Comparison of the CardioWest total artificial heart, the novacor left ventricular assist system and the thoratec ventricular assist system in bridge to transplantation.

J G Copeland1, R G Smith, F A Arabia, P E Nolan, V K Mehta, M S McCarthy, K A Chisholm.   

Abstract

BACKGROUND: Device selection has historically been supported by minimal comparative data. Since 1994, we have implanted 43 patients with the CardioWest Total Artificial Heart (CW), 23 with the Novacor Left Ventricular Assist System (N), and 26 with the Thoratec Ventricular Assist System (T). This experience provides a basis for our device selection criteria.
METHODS: We reviewed retrospectively the results for survival, stroke, and infection in the CW, N, and T groups. Statistical methods included the Student's t-test, chi2 analysis, and Kaplan-Meier actuarial survival curves.
RESULTS: The T group patients were younger and smaller sized than the CW or N group. The CW group had the highest mean central venous pressure (CVP) and lowest mean cardiac index. Survival to transplantation was 75% for CW, 57% for N, and 38% for T. Multiple organ failure postimplant caused most deaths in the CW and T groups. Right heart failure and stroke caused most N deaths. Linearized stroke rates (event/patient-month) were 0.03 for CW, 0.28 for N, and 0.08 for T. Serious infections were found in 20% of CW, 30% of N, and 8% of T patients, but linearized rates showed little difference and death from infection was rare.
CONCLUSIONS: The N device should be used in "stable" patients with body surface area (BSA) greater than 1.7 m2 and with minimal right heart failure. Unstable patients with biventricular failure should receive a CW if the BSA is greater than 1.7 m2 or a T if they are smaller.

Entities:  

Mesh:

Year:  2001        PMID: 11265873     DOI: 10.1016/s0003-4975(00)02625-4

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  12 in total

Review 1.  Mechanical circulatory support.

Authors:  Deborah J Kozik; Mark D Plunkett
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

2.  The helical flow pump with a hydrodynamic levitation impeller.

Authors:  Yusuke Abe; Kohei Ishii; Takashi Isoyama; Itsuro Saito; Yusuke Inoue; Toshiya Ono; Hidemoto Nakagawa; Emiko Nakano; Kyoko Fukazawa; Kazuhiko Ishihara; Kazuyoshi Fukunaga; Minoru Ono; Kou Imachi
Journal:  J Artif Organs       Date:  2012-08-28       Impact factor: 1.731

Review 3.  A ventricular assist device as a bridge to recovery, decision making, or transplantation in patients with advanced cardiac failure.

Authors:  Siyamek Neragi-Miandoab
Journal:  Surg Today       Date:  2012-07-20       Impact factor: 2.549

Review 4.  An insight into short- and long-term mechanical circulatory support systems.

Authors:  Markus Ferrari; Peter Kruzliak; Kyriakos Spiliopoulos
Journal:  Clin Res Cardiol       Date:  2014-10-28       Impact factor: 5.460

5.  Cardiac Replacement in Our Lifetime.

Authors:  Jack G Copeland
Journal:  Tex Heart Inst J       Date:  2019-02-01

6.  Left Ventricular Assist Devices for the Treatment of Congestive Heart Failure.

Authors:  Michel Haddad; Khanh Lam; Paul Hendry; Thierry Mesana; Richard Davies
Journal:  Curr Treat Options Cardiovasc Med       Date:  2005-05

7.  End-stage heart failure with multiple intracardiac thrombi: a rescue strategy.

Authors:  Maximillian G Posch; Larry O Thompson; Michael M Koerner; Mehmet H Akay; George P Noon; Matthias Loebe
Journal:  Tex Heart Inst J       Date:  2004

Review 8.  The total artificial heart.

Authors:  Jason A Cook; Keyur B Shah; Mohammed A Quader; Richard H Cooke; Vigneshwar Kasirajan; Kris K Rao; Melissa C Smallfield; Inna Tchoukina; Daniel G Tang
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

9.  Low incidence of neurologic events during long-term support with the HeartMate XVE left ventricular assist device.

Authors:  Mark S Slaughter; Michael A Sobieski; Colleen Gallagher; Muhyaldeen Dia; Marc A Silver
Journal:  Tex Heart Inst J       Date:  2008

10.  First Turkish experience with the MicroMed DeBakey VAD.

Authors:  Deniz Suha Kucukaksu; Erol Sener; Akif Undar; George P Noon; Oguz Tasdemir
Journal:  Tex Heart Inst J       Date:  2003