Literature DB >> 25270081

[Damage to implants due to high-frequency electrocautery : analysis of four fractured hip endoprostheses shafts].

C Konrads1, M N Wente, W Plitz, M Rudert, M Hoberg.   

Abstract

In revision surgery of joints, high-frequency electrocauterization instruments are used for homeostasis and dissection of soft tissue. If there is contact of these instruments with the metal implants, flashover can occur. This can lead to thermal microstructural changes in the material and as a consequence may reduce the fatigue strength of the implant. Four cases of hip revision surgeries were analysed. In all cases flashovers occurred and secondarily, the titanium hip endoprosthesis stem broke in the neck section of the prosthesis. The conducted investigations showed that contact between the high-frequency instrument and the anterolateral aspect of the endoprosthesis neck had occurred. Electrothermal implant damage was found in the broken area. If in hip revision surgery the stem is not to be replaced, contact between high-frequency instruments and the metal implant should be avoided.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25270081     DOI: 10.1007/s00132-014-3023-6

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  6 in total

1.  Prosthetic component failures in hip arthroplasty surgery.

Authors:  D A Heck; C M Partridge; J D Reuben; W L Lanzer; C G Lewis; E M Keating
Journal:  J Arthroplasty       Date:  1995-10       Impact factor: 4.757

2.  [Damage to a hip endoprosthesis caused by high-frequency electrocautery].

Authors:  G Huber; T Weik; M M Morlock
Journal:  Orthopade       Date:  2009-07       Impact factor: 1.087

Review 3.  The incidence of implant fractures after total hip arthroplasty.

Authors:  Patrick Sadoghi; Wolfram Pawelka; Michael C Liebensteiner; Alexandra Williams; Andreas Leithner; Gerold Labek
Journal:  Int Orthop       Date:  2014-01       Impact factor: 3.075

4.  Fracture of femoral prostheses in total hip replacement. A clinical study.

Authors:  J Charnley
Journal:  Clin Orthop Relat Res       Date:  1975-09       Impact factor: 4.176

5.  [On the problem of femoral stem failure in hip replacement endoprostheses (author's transl)].

Authors:  M Ungethüm
Journal:  Z Orthop Ihre Grenzgeb       Date:  1980-04

6.  Modular titanium alloy neck adapter failures in hip replacement--failure mode analysis and influence of implant material.

Authors:  Thomas M Grupp; Thomas Weik; Wilhelm Bloemer; Hanns-Peter Knaebel
Journal:  BMC Musculoskelet Disord       Date:  2010-01-04       Impact factor: 2.362

  6 in total
  2 in total

Review 1.  Metallic Implants Used in Lumbar Interbody Fusion.

Authors:  Jakub Litak; Michał Szymoniuk; Wojciech Czyżewski; Zofia Hoffman; Joanna Litak; Leon Sakwa; Piotr Kamieniak
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

2.  Femoral Stem Fracture in Hip Revision Arthroplasty: A Systematic Literature Review of the Real-World Evidence.

Authors:  Matthias Sukopp; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Z Orthop Unfall       Date:  2021-04-13       Impact factor: 0.923

  2 in total