Accuracy of the volume and pressure displays of high frequency oscillators.

J A Leipälä1, S Iwasaki, A Milner, A Greenough.   

Abstract

OBJECTIVE: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators.
METHODS: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5-15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator's patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz.
RESULTS: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators.
CONCLUSION: Frequency affects the accuracy of displayed volumes and, to a lesser extent, displayed pressures of high frequency oscillators. The results emphasise that data displayed by new devices should not be uncritically accepted.

Entities:  

Mesh:

Year:  2004        PMID: 14977906      PMCID: PMC1756021          DOI: 10.1136/adc.2002.023937

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  13 in total

1.  In vitro performance characteristics of high-frequency oscillatory ventilators.

Authors:  J J Pillow; M H Wilkinson; H L Neil; C A Ramsden
Journal:  Am J Respir Crit Care Med       Date:  2001-09-15       Impact factor: 21.405

2.  Dependence of intrapulmonary pressure amplitudes on respiratory mechanics during high-frequency oscillatory ventilation in preterm lambs.

Authors:  J Jane Pillow; Peter D Sly; Zoltan Hantos; Jason H T Bates
Journal:  Pediatr Res       Date:  2002-10       Impact factor: 3.756

3.  Measurement of tidal lung volumes in neonates during high-frequency oscillation.

Authors:  E W Hoskyns; A D Milner; I E Hopkin
Journal:  J Biomed Eng       Date:  1992-01

4.  Factors influencing mechanical performance of neonatal high-frequency ventilators.

Authors:  J J Fredberg; G M Glass; B R Boynton; I D Frantz
Journal:  J Appl Physiol (1985)       Date:  1987-06

5.  The effect of frequency and mean airway pressure on volume delivery during high-frequency oscillation.

Authors:  V Chan; A Greenough; A D Milner
Journal:  Pediatr Pulmonol       Date:  1993-03

6.  Effect of I/E ratio on mean alveolar pressure during high-frequency oscillatory ventilation.

Authors:  J J Pillow; H Neil; M H Wilkinson; C A Ramsden
Journal:  J Appl Physiol (1985)       Date:  1999-07

7.  Reliable tidal volume estimates at the airway opening with an infant monitor during high-frequency oscillatory ventilation.

Authors:  P Scalfaro; J J Pillow; P D Sly; J Cotting
Journal:  Crit Care Med       Date:  2001-10       Impact factor: 7.598

8.  CO2 elimination as a function of frequency and tidal volume in rabbits during HFO.

Authors:  J W Watson; A C Jackson
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1984-08

9.  High-frequency oscillatory ventilation for the prevention of chronic lung disease of prematurity.

Authors:  Alice H Johnson; Janet L Peacock; Anne Greenough; Neil Marlow; Elizabeth S Limb; Louise Marston; Sandra A Calvert
Journal:  N Engl J Med       Date:  2002-08-29       Impact factor: 91.245

10.  Measurement of tidal volume using a pneumotachometer during high-frequency oscillation.

Authors:  S E Courtney; K R Weber; W A Spohn; S W Malin; C V Bender; R W Gotshall
Journal:  Crit Care Med       Date:  1990-06       Impact factor: 7.598

View more
  3 in total

Review 1.  What is new in ventilation strategies for the neonate?

Authors:  Anne Greenough; Atul Sharma
Journal:  Eur J Pediatr       Date:  2007-06-02       Impact factor: 3.183

Review 2.  Trends in mechanical ventilation: are we ventilating our patients in the best possible way?

Authors:  Raffaele L Dellaca'; Chiara Veneroni; Ramon Farre'
Journal:  Breathe (Sheff)       Date:  2017-06

3.  High-frequency oscillatory ventilation combined with volume guarantee in a neonatal animal model of respiratory distress syndrome.

Authors:  Manuel Sánchez Luna; Martín Santos González; Francisco Tendillo Cortijo
Journal:  Crit Care Res Pract       Date:  2013-07-18