Evaluation of renal function with 99mTc-MAG3 using semiautomated regions of interest.

Y Inoue1, K Yoshikawa, N Yoshioka, T Watanabe, S Saegusa, Y Kaneko, I Yokoyama, K Ohtomo.   

Abstract

UNLABELLED: The need to draw regions of interest (ROIs) manually may reduce the convenience and reliability of estimating renal function from renal scintigraphy. We developed a semiautomated method to define ROIs for renal scintigraphy with 99mTc-mercaptoacetyltriglycine (MAG3) and evaluated the clinical applicability of the method to the estimation of renal function by camera-based methods.
METHODS: Dynamic renal scintigraphy with 99mTc-MAG3 was performed on 21 patients. An operator placed a large rectangular ROI over each kidney, a circular ROI within the liver, and a rectangular ROI between the kidneys. Using these ROIs, semiautomated renal ROIs were determined on the basis of the temporal changes in counts, in addition to the absolute counts, and a subrenal background ROI was automatically assigned for each renal ROI. Background-subtracted renograms were generated using these renal and subrenal ROIs, and renogram parameters were derived from the slope of the renogram and the area under the renogram. Clearance was calculated using the renogram parameters and equations determined previously with manual ROIs and correlated with clearance measured by a single-sample method. The relative function of the right kidney determined by the semiautomated method was compared with that determined by the manual method. Data processing was performed independently by another operator to assess interoperator reproducibility.
RESULTS: ROIs defined by the semiautomated method were visually judged to be acceptable for clinical use in all patients with a wide range of renal function. Clearance was successfully predicted with the semiautomated ROIs (r = 0.968 using the slope of the renogram; r = 0.934 using the area under the renogram), and relative function calculated with the semiautomated ROIs was almost identical to that calculated with manual ROIs. There was almost complete concordance in absolute and relative function between the two operators.
CONCLUSION: The semiautomated method can define ROIs for 99mTc-MAG3 renal scintigraphy with limited operator intervention. Camera-based methods using the semiautomated ROIs allow estimation of renal function with high accuracy and little interoperator variability and are suggested to be suitable for clinical use.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11138677

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  7 in total

1.  Fully automatic region of interest selection in glomerular filtration rate estimation from 99mTc-DTPA renogram.

Authors:  Kun-Ju Lin; Jia-Yann Huang; Yung-Sheng Chen
Journal:  J Digit Imaging       Date:  2011-12       Impact factor: 4.056

2.  Relative Ectopic Kidney Function Quantification Using DMSA Tomoscintigraphy Modality.

Authors:  Ines Njeh; Kalil Chtourou; Ahmed BenHamida
Journal:  J Healthc Eng       Date:  2020-05-26       Impact factor: 2.682

3.  Totally automatic definition of renal regions of interest from 99mTc-MAG3 renograms: validation in patients with normal kidneys and in patients with suspected renal obstruction.

Authors:  Ernest V Garcia; Russell Folks; Samuel Pak; Andrew Taylor
Journal:  Nucl Med Commun       Date:  2010-05       Impact factor: 1.690

4.  Reference Ranges in [(99m)Tc]Mercaptoacetyltriglycerine Renography: Comparison of a Semi-automated (Xeleris, GE) and Manual (Picker, Odyssey) Processing Software.

Authors:  Kate I Rewers; Svend Hvidsten; Oke Gerke; Søren Schifter
Journal:  Mol Imaging Biol       Date:  2015-10       Impact factor: 3.488

5.  Renal access through the inferior calyx is associated with higher risk of severe bleeding after percutaneous nephrolithotomy.

Authors:  Jing Tan; Binghai Chen; Leye He; Guangming Yin; Zhiqiang Jiang; Kun Yao; Xianzheng Jiang
Journal:  Arch Med Sci       Date:  2015-04-23       Impact factor: 3.318

6.  Evaluation of Congenital Hydronephrosis with Static and Dynamic Magnetic Resonance Urography in Comparation to Dynamic Renal Scintigraphy.

Authors:  Amra Dzananovic; Amela Begic; Danka Pokrajac
Journal:  Acta Inform Med       Date:  2019-09

7.  In Vitro and In Vivo (1)H-MR Spectroscopic Examination of the Renal Cell Carcinoma.

Authors:  F Süllentrop; J Hahn; D Moka
Journal:  Int J Biomed Sci       Date:  2012-06
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.