Determination of the accuracy of implant reconstruction and dose delivery in brachytherapy in The Netherlands and Belgium.

R J Elfrink1, I K Kolkman-Deurloo, H J van Kleffens, A Rijnders, B Schaeken, T H Aalbers, W J Dries, J L Venselaar.   

Abstract

PURPOSE: To gain insight into the accuracy of brachytherapy treatments, the accuracy of implant reconstruction and dose delivery was investigated in 33 radiotherapy institutions in The Netherlands and Belgium.
MATERIALS AND METHODS: The accuracy of the implant reconstruction method was determined using a cubic phantom containing 25 spheres at well-known positions. Reconstruction measurements were obtained on 41 brachytherapy localizers, 33 of which were simulators. The reconstructed distances between the spheres were compared with the true distances. The accuracy of the dose delivery was determined for high dose rate (HDR), pulsed dose rate (PDR) and low dose rate (LDR) afterloading systems using a polymethyl methacrylate cylindrical phantom containing a NE 2571 ionization chamber in its centre. The institutions were asked to deliver a prescribed dose at the centre of the phantom. The measured dose was compared with the prescribed dose.
RESULTS: The average reconstruction accuracy was -0.07 mm (+/-0.4 mm, 1 SD) for 41 localizers. The average deviation of the measured dose from the prescribed dose was +0.9% (+/-1.3%, 1 SD) for 21 HDR afterloading systems, +1.0% (+/-2.3%, 1 SD) for 12 PDR afterloaders, and +1.8% (+/-2.5%, 1 SD) for 15 LDR afterloaders.
CONCLUSIONS: This comparison showed a good accuracy of brachytherapy implant reconstruction and dose delivery in The Netherlands and Belgium.

Entities:  

Mesh:

Year:  2001        PMID: 11369071     DOI: 10.1016/s0167-8140(01)00300-0

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  6 in total

Review 1.  Dosimetric audit in brachytherapy.

Authors:  A L Palmer; D A Bradley; A Nisbet
Journal:  Br J Radiol       Date:  2014-05-06       Impact factor: 3.039

2.  Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.

Authors:  Catharine H Clark; Edwin G A Aird; Steve Bolton; Elizabeth A Miles; Andrew Nisbet; Julia A D Snaith; Russell A S Thomas; Karen Venables; David I Thwaites
Journal:  Br J Radiol       Date:  2015-09-02       Impact factor: 3.039

3.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30

Review 4.  Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.

Authors:  Christian Kirisits; Mark J Rivard; Dimos Baltas; Facundo Ballester; Marisol De Brabandere; Rob van der Laarse; Yury Niatsetski; Panagiotis Papagiannis; Taran Paulsen Hellebust; Jose Perez-Calatayud; Kari Tanderup; Jack L M Venselaar; Frank-André Siebert
Journal:  Radiother Oncol       Date:  2013-11-30       Impact factor: 6.280

5.  Prostate volume and implant configuration during 48 hours of temporary prostate brachytherapy: limited effect of oedema.

Authors:  Anna M Dinkla; Bradley R Pieters; Kees Koedooder; Niek van Wieringen; Rob van der Laarse; Arjan Bel
Journal:  Radiat Oncol       Date:  2014-12-11       Impact factor: 3.481

6.  Independent assessment of source position for gynecological applicator in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Nakamura; Shie Nishioka; Kotaro Iijima; Akihisa Wakita; Yukinao Abe; Naoki Tohyama; Shinji Kawamura; Toshiyuki Minemura; Jun Itami
Journal:  J Contemp Brachytherapy       Date:  2017-10-19
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.