Monte Carlo evaluation of object shape effects in iodine-131 SPET tumor activity quantification.

Y K Dewaraja1, M Ljungberg, K F Koral.   

Abstract

In our clinical iodine-131 single-photon emission tomography (SPET) quantification for radioimmunotherapy, calibration and partial volume correction are based on measurements with phantoms containing spheres to simulate patient tumors even though real tumors are frequently nonspherical. In this study, Monte Carlo simulation was used to evaluate how object shape influences "spill-out" and "spill-in", which are major sources of quantification error associated with the poor spatial resolution of 131I SPET. Objects that varied in shape (spheres, cylinders, and an irregular structure) but were identical in activity and volume were simulated. Iterative reconstruction employed both attenuation and triple-energy-window scatter compensation. VOIs were defined in the reconstructed images both using physical boundaries and using expanded boundaries to allow for the limited resolution. When physical boundaries were used, both spill-out and spill-in were more significant for nonspherical structures than for spherical structures. Over the range of object volumes (50-200 ml) and at all background levels, VOI counts in cylinders were lower than VOI counts in spheres. This underestimation increased with decrease in object size (for the cold background -18% at 200 ml and -39% at 50 ml). It also decreased with increase in background activity because spill-in partially compensated for spill-out. It was shown that with a VOI larger than physical size, the results are independent of object shape and size only in the case of cold background. Activity quantification was carried out using a procedure similar to that used in our clinic. Quantification of nonspherical objects was improved by simple sphere-based partial volume correction, but the error was still large in some cases (for example, -39% for a 50-ml cylinder in a cold background and -35% for a 200-ml irregular structure defined on the basis of a typical tumor outlined on an X-ray computed tomography scan of a patient with non-Hodgkin's lymphoma). Partial volume correction by patient-specific Monte Carlo simulation may provide better quantification accuracy.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11504087      PMCID: PMC2831207          DOI: 10.1007/s002590100551

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  16 in total

1.  Recovery correction for quantitation in emission tomography: a feasibility study.

Authors:  L Geworski; B O Knoop; M L de Cabrejas; W H Knapp; D L Munz
Journal:  Eur J Nucl Med       Date:  2000-02

2.  A parallel Monte Carlo code for planar and SPECT imaging: implementation, verification and applications in (131)I SPECT.

Authors:  Yuni K Dewaraja; Michael Ljungberg; Amitava Majumdar; Abhijit Bose; Kenneth F Koral
Journal:  Comput Methods Programs Biomed       Date:  2002-02       Impact factor: 5.428

3.  Initial results for Hybrid SPECT--conjugate-view tumor dosimetry in 131I-anti-B1 antibody therapy of previously untreated patients with lymphoma.

Authors:  K F Koral; Y Dewaraja; J Li; C L Barrett; D D Regan; K R Zasadny; S G Rommelfanger; I R Francis; M S Kaminski; R L Wahl
Journal:  J Nucl Med       Date:  2000-09       Impact factor: 10.057

4.  Development and validation of a Monte Carlo simulation of photon transport in an Anger camera.

Authors:  D J De Vries; S C Moore; R E Zimmerman; S P Mueller; B Friedland; R C Lanza
Journal:  IEEE Trans Med Imaging       Date:  1990       Impact factor: 10.048

5.  Quantitative imaging of iodine-131 distributions in brain tumors with pinhole SPECT: a phantom study.

Authors:  M F Smith; D R Gilland; R E Coleman; R J Jaszczak
Journal:  J Nucl Med       Date:  1998-05       Impact factor: 10.057

6.  Computerized three-dimensional segmented human anatomy.

Authors:  I G Zubal; C R Harrell; E O Smith; Z Rattner; G Gindi; P B Hoffer
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

7.  Single-photon emission tomographic quantification in spherical objects: effects of object size and background.

Authors:  F Zito; M C Gilardi; P Magnani; F Fazio
Journal:  Eur J Nucl Med       Date:  1996-03

8.  Neuroblastoma imaging using a combined CT scanner-scintillation camera and 131I-MIBG.

Authors:  H R Tang; A J Da Silva; K K Matthay; D C Price; J P Huberty; R A Hawkins; B H Hasegawa
Journal:  J Nucl Med       Date:  2001-02       Impact factor: 10.057

9.  Accuracy of 131I tumor quantification in radioimmunotherapy using SPECT imaging with an ultra-high-energy collimator: Monte Carlo study.

Authors:  Y K Dewaraja; M Ljungberg; K F Koral
Journal:  J Nucl Med       Date:  2000-10       Impact factor: 10.057

10.  Quantitation in positron emission computed tomography: 5. Physical--anatomical effects.

Authors:  J C Mazziotta; M E Phelps; D Plummer; D E Kuhl
Journal:  J Comput Assist Tomogr       Date:  1981-10       Impact factor: 1.826

View more
  8 in total

1.  A 3-dimensional absorbed dose calculation method based on quantitative SPECT for radionuclide therapy: evaluation for (131)I using monte carlo simulation.

Authors:  Michael Ljungberg; Katarina Sjögreen; Xiaowei Liu; Eric Frey; Yuni Dewaraja; Sven-Erik Strand
Journal:  J Nucl Med       Date:  2002-08       Impact factor: 10.057

2.  Comparison of residence time estimation methods for radioimmunotherapy dosimetry and treatment planning--Monte Carlo simulation studies.

Authors:  B He; R L Wahl; Y Du; G Sgouros; H Jacene; I Flinn; E C Frey
Journal:  IEEE Trans Med Imaging       Date:  2008-04       Impact factor: 10.048

Review 3.  Accuracy and precision of radioactivity quantification in nuclear medicine images.

Authors:  Eric C Frey; John L Humm; Michael Ljungberg
Journal:  Semin Nucl Med       Date:  2012-05       Impact factor: 4.446

4.  High long-term test-retest reliability for extrastriatal 11C-raclopride binding in healthy older adults.

Authors:  Nina Karalija; Lars Jonassson; Jarkko Johansson; Goran Papenberg; Alireza Salami; Micael Andersson; Katrine Riklund; Lars Nyberg; Carl-Johan Boraxbekk
Journal:  J Cereb Blood Flow Metab       Date:  2019-09-10       Impact factor: 6.200

5.  FDG PET/CT in Crohn's disease: correlation of quantitative FDG PET/CT parameters with clinical and endoscopic surrogate markers of disease activity.

Authors:  Babak Saboury; Ali Salavati; Alex Brothers; Sandip Basu; Thomas C Kwee; Marnix G E H Lam; Roland Hustinx; Edouard Louis; Drew A Torigian; Abass Alavi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-11-20       Impact factor: 9.236

6.  Is absolute quantification of dopaminergic neurotransmission studies with 123I SPECT ready for clinical use?

Authors:  Habib Zaidi; Georges El Fakhri
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-07       Impact factor: 9.236

7.  Tumor necrosis factor-related apoptosis-inducing ligand additive with Iodine-131 of inhibits non-small cell lung cancer cells through promoting apoptosis.

Authors:  Ning Yang; Shuzhan Yao; Dong Liu
Journal:  Oncol Lett       Date:  2018-05-04       Impact factor: 2.967

8.  Validation of Monte Carlo 131 I radiopharmaceutical dosimetry workflow using a 3D-printed anthropomorphic head and neck phantom.

Authors:  David P Adam; Joseph J Grudzinski; Ian Bormett; Benjamin L Cox; Ian R Marsh; Tyler J Bradshaw; Paul M Harari; Bryan P Bednarz
Journal:  Med Phys       Date:  2022-06-06       Impact factor: 4.506

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.