Reliable tidal volume estimates at the airway opening with an infant monitor during high-frequency oscillatory ventilation.

P Scalfaro1, J J Pillow, P D Sly, J Cotting.   

Abstract

OBJECTIVE: To assess the suitability of a hot-wire anemometer infant monitoring system (Florian, Acutronic Medical Systems AG, Hirzel, Switzerland) for measuring flow and tidal volume (Vt) proximal to the endotracheal tube during high-frequency oscillatory ventilation.
DESIGN: In vitro model study.
SETTING: Respiratory research laboratory. SUBJECT: In vitro lung model simulating moderate to severe respiratory distress. INTERVENTION: The lung model was ventilated with a SensorMedics 3100A ventilator. Vt was recorded from the monitor display (Vt-disp) and compared with the gold standard (Vt-adiab), which was calculated using the adiabatic gas equation from pressure changes inside the model.
MEASUREMENTS AND MAIN RESULTS: A range of Vt (1-10 mL), frequencies (5-15 Hz), pressure amplitudes (10-90 cm H2O), inspiratory times (30% to 50%), and Fio2 (0.21-1.0) was used. Accuracy was determined by using modified Bland-Altman plots (95% limits of agreement). An exponential decrease in Vt was observed with increasing oscillatory frequency. Mean DeltaVt-disp was 0.6 mL (limits of agreement, -1.0 to 2.1) with a linear frequency dependence. Mean DeltaVt-disp was -0.2 mL (limits of agreement, -0.5 to 0.1) with increasing pressure amplitude and -0.2 mL (limits of agreement, -0.3 to -0.1) with increasing inspiratory time. Humidity and heating did not affect error, whereas increasing Fio2 from 0.21 to 1.0 increased mean error by 6.3% (+/-2.5%).
CONCLUSIONS: The Florian infant hot-wire flowmeter and monitoring system provides reliable measurements of Vt at the airway opening during high-frequency oscillatory ventilation when employed at frequencies of 8-13 Hz. The bedside application could improve monitoring of patients receiving high-frequency oscillatory ventilation, favor a better understanding of the physiologic consequences of different high-frequency oscillatory ventilation strategies, and therefore optimize treatment.

Entities:  

Mesh:

Year:  2001        PMID: 11588453     DOI: 10.1097/00003246-200110000-00013

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  11 in total

1.  Accuracy of the volume and pressure displays of high frequency oscillators.

Authors:  J A Leipälä; S Iwasaki; A Milner; A Greenough
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-03       Impact factor: 5.747

2.  Noninvasive high frequency oscillatory ventilation through nasal prongs: bench evaluation of efficacy and mechanics.

Authors:  Daniele De Luca; Virgilio P Carnielli; Giorgio Conti; Marco Piastra
Journal:  Intensive Care Med       Date:  2010-09-21       Impact factor: 17.440

3.  Breath-to-breath analysis of abdominal and rib cage motion in surfactant-depleted piglets during high-frequency oscillatory ventilation.

Authors:  Dick G Markhorst; Jos R C Jansen; Adrianus J van Vught; Huibert R van Genderingen
Journal:  Intensive Care Med       Date:  2005-01-20       Impact factor: 17.440

4.  Volume-targeted modes of modern neonatal ventilators: how stable is the delivered tidal volume?

Authors:  Thomas Jaecklin; Denis R Morel; Peter C Rimensberger
Journal:  Intensive Care Med       Date:  2006-11-22       Impact factor: 17.440

5.  [Monitoring tidal volumes when using the Ventrain® emergency ventilator].

Authors:  A R Schmidt; K Ruetzler; T Haas; A Schmitz; M Weiss
Journal:  Anaesthesist       Date:  2016-05-31       Impact factor: 1.041

6.  Comparison of pneumotachography and anemometery for flow measurement during mechanical ventilation with volatile anesthetics.

Authors:  Jarred R Mondoñedo; Jacob Herrmann; John S McNeil; David W Kaczka
Journal:  J Clin Monit Comput       Date:  2016-11-14       Impact factor: 2.502

7.  Effect of the I/E ratio on CO2 removal during high-frequency oscillatory ventilation with volume guarantee in a neonatal animal model of RDS.

Authors:  Manuel Sánchez-Luna; Noelia González-Pacheco; Martín Santos; Ángel Blanco; Cristina Orden; Jaques Belik; Francisco J Tendillo
Journal:  Eur J Pediatr       Date:  2016-09-05       Impact factor: 3.183

8.  Using very high frequencies with very low lung volumes during high-frequency oscillatory ventilation to protect the immature lung. A pilot study.

Authors:  N González-Pacheco; M Sánchez-Luna; C Ramos-Navarro; N Navarro-Patiño; A R-S de la Blanca
Journal:  J Perinatol       Date:  2016-01-07       Impact factor: 2.521

9.  Secretory phospholipase A2 pathway in various types of lung injury in neonates and infants: a multicentre translational study.

Authors:  Daniele De Luca; Ettore Capoluongo; Vincent Rigo
Journal:  BMC Pediatr       Date:  2011-11-08       Impact factor: 2.125

10.  Functional lung imaging during HFV in preterm rabbits.

Authors:  Jordan Thurgood; Stuart Hooper; Melissa Siew; Megan Wallace; Stephen Dubsky; Marcus Kitchen; R Aidan Jamison; Richard Carnibella; Andreas Fouras
Journal:  PLoS One       Date:  2012-10-30       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.