| Literature DB >> 11308201 |
R M Davila1, T Rawles, M J Mack.
Abstract
Minimal access techniques with cardiopulmonary bypass use smaller cannula systems for management of cardiopulmonary bypass. To augment flow rates through the smaller cannula, the technique of vacuum-assisted venous drainage has been used. We describe a complication of vacuum-assisted venous drainage by inadvertent positive pressurization of the venous circuit resulting in a paradoxic air embolus across a patent atrial septal defect. Hazards of the current cardiopulmonary bypass systems and techniques for avoiding this potential complication are discussed.Entities:
Mesh:
Year: 2001 PMID: 11308201 DOI: 10.1016/s0003-4975(00)02198-6
Source DB: PubMed Journal: Ann Thorac Surg ISSN: 0003-4975 Impact factor: 4.330