OBJECTIVE: The purpose of this study was to evaluate the diagnostic efficacy of Ektaspeed Plus film (EPF), a charge-coupled device (CCD), and photostimulable phosphor (PSP)-based digital images for detection of simulated periapical lesions. STUDY
DESIGN: Lesions were simulated in the periapical areas of 24 human mandibular sections invested in acrylic using burs of sizes #1, 2, 4, and 6 and imaged using EPF, CCD, and PSP sensors. Percent correct response scores, sensitivity, and specificity values were computed for all variables. Repeated-measures analysis of variance and post hoc testing were performed to determine the effects of imaging modality, observer, and lesion sizes with respect to lesion detection.
RESULTS: EPF displayed the highest sensitivity and specificity, followed by PSP and CCD images (P <.001). Percent correct score was the highest for 3 of 4 observers when EPF was used. Analysis of variance revealed significance (P <.001) with respect to all variables. Observers with experience in digital image-viewing performed better than those without such experience (P <.001). Intraobserver agreement was fair (kappa = 0.5).
CONCLUSIONS: EPF outperformed CCD and PSP images when observers could manipulate image characteristics.

Entities:  

Mesh:

Year:  2001        PMID: 11458252     DOI: 10.1067/moe.2001.115974

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  8 in total

1.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

2.  Cone Beam CT use in the pre-prosthetic evaluation of endodontically treated of the rear maxilla.

Authors:  M Gargari; L Ottria; M Nezzo; L Neroni; E Fanucci
Journal:  Oral Implantol (Rome)       Date:  2012-11-16

3.  Different digital imaging techniques in dental practice.

Authors:  Amela Lacević; Edina Vranić
Journal:  Bosn J Basic Med Sci       Date:  2004-05       Impact factor: 3.363

4.  Comparison of Cone-beam Computed Tomography and Periapical Radiography to Determine the Proximity of Periapical Lesions to Anatomical Structures in Premaxillary Area prior to Surgical Endodontics: A Clinical Study.

Authors:  Kesha Sheth; Sonali Kapoor; Shilpi Daveshwar
Journal:  Int J Clin Pediatr Dent       Date:  2020 Jul-Aug

5.  A comparison of the diagnostic accuracy of in vivo and in vitro photostimulable phosphor digital images in the detection of occlusal caries lesions.

Authors:  K Kamburoglu; B Senel; S P Yüksel; T Ozen
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

6.  A comparative study on image quality of two digital intraoral sensors.

Authors:  Cinar Aziman; Kristina Hellén-Halme; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2019-05-14       Impact factor: 2.419

7.  Diagnostic Accuracy of Cone-Beam Computed Tomography and Periapical Radiography in Internal Root Resorption.

Authors:  Zahrasadat Madani; Ehsan Moudi; Ali Bijani; Elham Mahmoudi
Journal:  Iran Endod J       Date:  2015-12-24

8.  Accuracy of Cone Beam Computed Tomography, Photostimulable Phosphor Plate Digital Radiography and Conventional Radiography for Detection of Artificial Cancellous Bone Defects.

Authors:  Roghieh Bardal; Hassan Jahanihashemi; Maryam Mostafavi; Esmaeil Kalhor; Maryam Tofangchiha; Mahdieh Dehghani
Journal:  J Dent (Tehran)       Date:  2015-11
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.