Accuracy of plastic replica of aortic aneurysm using 3D-CT data for transluminal stent-grafting: experimental and clinical evaluation.

K Kato1, T Ishiguchi, K Maruyama, S Naganawa, T Ishigaki.   

Abstract

PURPOSE: To plan stent-grafting for aortic aneurysms with complicated morphology, we prepared life-sized aortic replicas by laser stereolithography using helical 3D--CT data. The accuracy of the replica was evaluated by measurement of vessel phantoms and clinical 3D--CT data.
METHOD: An imaginary aortic wall was created from helical CT images of the aorta, and a hollow plastic replica was produced by laser stereolithography. The accuracy of the replica was evaluated in five abdominal aortic aneurysms by experimental phantom studies and measurements of the replicas.
RESULTS: The mean difference in measurements between 3D--CT images and model vessels and between 3D--CT images and aortic replicas was 0.2 mm each. Therefore, the difference in measurements between real aortic aneurysms and the replicas was at most 0.4 mm.
CONCLUSION: The accuracy of the replica is satisfactory, making it useful for preoperative evaluation and simulation for stent-grafting.

Entities:  

Mesh:

Year:  2001        PMID: 11242232     DOI: 10.1097/00004728-200103000-00026

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  5 in total

1.  Rapid vessel prototyping: vascular modeling using 3t magnetic resonance angiography and rapid prototyping technology.

Authors:  Michael Markl; Ralf Schumacher; Jürg Küffer; Thorsten A Bley; Jürgen Hennig
Journal:  MAGMA       Date:  2005-12-21       Impact factor: 2.310

Review 2.  3D Bioprinting of cardiac tissue and cardiac stem cell therapy.

Authors:  Matthew Alonzo; Shweta AnilKumar; Brian Roman; Nishat Tasnim; Binata Joddar
Journal:  Transl Res       Date:  2019-04-20       Impact factor: 7.012

Review 3.  Cardiac 3D Printing and its Future Directions.

Authors:  Marija Vukicevic; Bobak Mosadegh; James K Min; Stephen H Little
Journal:  JACC Cardiovasc Imaging       Date:  2017-02

4.  Advanced "tactile" medical imaging for separation surgeries of conjoined twins.

Authors:  Andrew M Christensen; Stephen M Humphries; Keith Y C Goh; Dale Swift
Journal:  Childs Nerv Syst       Date:  2004-07-22       Impact factor: 1.475

5.  The accuracy and reliability of 3D printed aortic templates: a comprehensive three-dimensional analysis.

Authors:  Pawel Rynio; Maciej Wojtuń; Łukasz Wójcik; Miłosz Kawa; Aleksander Falkowski; Piotr Gutowski; Arkadiusz Kazimierczak
Journal:  Quant Imaging Med Surg       Date:  2022-02
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.