E Yakoumakis1, I A Tsalafoutas, D Nikolaou, I Nazos, E Koulentianos, C Proukakis.
Abstract
In this study, measurements of dose-area product (DAP) and entrance surface dose (ESD) were carried out in a sample of 25 adult patients who underwent intravenous urography (IVU). These measured quantities were used to estimate the effective dose E from the IVU examination, a quantity closely correlated to radiation risk. Estimating E involves the use of conversion coefficients that have been determined for specific X-ray views in a mathematical phantom. These are obtained under conditions which are not usually met in clinical practice. As a result, the E estimates using the two different measurable quantities can be quite different. Analysis of the calculation procedure suggests that the E estimate using the DAP measurements, in addition to being more practical, could be more accurate than using ESD measurements, as DAP is sensitive to the X-ray field size settings. Furthermore, it is shown that in the absence of the appropriate equipment, a reliable E estimate can be obtained from the ESD calculated using the exposure data for each X-ray view.Entities:
Mesh:
Year: 2001 PMID: 11511498 DOI: 10.1259/bjr.74.884.740727