray examinations3.1
Phillips, model U‐5BC‐41T3.4
KCMCPhillips, model Emerald 125 RPX Omatautomatic, Kodak model M6B150 cm: chest PA 100 cm: rest x‐ray examinations4.0 a
Phillips, model Emerald 1254.7
BMCSiemens, model 125/20/40–80manual150 cm: chest PA 100 cm: rest x‐ray examinations2.8
Shimadzu, model Circlex 1/2P/18C3.7
MCHShimadzu, model Circlex 1/2P/80–806manual150 cm: chest PA 100 cm: rest x‐ray examinations2.1
ARHPhillips medio 51, model Rotalixmanual150 cm: chest PA 100 cm: rest x‐ray examinations3.6

HVL measurements done at 120 kVp.

Table III

Mean entrance surface dose to patients at hospitals.

Hospital x‐ray examinationMean tube voltage range (kVp)Mean tube current‐time product range (mAs)Speed class of film‐screen combinationMean ESD range (mGy)
MMCchest PA76–839.8–22400/2000.2–0.36
pelvis AP75–80120–143.7400/2002.53–5.4
lumbar spine AP70–8089–120400/2003.87–4.94
lumbar spine LATERAL79–80207.1–240.7400/20010.26–16.78
KCMCchest PA pelvis AP1201.24000.08–0.01a
lumbar spine AP lumbar spine LATERAL70–73554003.84–4.60b
BMCchest PA59.7–68.525.3–40400/2000.22–0.39
pelvis AP77–81125400/2002.9–3.1
lumbar spine AP88–90151.7–1802006.5–7.7
lumbar spine LATERAL90–120.4160–2504009.83–14.4
ARHchest PA70–87.510–21.62000.20–0.34c
pelvis AP9033–352003.0–3.9c
lumbar spine AP80–9070–80400/2003.1–3.3c
lumbar spine LATERAL9070–90400/2004.0–6.2

Note: Some ESD ranges include patient doses obtained with additional filtration of 1 mm Al, hence HVL of , , and Al (see also Table II).

Table IV

The basis employed during the assessment of the quality of radiographic images.

Image criteriaDegree of visibilityScore
‐Visualization of characteristic features.‐feature detected and fully reproduced‐good
‐feature just visible‐satisfactory
‐feature not visible‐poor
‐Reproduction of anatomical structures.‐detail visible and clearly defined‐good
‐detail not visible or not clear‐poor
‐Visually sharp reproduction‐details clearly defined‐good
‐details just clear‐satisfactory
‐details not clear‐poor
Table V

Compliance with European guidelines for 50 chest PA x‐ray examinations.

Image criteriaNumber of scores per score category a
Good Satisfactory Poor
‐Performed at deep inspiration (as assessed by the position of the ribs above the diaphragm either 6 anteriorly or 10 posteriorly)5000
‐Symmetrical reproduction of the thorax14306
‐Medial border of the scapulae to be outside the lung field02525
‐Reproduction of the whole rib cage above the diaphragm21195
‐Reproduction of the vascular pattern in the whole lung particularly the peripheral vessels122810
‐Visually sharp reproduction of
  (a) the trachea and proximal bronchi, the borders of the heart and aorta151012
  (b) the diaphragm and costo‐phranic angles13128
‐Visualization of the retrocardia lung and the mediastinum17266

‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear

Table VIII

Compliance with European guidelines for 50 pelvis AP x‐ray examinations.

Image criteriaNumber of scores per score category a
Good Satisfactory Poor
‐Symmetrical reproductionofthe pelvis30200
‐Visualization of the sacrum and its intervertebral foramina101426
‐Visualization of the sacroiliac joints122018
‐Reproduction of the neck of the femora which should not be distorted by foreshortening or rotation04010
‐Reproduction of spongiosa and corticalis and visualization of the trochanters26816

‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear

Type of x‐ray equipment and film processors at hospitals. HVL measurements done at 120 kVp. Mean entrance surface dose to patients at hospitals. Note: Some ESD ranges include patient doses obtained with additional filtration of 1 mm Al, hence HVL of , , and Al (see also Table II). The basis employed during the assessment of the quality of radiographic images.

RESULTS AND DISCUSSIONS

Tables V to VIII give the results of image quality assessment based on the European guidelines. Typical densitometry results on 400‐speed class of film screen combination at each hospital are given in Table IX. From Tables V to VIII, it can be seen that the scores rated good (i.e., all anatomical structures seen) were in percentage 39%, 31.2%, 25%, and 2.7% for chest PA, pelvis AP, lumbar spine AP, and lumbar spine LAT, respectively. On the other hand, the number of poor scores led in lumbar spine LAT projections (53%) followed by pelvis AP x‐ray examinations (28%). For lumbar spine AP and chest PA x‐ray examinations, the poor scores were 21% and 20%, respectively. Despite poor performance in lumbar spine LAT and pelvis AP x‐ray projections, such situation did not necessarily constitute to waste film. The majority of radiographs rated satisfactory could still meet the clinical needs of the intended diagnosis. The low optimization levels of film processing conditions, different levels of experiences among the radiology staff, and variable x‐ray techniques form a possible explanation for such performance in lumbar spine LAT and pelvis AP projections. The data suggest that the image quality criteria for good pelvis AP and lumbar spine LAT x‐ray examinations were the most difficult to meet, although this observation is not reflected in the list of image criteria that could be seen with some difficulties (Table X). The difficulty is attributable to varying technical and clinical needs. The majority of observers noted the detailed nature of the quality criteria and were of opinion that the variations in x‐ray techniques and film processing conditions were the main source of significant numbers of radiographs rated satisfactory that otherwise could be good radiographs. The observers also noted the difficulty of seeing image criteria as a function of increasing film contrast requirements.
Table IX

Typical film optical density ranges of optimized radiographs on 400‐speed class of film‐screen combination.

X‐ray examinationHospitalFilm density rangeMean film density a
Chest PAMMC0.35–2.14 1.1±0.41
Chest PAKCMC0.25–1.40 0.7±0.2
Chest PABMC0.51–1.80 1.62±0.16
Chest PAMCH0.28–1.32 0.76±0.35
Pelvis APMMC1.67–2.33 1.9±0.25
Pelvis APKCMC1.0–2.4 1.4±0.3
Pelvis APBMC0.22–2.36 1.21±0.56
Pelvis APMCH0.46–1.86 1.15±0.60
Lumbar spine APMMC1.02–1.75 1.35±0.28
Lumbar spine APKCMC0.8–1.6 1.2±0.1
Lumbar spine APBMC0.47–1.78 0.98±0.43
Lumbar spine APMCH0.40–1.24 0.82±0.29
Lumbar spine LATMMC1.25–2.33 1.81±0.53
Lumbar spine LATKCMC0.7–2.5 1.2±0.4
Lumbar spine LATMCH0.46–1.98 1.06±0.5

At one standard deviation.

Table X

Specific image criteria, which were generally difficult to see.

X‐ray examinationImage criteriaReference hospital(s)Number of radiographs
Chest PA‐symmetrical reproduction of the thorax‐KCMC4
‐reproduction of the whole rib cage above the diaphragm‐KCMC, MCH, ARH8
‐Visualization of the retrocardiac lung and the mediastinum‐MMC, KCMC6
‐Visually sharp reproduction of
(a) the trachea and proximal bronchi, the borders of the heart and aorta‐MMC, ARH3
(b) the diaphragm and casto‐phrenic angles‐MCH, ARH8
Lumbar spine AP‐Visually sharp reproduction of the Pedicles‐MMC, KCMC5
‐Reproduction of the spinious and and transverse processes‐MMC, KCMC8
Reproduction of the adjacent soft tissues particularly the psoas shadows‐KCMC, ARH6
Lumbar spine LAT‐Reproduction by tangential projection of the inferior end plate of L5 and the superior end plate of S1‐MMC3
‐Visualization of the anterior border of the upper sacrum‐MMC5
‐Reproduction of vertebral pieces of the upper sacrum‐MMC5
Pelvis AP‐Symmetrical reproduction of the pelvis‐MMC, KCMC6
‐visualization of the sacrum and its intervertebral foramina‐MMC, KCMC, ARH10
‐Visualization of the sacroiliac joints‐ARH3
‐Reproduction of spongiosa and and corticalis and visualization of the trochanters‐MCH4
Compliance with European guidelines for 50 chest PA x‐ray examinations. ‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear Compliance with European guidelines for 50 lumbar spine AP x‐ray examinations. ‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear Compliance with European guidelines for 50 lumbar spine LAT x‐ray examinations. ‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear Compliance with European guidelines for 50 pelvis AP x‐ray examinations. ‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear Typical film optical density ranges of optimized radiographs on 400‐speed class of film‐screen combination. At one standard deviation. Specific image criteria, which were generally difficult to see. The two observations above suggest that the application of the quality criteria can be prohibitive in lower healthcare levels (e.g., district hospitals, health centers, dispensaries, etc.) where radiographs reading and optimized film processing conditions are normally less satisfactory. However, some of these “difficult criteria” were not a prerequisite under the type of projections requested. For example, “visualization of the retro cardiac lung and the mediastinum” could better be seen in lateral view of chest than in PA view (reference ARH). Likewise a “visual sharp reproduction of the pedicles of the pelvis, particularly the psoas shadows” needs a special view of lumbar spine LAT for better visualization than the ordinary lateral view, which was of interest (reference KCMC). Such two different experiences show the need to adopt the European guidelines after careful examination of different specific clinical situations. The opinion has also been expressed by the publisher of the European guidelines, who clearly stated that the quality criteria define a level of performance considered necessary to produce images of standard quality for a particular anatomical region, without regard to specified clinical indication. Despite that, the overall results show that the majority of chest PA, lumbar spine AP, and pelvis AP radiographs passed the quality criteria by more than 70%, while the performance of lumbar spine LAT x‐ray examinations was slightly less than 50% and therefore less satisfactory. Generally, the compliance to European guidelines compare well with a similar study done in some African countries. The film densitometry results achieved in this study (Table IX) are also similar to the suggested density range values elsewhere. Excluding the data on lumbar spine LAT x‐ray examinations, the results (Tables V, VI, and VIII) indicate improvement on the performance of the European criteria from up to 50% of the radiographs rated poor earlier to 28% of poor radiographs observed during this study. The performance can implicitly be associated to the reduction of the number of retake films; the observation was later confirmed by the radiology staff.
Table VI

Compliance with European guidelines for 50 lumbar spine AP x‐ray examinations.

Image criteriaNumber of scores per score category a
Good Satisfactory Poor
‐Linear reproduction of the upper and lower plate surfaces in the centered beam area and visualization of the intervertebral spaces17330
‐Visually sharp reproduction of the pedicles03416
‐Visualization of the intervertebral joints26240
‐Reproduction of the spinuos and transverse process231017
‐Visually sharp reproduction of the cortex and trabecular structures04010
‐Reproduction of the adjacent soft tissues, particularly the psoas shadows82220

‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear

CONCLUSION

The performance of European guidelines on quality criteria for radiographic images in Tanzania has been presented. The usefulness of the criteria in optimization studies of x‐ray examinations has been demonstrated. The compliance of majority radiographs (obtained after implementation of dose reduction measures) to the guidelines is evidence that the radiation protection of patients was sufficiently optimized. The minimization of patient doses implies that the radiation risk associated with the x‐ray examinations was also minimized following the reduction in the number of waste film and hence retake films. Provided that the radiology staffs are familiar with the criteria and local clinical requirements are met, the European guidelines on quality criteria for radiographic images can practically be adopted in Tanzania. It is recommended that this study be extended to wider scale for reliable contribution to the medical regulations of ionizing radiation in pipeline. However, it is evident that relevant education and training programs must be initiated to the radiology staff for best results.

ACKNOWLEDGMENTS

The work was supported by a grant from the International Atomic Energy (Contract No. 9336/RO). The authors are further grateful to the International Center for Theoretical Physics for providing the opportunity to one of them (W.M.) to undertake a scientific study visit related to the work in Trieste, Italy. Last, but not least, the authors are especially thankful to Dr. H. Dieffenthal (KCMC), Dr. J. J Lyimo (Arusha x‐ray center), Dr. L. Iseme, and Dr. Z. Nhanga (MCH) for evaluating the quality of radiographic images and valuable discussions.
Table VII

Compliance with European guidelines for 50 lumbar spine LAT x‐ray examinations.

Image criteriaNumber of scores per score category a
Good Satisfactory Poor
‐Reproduction by tangential projection of the inferior end plate of L5 and the superior end of S142026
‐Visualization of the anterior border of the upper scrum03614
‐Reproduction of vertebral pieces of the upper scrum01040

‐good: feature detected and fully reproduced, detail visible and clearly defined ‐satisfactory: feature just visible, detail just visible but not clearly defined ‐poor: feature invisible, detail invisible or not clear

  1 in total

1.  The potential for reduction of radiation doses to patients undergoing some common X ray examinations in Tanzania.

Authors:  W E Muhogora; A M Nyanda
Journal:  Radiat Prot Dosimetry       Date:  2001       Impact factor: 0.972

  1 in total
  5 in total

1.  The use of reference image criteria in X-ray diagnostics: an application for the optimisation of lumbar spine radiographs.

Authors:  A Almén; A Tingberg; J Besjakov; S Mattsson
Journal:  Eur Radiol       Date:  2004-04-01       Impact factor: 5.315

2.  European Guidelines for AP/PA chest X-rays: routinely satisfiable in a paediatric radiology division?

Authors:  Sebastian Tschauner; Robert Marterer; Michael Gübitz; Peter I Kalmar; Emina Talakic; Sabine Weissensteiner; Erich Sorantin
Journal:  Eur Radiol       Date:  2015-05-24       Impact factor: 5.315

3.  Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations.

Authors:  Gaetano Compagnone; Laura Pagan; Carlo Bergamini
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

4.  Interventional Radiologists: a Necessary Evaluation of Technical, Protective and Technological Operation.

Authors:  Sayed Ali Rahimi; Maryam Pourkaveh
Journal:  Acta Inform Med       Date:  2016-02-02

5.  Diagnostic X-ray facilities as per quality control performances in Tanzania.

Authors:  Yesaya Yohana Sungita
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.