Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults.

W B White1, Y A Anwar.   

Abstract

BACKGROUND: Non-invasive self blood pressure monitoring has become increasingly popular. To assure the accuracy of devices used for this purpose, all need to be validated independently prior to marketing. The objective of this study was to assess the accuracy of the HEM-907, a new semi-automatic, non-invasive, oscillometric blood pressure monitoring device specifically designed to be used in the clinic or physician's office setting.
METHODS: Blood pressure measurements taken employing this device were compared with the results obtained by two experienced observers using a mercury sphygmomanometer on 100 subjects and patients (384 measurements). The limits of agreement were calculated for the device compared with the results of the two observers according to the standards of the Association for the Advancement of Medical Instrumentation (AAMI).
RESULTS: The agreement between the two observers was -0.36+/-2.32mmHg for systolic blood pressure and 0.02+/-2.42mmHg for diastolic blood pressure. The agreement between the device and the observers was 1.56+/-4.42mmHg and 3.49+/-4.61mmHg for systolic and diastolic blood pressure respectively.
CONCLUSIONS: The Omron HEM-907 satisfied the AAMI criteria for accuracy for a non-invasive blood pressure monitoring device.

Entities:  

Mesh:

Year:  2001        PMID: 11433132     DOI: 10.1097/00126097-200104000-00007

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  63 in total

1.  Why use automated office blood pressure measurements in clinical practice?

Authors:  Emmanuel A Andreadis; Epameinondas T Angelopoulos; Gerasimos D Agaliotis; Athanasios P Tsakanikas; George P Mousoulis
Journal:  High Blood Press Cardiovasc Prev       Date:  2011-09-01

2.  The effects of pre-ejection period on post-exercise systolic blood pressure estimation using the pulse arrival time technique.

Authors:  Mico Yee Man Wong; Emma Pickwell-MacPherson; Yuan Ting Zhang; Jack C Y Cheng
Journal:  Eur J Appl Physiol       Date:  2010-09-08       Impact factor: 3.078

3.  Evaluation of a mindfulness-based intervention program to decrease blood pressure in low-income African-American older adults.

Authors:  Priya Palta; G Page; R L Piferi; J M Gill; M J Hayat; A B Connolly; S L Szanton
Journal:  J Urban Health       Date:  2012-04       Impact factor: 3.671

4.  The acute effects of running on blood pressure estimation using pulse transit time in normotensive subjects.

Authors:  Mico Yee-Man Wong; Emma Pickwell-MacPherson; Yuan-Ting Zhang
Journal:  Eur J Appl Physiol       Date:  2009-06-20       Impact factor: 3.078

Review 5.  Recent advances in automated blood pressure measurement.

Authors:  Martin G Myers
Journal:  Curr Hypertens Rep       Date:  2008-10       Impact factor: 5.369

Review 6.  Automated Office-Based Blood Pressure Measurement: an Overview and Guidance for Implementation in Primary Care.

Authors:  Romsai T Boonyasai; Erika L McCannon; Joseph E Landavaso
Journal:  Curr Hypertens Rep       Date:  2019-04-04       Impact factor: 5.369

Review 7.  Automated office blood pressure measurement in primary care.

Authors:  Martin G Myers; Janusz Kaczorowski; Martin Dawes; Marshall Godwin
Journal:  Can Fam Physician       Date:  2014-02       Impact factor: 3.275

8.  Physical activity patterns of American Indian and Alaskan Native people living in Alaska and the Southwestern United States.

Authors:  Diana Redwood; Mary C Schumacher; Anne P Lanier; Elizabeth D Ferucci; Elvin Asay; Laurie J Helzer; Lillian Tom-Orme; Sandra L Edwards; Maureen A Murtaugh; Martha L Slattery
Journal:  Am J Health Promot       Date:  2009 Jul-Aug

9.  Development, implementation, and evaluation of a computerized self-administered diet history questionnaire for use in studies of American Indian and Alaskan native people.

Authors:  Martha L Slattery; Maureen A Murtaugh; Mary Catherine Schumacher; Jennifer Johnson; Sandra Edwards; Roger Edwards; Joan Benson; Lillian Tom-Orme; Anne P Lanier
Journal:  J Am Diet Assoc       Date:  2008-01

Review 10.  Determining which automatic digital blood pressure device performs adequately: a systematic review.

Authors:  Y Wan; C Heneghan; R Stevens; R J McManus; A Ward; R Perera; M Thompson; L Tarassenko; D Mant
Journal:  J Hum Hypertens       Date:  2010-04-08       Impact factor: 3.012

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.