The influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics.

J J Liau1, C C Hu, C K Cheng, C H Huang, W H Lo.   

Abstract

OBJECTIVE: To investigate the influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics.
DESIGN: A finite element analysis was used to investigate the alteration of contact characteristics of the tibiofemoral joint due to inserting a pressure sensitive film between the contacting surfaces.
BACKGROUND: The discrepancy between actual contact behaviors of tibiofemoral joint of knee prosthesis and the measurement using Fuji pressure sensitive film was not discussed extensively. The change of direct contact circumstance of the tibiofemoral joint due to inserting a pressure sensitive film was not well reported.
METHOD: A two-dimensional finite element model of the tibiofemoral joint of knee prosthesis in the sagittal plane was constructed. Four different radii of the femoral component were used to investigate the conformity effect. Two-layers of plane strain element were used to model the pressure sensitive film. The influence of inserting a pressure sensitive film on the actual contact characteristics was analyzed by comparing the results of the change in contact radius.
RESULTS: Inserting a pressure sensitive film between contacting surfaces would disturb the original contact behaviors, especially in the lowest conformity design. The maximum difference of contact radius between the model simulating actual contact circumstance and the model with inserting a pressure sensitive film between contacting surface was 77% in the lowest conformity condition at the smallest load.
CONCLUSIONS: This study proposes a quantitative analysis of contact characteristics in the tibiofemoral joint of knee prosthesis between the models with and without inserting a pressure sensitive film into the contact surface. The measurement of contact area in artificial tibiofemoral joints by using Fuji pressure sensitive film is always overestimated its true contact area by 14-77%. RELEVANCE: This study revealed the measurement of contact characteristics of artificial tibiofemoral joints by using Fuji pressure sensitive film which depends on not only the applied load, but also the conformity and material properties of the contact surface. Therefore, the information of the conformity and material properties of the contact surface should be provided as well as the applied load wherever a measurement of contact characteristics using Fuji pressure sensitive film is analyzed and interpreted.

Mesh:

Year:  2001        PMID: 11222935     DOI: 10.1016/s0268-0033(00)00040-1

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  7 in total

Review 1.  Techniques of Force and Pressure Measurement in the Small Joints of the Wrist.

Authors:  Michael J Schreck; Meghan Kelly; Colin D Canham; John C Elfar
Journal:  Hand (N Y)       Date:  2017-02-06

2.  Intra-articular compressive stress of the elbow joint in extension: an experimental study using Fuji films.

Authors:  C Chantelot; G Wavreille; C Dos Remedios; B Landejerit; C Fontaine; H Hildebrand
Journal:  Surg Radiol Anat       Date:  2008-01-29       Impact factor: 1.246

3.  A nondestructive, reproducible method of measuring joint reaction force at the distal radioulnar joint.

Authors:  Colin D Canham; Michael J Schreck; Noorullah Maqsoodi; Madison Doolittle; Mark Olles; John C Elfar
Journal:  J Hand Surg Am       Date:  2015-04-16       Impact factor: 2.230

4.  Biomechanical evaluation of meniscal root repair: a porcine study.

Authors:  Ekkehard F Röpke; Sebastian Kopf; Steffen Drange; Roland Becker; Christoph H Lohmann; Christian Stärke
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-07-16       Impact factor: 4.342

5.  Distal Radioulnar Joint Reaction Force Following Ulnar Shortening: Diaphyseal Osteotomy Versus Wafer Resection.

Authors:  Colin D Canham; Michael J Schreck; Noorullah Maqsoodi; Susan Messing; Mark Olles; John C Elfar
Journal:  J Hand Surg Am       Date:  2015-10-06       Impact factor: 2.230

6.  The Effect of Axial Compression and Distraction on Cervical Facet Cartilage Apposition During Shear and Bending Motions.

Authors:  Ryan D Quarrington; Darcy W Thompson-Bagshaw; Claire F Jones
Journal:  Ann Biomed Eng       Date:  2022-03-07       Impact factor: 3.934

7.  Biomechanical Measurement Error Can Be Caused by Fujifilm Thickness: A Theoretical, Experimental, and Computational Analysis.

Authors:  Ahmed Sarwar; Simli Srivastava; Chris Chu; Alan Machin; Emil H Schemitsch; Habiba Bougherara; Zahra S Bagheri; Radovan Zdero
Journal:  Biomed Res Int       Date:  2017-08-16       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.