Patient-derived quality specifications for instruments used in self-monitoring of blood glucose.

S Skeie1, G Thue, S Sandberg.   

Abstract

BACKGROUND: Instruments for self-monitoring of glucose (SMBG) are increasingly used by diabetic patients. Information is limited on how patients use and interpret SMBG results, and no quality specifications for such instruments are based on the opinions of patients.
METHODS: Type 1 diabetic patients (n = 201) filled in a questionnaire eliciting daily limits for blood glucose (BG) and changes of BG considered significant at different glucose concentrations. From these responses, patient-derived quality specifications were calculated in different clinical situations with low, intermediate, and high BG concentrations.
RESULTS: Mean age of the patients was 31.8 years, mean diabetes duration was 14.7 years, and mean SMBG duration was 10.0 years with a mean frequency of 11.2 measurements/week. The threshold for hypoglycemic symptoms was 3.0 mmol/L (54 mg/dL), and the mean daily BG target window was 4.3-10.4 mmol/L (77-187 mg/dL). The mean absolute BG changes producing actions from the patients ranged from 1.1 mmol/L (20 mg/dL) to 3.6 mmol/L (65 mg/dL). The analytical quality specifications for imprecision depended on the clinical situation. Excluding the hypoglycemic situation, the analytical CV needed to fulfill the expectations of 75% of the patients was 6.4-9.7%. The analytical quality specification for CV at hypoglycemic concentrations was 3.1%.
CONCLUSIONS: Instruments for self-measurements of glucose with an imprecision (CV) of < or = 5% and bias < or = 5% meet the expectations of >75% of patients in clinical situations other than hypoglycemia.

Entities:  

Mesh:

Year:  2001        PMID: 11148179

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  5 in total

1.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Diabetes Care       Date:  2011-06       Impact factor: 19.112

2.  Quality control of self-monitoring of blood glucose: why and how?

Authors:  Bogdan Solnica; Jerzy W Naskalski
Journal:  J Diabetes Sci Technol       Date:  2007-03

3.  Self-monitoring of blood glucose: the use of the first or the second drop of blood.

Authors:  Johanna Hortensius; Robbert J Slingerland; Nanne Kleefstra; Susan J J Logtenberg; Klaas H Groenier; Sebastiaan T Houweling; Henk J G Bilo
Journal:  Diabetes Care       Date:  2011-02-02       Impact factor: 19.112

Review 4.  Interpretation of EQA results and EQA-based trouble shooting.

Authors:  Gunn Berit Berge Kristensen; Piet Meijer
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

5.  Commutability of external quality assessment materials for point-of-care glucose testing using the Clinical and Laboratory Standards Institute and International Federation of Clinical Chemistry approaches.

Authors:  Yan Wang; Mario Plebani; Laura Sciacovelli; Shunli Zhang; Qingtao Wang; Rui Zhou
Journal:  J Clin Lab Anal       Date:  2020-04-27       Impact factor: 2.352

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.