Efficacy of biventricular sensing and treatment of ventricular arrhythmias.

M Bocchiardo1, M Achtelik, F Gaita, H J Trappe, I Lozano, S Higgins, E Daoud, P Yong, C Duby.   

Abstract

While much is known concerning the hemodynamic effects of biventricular (BV) pacing, little has been reported concerning the efficacy of BV sensing and pacing in the detection and treatment of ventricular tachyarrhythmias. Two hundred nineteen heart failure (HF) patients with VT or VF and a QRS > or = 120 ms during sinus rhythm received an ICD capable of BV pacing and sensing. Detection times of induced VF and success rates for terminating induced VT were measured. The ICD system used a left ventricular epicardial lead implanted via thoracotomy (52 patients) or a specially designed percutaneous, over-the-wire lead inserted in the coronary venous system. VF detection times and VT termination rates by antitachycardia pacing (ATP) were compared with those measured in a population of recipients of ICD using a RV lead alone. Median induced VF detection times were comparable (2.0-s BV vs 1.8-s RV). Termination of induced VT on the first attempt was comparable with BV pacing (87.4%) versus RV pacing (89.6%). The time to detect induced VF was not different with ICDs using BV sensing versus conventional ICDs using RV sensing alone. Similarly, the rates of successful termination of induced VT by ATP with BV or RV pacing were comparable.

Entities:  

Mesh:

Year:  2000        PMID: 11139974     DOI: 10.1111/j.1540-8159.2000.tb07069.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

Review 1.  Cardiac resynchronization: a brief synopsis part II: implant and followup methodology.

Authors:  David M Kalinchak; Mark H Schoenfeld
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

Review 2.  Cardiac resynchronization therapy: strategies for device programming, troubleshooting and follow-up.

Authors:  Safwat Gassis; Angel R León
Journal:  J Interv Card Electrophysiol       Date:  2005-09       Impact factor: 1.900

3.  Innovations in the development of cardiac resynchronization therapy devices with defibrillation capability-the Boston Scientific experience.

Authors:  Patrick Yong; Erdulfo Deperalta
Journal:  J Cardiovasc Transl Res       Date:  2008-09-09       Impact factor: 4.132

4.  Entrainment of ventricular tachycardia with a permanent biventricular pacemaker.

Authors:  Jerry C Luck; Luna Bhatta; Stephen E Artman; Deborah L Wolbrette; Mary Angela Pantelloni; Gerald V Naccarelli
Journal:  J Interv Card Electrophysiol       Date:  2002-08       Impact factor: 1.900

5.  Frequent ICD shocks due to double sensing in patients with bi-ventricular implantable cardioverter defibrillators.

Authors:  Amin Al-Ahmad; Paul J Wang; Munther K Homoud; N A Mark Estes; Mark S Link
Journal:  J Interv Card Electrophysiol       Date:  2003-12       Impact factor: 1.900