Literature DB >> 11448320

Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph.

L M Zangwill1, C Bowd, C C Berry, J Williams, E Z Blumenthal, C A Sánchez-Galeana, C Vasile, R N Weinreb.   

Abstract

OBJECTIVE: To compare the ability of 3 instruments, the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT), to discriminate between healthy eyes and eyes with early to moderate glaucomatous visual field loss. SUBJECTS AND METHODS: Forty-one patients with early to moderate glaucomatous visual field loss and 50 healthy subjects were included in the study. The HRT, GDx, and OCT imaging and visual field testing were completed on 1 eye from each subject within a 6-month interval. Statistical differences in sensitivity at fixed specificities of 85%, 90%, and 95% were evaluated. In addition, areas under the receiver operating characteristic (ROC) curve were compared.
RESULTS: No significant differences were found between the area under the ROC curve and the best parameter from each instrument: OCT thickness at the 5-o'clock inferior temporal position (mean +/- SE, 0.87 +/- 0.04), HRT mean height contour in the nasal inferior region (mean +/- SE, 0.86 +/- 0.04), and GDx linear discriminant function (mean +/- SE, 0.84 +/- 0.04). Twelve HRT, 2 GDx, and 9 OCT parameters had an area under the ROC curve of at least 0.81. At a fixed specificity of 90%, significant differences were found between the sensitivity of OCT thickness at the 5-o'clock inferior temporal position (71%) and parameters with sensitivities less than 52%. Qualitative assessment of stereophotographs resulted in a sensitivity of 80%.
CONCLUSION: Although the area under the ROC curves was similar among the best parameters from each instrument, qualitative assessment of stereophotographs and measurements from the OCT and HRT generally had higher sensitivities than measurements from the GDx.

Entities:  

Mesh:

Year:  2001        PMID: 11448320     DOI: 10.1001/archopht.119.7.985

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  70 in total

1.  Acutance, an objective measure of retinal nerve fibre image clarity.

Authors:  Y F Choong; F Rakebrandt; R V North; J E Morgan
Journal:  Br J Ophthalmol       Date:  2003-03       Impact factor: 4.638

2.  Glaucomatous progression in series of stereoscopic photographs and Heidelberg retina tomograph images.

Authors:  Neil O'Leary; David P Crabb; Steven L Mansberger; Brad Fortune; Michael D Twa; Michael J Lloyd; Aachal Kotecha; David F Garway-Heath; George A Cioffi; Chris A Johnson
Journal:  Arch Ophthalmol       Date:  2010-05

3.  African Descent and Glaucoma Evaluation Study (ADAGES): II. Ancestry differences in optic disc, retinal nerve fiber layer, and macular structure in healthy subjects.

Authors:  Christopher A Girkin; Pamela A Sample; Jeffrey M Liebmann; Sonia Jain; Christopher Bowd; Lida M Becerra; Felipe A Medeiros; Lyne Racette; Keri A Dirkes; Robert N Weinreb; Linda M Zangwill
Journal:  Arch Ophthalmol       Date:  2010-05

4.  Ability of cirrus HD-OCT optic nerve head parameters to discriminate normal from glaucomatous eyes.

Authors:  Jean-Claude Mwanza; Jonathan D Oakley; Donald L Budenz; Douglas R Anderson
Journal:  Ophthalmology       Date:  2010-10-28       Impact factor: 12.079

5.  Comparison of algorithms for detection of localised nerve fibre layer defects using scanning laser polarimetry.

Authors:  F A Medeiros; R Susanna
Journal:  Br J Ophthalmol       Date:  2003-04       Impact factor: 4.638

6.  A few remarks about glaucoma.

Authors:  A Wegner
Journal:  Br J Ophthalmol       Date:  2002-08       Impact factor: 4.638

Review 7.  Imaging in glaucoma.

Authors:  Daniel M Stein; Gadi Wollstein; Joel S Schuman
Journal:  Ophthalmol Clin North Am       Date:  2004-03

8.  Novel use of 3T MRI in assessment of optic nerve volume in glaucoma.

Authors:  Norlina M Ramli; Sabrilhakim Sidek; Fadzlina A Rahman; Mohammadreza Peyman; Mimiwati Zahari; Kartini Rahmat; Norlisah Ramli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-04-27       Impact factor: 3.117

9.  Diagnostic Performance of the ISNT Rule for Glaucoma Based on the Heidelberg Retinal Tomograph.

Authors:  Errol Wei'en Chan; Jiemin Liao; Reuben Chao Ming Foo; Seng Chee Loon; Tin Aung; Tien Yin Wong; Ching-Yu Cheng
Journal:  Transl Vis Sci Technol       Date:  2013-06-28       Impact factor: 3.283

10.  Scanning laser polarimetry with variable corneal compensation and detection of glaucomatous optic neuropathy.

Authors:  Stefano Da Pozzo; Pierluigi Iacono; Roberta Marchesan; Anna Fantin; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-09       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.