Characteristics of Kodak Insight, an F-speed intraoral film.

J B Ludlow1, E Platin, A Mol.   

Abstract

OBJECTIVES: This study reports film speed, contrast, exposure latitude, resolution, and response to processing solution depletion of Kodak Insight intraoral film.
METHODS: Densitometric curves were generated by using International Standards Organization protocol. Additional curves were generated for Ultra-speed, Ektaspeed Plus, and Insight films developed in progressively depleted processing solutions. Eight observers viewed images of a resolution test tool for maximum resolution assessment. Images of an aluminum step-wedge were reviewed to determine useful exposure latitude.
RESULTS: Insight's sensitivity in fresh automatic processor solutions places it in the F-speed group. An average gradient of 1.8 was found with all film types. Insight provided 93% of the useful exposure latitude of Ektaspeed Plus film. Insight maintained contrast in progressively depleted processing solutions. Like Ektaspeed Plus, Insight was able to resolve at least 20 line-pairs per millimeter.
CONCLUSIONS: Under International Standards Organization conditions, Insight required only 77% of the exposure of Ektaspeed Plus film. Insight film provided stable contrast in depleted processing solutions.

Mesh:

Substances:

Year:  2001        PMID: 11174582     DOI: 10.1067/moe.2001.110575

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  3 in total

1.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 1: the influence of X-ray generator on periodontal measurements using conventional and digital receptors.

Authors:  Bart Vandenberghe; Livia Corpas; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-05       Impact factor: 3.573

2.  Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films.

Authors:  Mehmet Ilgüy; Semanur Dinçer; Dilhan Ilgüy; Gündüz Bayirli
Journal:  J Digit Imaging       Date:  2008-10-24       Impact factor: 4.056

3.  Effects of developer exhaustion on DFL Contrast FV-58 and Kodak Insight dental films.

Authors:  Fabiano Pachêco de Carvalho; M M F da Silveira; M A G Frazão; S T de Santana; M L dos Anjos Pontual
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419