The role of transesophageal echocardiography in transcatheter closure of secundum atrial septal defects by the Amplatzer septal occluder.

U Mazic1, P Gavora, J Masura.   

Abstract

BACKGROUND: Our purpose was to determine the role of transesophageal echocardiography (TEE) in the closure of atrial septal defects by the Amplatzer septal occluder (ASO) (AGA Medical, Golden Valley, Minn).
METHODS: A total of 240 patients with atrial septal defect (ASD) secundum were examined by transthoracic 2-dimensional echocardiography (TTE) and TEE to determine the ASD morphologic features, diameter, and rims. During transcatheter closure TEE was used for determination of the ASD diameter and guidance of the ASO implantation.
RESULTS: Sixteen (6%) patients were found not suitable for transcatheter closure with TTE, 35 (14%) with TEE, and 2 during catheterization. Twenty-eight patients (18%) had partial or total deficiency of the posterior, inferoanterior, or inferoposterior rim, 54 (27%) had a centrally positioned ASD, 92 (46%) had insufficient superoanterior rim, and 9 had multiple ASDs, whereas 8 had a septal aneurysm associated with a single defect and 4 a multiperforated aneurysm. A total of 170 patients underwent implantation of ASO. The ASO was correctly positioned in 144 at the first attempt. In the remainder TEE revealed unstable position of the left atrial disk (12), opening of both atrial disks in the left atrium (5), deployment of the device through the smaller defect in patients with multiple ASDs (3), and, in 1 patient, the device was too small and had to be replaced by a larger one.
CONCLUSIONS: Morphologic variations of the ASD are common. TEE is crucial for the determination of the ASD morphologic features, diameter, and rims, which are crucial for proper patient selection. TEE allows precise guiding and positioning of the ASO, which is essential for safe and effective transcatheter ASD closure.

Entities:  

Mesh:

Year:  2001        PMID: 11526362     DOI: 10.1067/mhj.2001.116770

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  16 in total

Review 1.  The role of cardiac ultrasound in stem cell therapy.

Authors:  Andrew M Kahn; Anthony N Demaria
Journal:  J Cardiovasc Transl Res       Date:  2009-01-22       Impact factor: 4.132

2.  Transthoracic echocardiographic guidance during transcatheter closure of atrial septal defects in children and adults.

Authors:  Abdullah Erdem; Turkay Sarıtas; Cenap Zeybek; Ilker Kemal Yucel; Nurdan Erol; Halil Demır; Ender Odemıs; Yalim Yalcın; Ahmet Celebı
Journal:  Int J Cardiovasc Imaging       Date:  2011-08-11       Impact factor: 2.357

Review 3.  Echocardiographic guidance of catheter-based treatments of atrial septal defect: transesophageal echocardiography remains the gold standard.

Authors:  C S Kleinman
Journal:  Pediatr Cardiol       Date:  2005 Mar-Apr       Impact factor: 1.655

4.  A noninvasive sizing method to choose fitted Amplatzer septal occluder by transthoracic echocardiography in patients with secundum atrial septal defects.

Authors:  Kuang-Jen Chien; Cheng-Liang Lee; Ta-Cheng Huang; Chu-Chuan Lin; Ken-Pen Weng; Shih-Hui Huang; Kai-Sheng Hsieh
Journal:  Heart Vessels       Date:  2007-07-20       Impact factor: 2.037

5.  Atrial septal defects type II: noninvasive evaluation of patients before implantation of an Amplatzer Septal Occluder and on follow-up by magnetic resonance imaging compared with TEE and invasive measurement.

Authors:  Christoph Weber; Michael Weber; Okan Ekinci; Thomas Neumann; Anja Deetjen; Andreas Rolf; Gerhard Adam; Christian W Hamm; Thorsten Dill
Journal:  Eur Radiol       Date:  2008-06-21       Impact factor: 5.315

6.  Clinical advantage of real-time three-dimensional transesophageal echocardiography for transcatheter closure of multiple atrial septal defects.

Authors:  Manabu Taniguchi; Teiji Akagi; Yasufumi Kijima; Shunji Sano
Journal:  Int J Cardiovasc Imaging       Date:  2013-04-23       Impact factor: 2.357

Review 7.  Challenges encountered during closure of atrial septal defects.

Authors:  W Knirsch; A Dodge-Khatami; E Valsangiacomo-Buechel; M Weiss; F Berger
Journal:  Pediatr Cardiol       Date:  2005 Mar-Apr       Impact factor: 1.655

8.  Predictors of successful transcatheter closure of atrial septal defect by cardiac magnetic resonance imaging.

Authors:  K Durongpisitkul; N L Tang; J Soongswang; D Laohaprasitiporn; A Nanal
Journal:  Pediatr Cardiol       Date:  2003-12-15       Impact factor: 1.655

9.  Transcatheter perimembranous ventricular septal defect closure under transthoracic echocardiographic guidance without fluoroscopy.

Authors:  Shouzheng Wang; Wenbin Ouyang; Yao Liu; Fengwen Zhang; Gaili Guo; Guangzhi Zhao; Xiangbin Pan
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

10.  Transcatheter Device Closure of Secundum Atrial Septal Defect in Adult Patient.

Authors:  Nabil Naser; Nura Hadziomerovic; Sevleta Avdic
Journal:  Acta Inform Med       Date:  2021-03