Regional improvement of signal-to-noise and contrast-to-noise ratios in dual-screen CR chest imaging--a phantom study.

X Liu1, C C Shaw.   

Abstract

The improvement of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in dual-screen computed radiography (CR) has been investigated for various regions in images of an anthropomorphic chest phantom. With the dual-screen CR technique, two image plates are placed in a cassette and exposed together during imaging. The exposed plates are separately scanned to form a front image and a back image, which are then registered and superimposed to form a composite image with improved SNRs and CNRs. The improvement can be optimized by applying specifically selected weighting factors during superimposition. In this study, dual-screen CR images of an anthropomorphic chest phantom were acquired and formed with four different combinations of standard resolution (ST) and high-resolution (HR) screens: ST-ST, ST-HR, HR-ST, and HR-HR. SNRs and their improvements were measured and compared over twelve representative regions-of-interest (ROIs) in these images. A 19.1%-45.7% increase of the SNR was observed, depending on the ROI and screen combination used. The optimal weighting factors were found to vary by only 4.5%-12.4%. Largest improvement was found in the lung field for all screen combinations. Improvement of CNRs was investigated over two ROIs in the lung field using the rib bones as the contrast objects and a 29.2%-43.9% improvement of the CNR was observed. Among the four screen combinations, ST-ST resulted in the most SNR and CNR improvement, followed in order by HR-ST, HR-HR, and ST-HR. The HR-ST combination yielded the lowest spatial variation of the optimal weighting factors with improved SNRs and CNRs close to those of the ST-ST combination.

Entities:  

Mesh:

Year:  2001        PMID: 11439477     DOI: 10.1118/1.1376136

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Noise simulation in cone beam CT imaging with parallel computing.

Authors:  Shu-Ju Tu; Chris C Shaw; Lingyun Chen
Journal:  Phys Med Biol       Date:  2006-02-15       Impact factor: 3.609

2.  Dependence of radiographic sensitivity of CR imaging plate on X-ray tube voltage.

Authors:  Yoshiyuki Asai; Masanobu Uemura; Masao Matsumoto; Hitoshi Kanamori
Journal:  Radiol Phys Technol       Date:  2007-12-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.