Literature DB >> 11215920

Stentless bioprostheses have ideal haemodynamics, even in the small aortic root.

L H Baur1, Y Houdas, K H Peels, J Braun, B van Straten, A Prat, A P Kappetein, M Wolters-Geldoff, E E van der Wall, A V Bruschke, H A Huysmans.   

Abstract

OBJECTIVE: To determine normal Doppler and 2D gradients and flow characteristics of the Freestyle stentless aortic bioprosthesis related to valve size.
BACKGROUND: The Freestyle stentless aortic bioprosthesis is one of the newer aortic xenografts. Only limited data are available of the echocardiographic flow characteristics during a mid-term follow-up period of this valve. Therefore valve performance related to valve size was measured during a follow-up period of two years.
METHODS: 175 consecutive patients with a Freestyle aortic bioprosthesis underwent an echocardiographic and Doppler examination according to a common protocol. Investigations were done within 4 weeks after operation, after 3 to 6 months, and after 1 and 2 years.
RESULTS: With a valve size from 19 to 27 mm mean gradients decreased from 8.0 +/- 5.1 mmHg at discharge to 5.8 +/- 3.8 mmHg after 3-6 months (p < 0.001). Thereafter gradients remained stable. The performance index, the ratio of the measured effective orifice area in the patient divided by the effective orifice area measured in vitro increased from 69 +/- 20% at discharge to 79 +/- 29% after one, two and three years. Performance index was especially very high in the smaller sized valves with a performance index of 85 +/- 17% for the 21 mm valve. During follow-up mean gradients remained below 10 mmHg even in the 21 mm valve.
CONCLUSION: Stentless xenografts have ideal haemodynamics, even in the small aortic root.

Entities:  

Mesh:

Year:  2000        PMID: 11215920     DOI: 10.1023/a:1026521211249

Source DB:  PubMed          Journal:  Int J Card Imaging        ISSN: 0167-9899


  27 in total

1.  The Ross operation in children: 10-year experience.

Authors:  R C Elkins; C J Knott-Craig; K E Ward; M M Lane
Journal:  Ann Thorac Surg       Date:  1998-02       Impact factor: 4.330

2.  Comparison of long-term results of Carpentier-Edwards and Hancock bioprosthetic valves.

Authors:  H Bolooki; G A Kaiser; S M Mallon; G M Palatianos
Journal:  Ann Thorac Surg       Date:  1986-11       Impact factor: 4.330

3.  Tissue buckling as a mechanism of bioprosthetic valve failure.

Authors:  I Vesely; D Boughner; T Song
Journal:  Ann Thorac Surg       Date:  1988-09       Impact factor: 4.330

4.  Echocardiographic profile of the normally functioning Omnicarbon valve.

Authors:  W Fehske; D Kessel; P G Kirchhoff; H Omran; M Manz; B Lüderitz
Journal:  J Heart Valve Dis       Date:  1994-05

5.  Normal echocardiographic characteristics of the Sorin Bicarbon bileaflet prosthetic heart valve in the mitral and aortic positions.

Authors:  L Badano; R Mocchegiani; D Bertoli; G DeGaetano; L Carratino; L Pasetti; M Caudullo; A Budini; B Mannello; G Passerone
Journal:  J Am Soc Echocardiogr       Date:  1997 Jul-Aug       Impact factor: 5.251

Review 6.  Thromboembolic complications of current cardiac valvular prostheses.

Authors:  L H Edmunds
Journal:  Ann Thorac Surg       Date:  1982-07       Impact factor: 4.330

7.  Evaluation of aortic insufficiency by Doppler color flow mapping.

Authors:  G J Perry; F Helmcke; N C Nanda; C Byard; B Soto
Journal:  J Am Coll Cardiol       Date:  1987-04       Impact factor: 24.094

8.  Doppler echocardiography in the assessment of the homograft aortic valve.

Authors:  W M Jaffe; H A Coverdale; A H Roche; P W Brandt; J A Ormiston; B G Barratt-Boyes
Journal:  Am J Cardiol       Date:  1989-06-15       Impact factor: 2.778

9.  Stentless porcine aortic root: valve of choice for the elderly patient with small aortic root?

Authors:  C F Sintek; A D Fletcher; S Khonsari
Journal:  J Thorac Cardiovasc Surg       Date:  1995-05       Impact factor: 5.209

10.  Mechanical stresses on closed cusps of porcine bioprosthetic valves: correlation with sites of calcification.

Authors:  H N Sabbah; M S Hamid; P D Stein
Journal:  Ann Thorac Surg       Date:  1986-07       Impact factor: 4.330

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.