Literature DB >> 11493849

Evaluation of various evoked potential techniques for spinal cord monitoring during scoliosis surgery.

K D Luk1, Y Hu, Y W Wong, K M Cheung.   

Abstract

STUDY
DESIGN: This prospective study compared the outcomes of different evoked potential (EP) techniques for intraoperative spinal cord monitoring.
OBJECTIVES: To evaluate the reliability of different EP techniques administered during scoliosis surgery. SUMMARY OF BACKGROUND DATA: A number of different methods of intraoperative spinal cord monitoring are available. Because each has its own advantages and limitations, multimodal spinal cord monitoring has been proposed to improve monitoring reliability.
MATERIALS AND METHODS: Cortical somatosensory-evoked potential (CSEP), cortical motor-evoked potential (CMEP), spinal somatosensory-evoked potential (SSEP), and spinal cord-evoked potential (SCEP) were applied simultaneously to 30 patients undergoing surgical correction for spinal deformity. The presence of the EP waveforms and their reproducibilities over separate tests were compared. In addition, the monitoring outcomes were evaluated with the clinical results.
RESULTS: Of the 30 patients, CSEP waveforms were successfully recorded in 28 cases (93%), SCEP in 25 cases (83%), CMEP in 24 cases (80%), and SSEP in 21 cases (70%). Latencies of each EP technique showed no significant variability. However, amplitudes showed significant differences between different techniques. SCEP and CMEP showed clearer waveforms of greater amplitude that could be detected faster than CSEP and SSEP waveforms. SCEP and SSEP waveforms were more easily influenced by the surgical procedure.
CONCLUSION: CSEP and CMEP are recommended for routine monitoring, so that both ascending and descending tracts are monitored. If adequate signals for either of these proposed monitoring methods cannot be easily obtained, SSEP can substitute for CSEP, whereas SCEP can substitute for CMEP.

Entities:  

Mesh:

Year:  2001        PMID: 11493849     DOI: 10.1097/00007632-200108150-00008

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  13 in total

1.  Intraoperative spinal cord and nerve root monitoring: a survey of Canadian spine surgeons.

Authors:  Lissa Peeling; Stephen Hentschel; Richard Fox; Hamilton Hall; Daryl R Fourney
Journal:  Can J Surg       Date:  2010-10       Impact factor: 2.089

Review 2.  Current opinions and recommendations on multimodal intraoperative monitoring during spine surgeries.

Authors:  Martin Sutter; Vedran Deletis; Jiri Dvorak; Andreas Eggspuehler; Dieter Grob; David Macdonald; Alfred Mueller; Francesco Sala; Tetsuya Tamaki
Journal:  Eur Spine J       Date:  2007-08-15       Impact factor: 3.134

3.  Characteristics of multi-channel Br(E)-MsEP waveforms for the lower extremity muscles in thoracic spine surgery: comparison based on preoperative motor status.

Authors:  Kazuyoshi Kobayashi; Kei Ando; Mikito Tsushima; Masaaki Machino; Kyotaro Ota; Masayoshi Morozumi; Satoshi Tanaka; Shunsuke Kanbara; Naoki Ishiguro; Shiro Imagama
Journal:  Eur Spine J       Date:  2018-11-15       Impact factor: 3.134

4.  Usefulness of multi-channels in intraoperative spinal cord monitoring: multi-center study by the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.

Authors:  Zenya Ito; Yukihiro Matsuyama; Kenichi Shinomiya; Muneharu Ando; Shigenori Kawabata; Tsukasa Kanchiku; Takanori Saito; Masato Takahashi; Shinichiro Taniguchi; Naoya Yamamoto; Kei Yamada; Kazunobu Kida; Yasushi Fujiwara; Sho Kobayashi; Kazuhiko Satomi; Toshikazu Tani
Journal:  Eur Spine J       Date:  2013-04-04       Impact factor: 3.134

5.  The prevention of neural complications in the surgical treatment of scoliosis: the role of the neurophysiological intraoperative monitoring.

Authors:  F Pastorelli; M Di Silvestre; R Plasmati; R Michelucci; T Greggi; A Morigi; M R Bacchin; S Bonarelli; A Cioni; F Vommaro; N Fini; F Lolli; P Parisini
Journal:  Eur Spine J       Date:  2011-03-18       Impact factor: 3.134

Review 6.  [Intraoperative electrophysiological monitoring with evoked potentials].

Authors:  R Nitzschke; N Hansen-Algenstaedt; J Regelsberger; A E Goetz; M S Goepfert
Journal:  Anaesthesist       Date:  2012-04       Impact factor: 1.041

7.  The diagnostic value of multimodal intraoperative monitoring (MIOM) during spine surgery: a prospective study of 1,017 patients.

Authors:  Martin Sutter; Andreas Eggspuehler; Dieter Grob; Dezsoe Jeszenszky; Arnaldo Benini; François Porchet; Alfred Mueller; Jiri Dvorak
Journal:  Eur Spine J       Date:  2007-07-31       Impact factor: 3.134

8.  Multimodal intraoperative monitoring during surgery of spinal deformities in 217 patients.

Authors:  Andreas Eggspuehler; Martin A Sutter; Dieter Grob; Dezsö Jeszenszky; Jiri Dvorak
Journal:  Eur Spine J       Date:  2007-07-14       Impact factor: 3.134

9.  Analysis of 1014 consecutive operative cases to determine the utility of intraoperative neurophysiological data.

Authors:  Namath Syed Hussain
Journal:  Asian J Neurosurg       Date:  2015 Jul-Sep

10.  The proximal thoracic curve in adolescent idiopathic scoliosis: surgical strategy and management outcomes.

Authors:  Tarek Anwar Elfiky; Dino Samartzis; Wai-Yuen Cheung; Yat-Wa Wong; Keith D K Luk; Kenneth M C Cheung
Journal:  Global Spine J       Date:  2011-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.