Literature DB >> 11098798

Quantification of fMRI artifact reduction by a novel plaster cast head holder.

V Edward1, C Windischberger, R Cunnington, M Erdler, R Lanzenberger, D Mayer, W Endl, R Beisteiner.   

Abstract

In light of artifact-induced high variability of activation in fMRI repeat studies, we developed and tested a clinically useful plaster cast head holder (PCH) with improved immobilization, repositioning, and comfort. With PCH, there were considerably lower levels of translational and rotational head motion components compared to head fixation with conventional restraining straps (CRS). Rotational components cannot be fully compensated by realignment and lead to "false activations." In addition, task-correlated head motion, which highly increases the risk of artifacts, was considerably reduced with PCH, especially in a motion prone subject. Compared with PCH, head motion was 133% larger with CRS in a highly cooperative subject. With a motion prone subject, head motion range was increased by 769% (PCH: 0.9 mm, CRS: 7.8 mm), which may indicate the usefulness of PCH for restless patients. In functional activation maps, PCH alone yielded fewer residual motion artifacts than CRS + image registration. Subject tolerance of the head holder during the long measurement times of up to 2.5 hr was good, and slice orientation on different days confirmed the quality of repositioning.

Entities:  

Mesh:

Year:  2000        PMID: 11098798      PMCID: PMC6871974          DOI: 10.1002/1097-0193(200011)11:3<207::aid-hbm60>3.0.co;2-j

Source DB:  PubMed          Journal:  Hum Brain Mapp        ISSN: 1065-9471            Impact factor:   5.038


  23 in total

1.  Automated image registration: I. General methods and intrasubject, intramodality validation.

Authors:  R P Woods; S T Grafton; C J Holmes; S R Cherry; J C Mazziotta
Journal:  J Comput Assist Tomogr       Date:  1998 Jan-Feb       Impact factor: 1.826

2.  Automated image registration: II. Intersubject validation of linear and nonlinear models.

Authors:  R P Woods; S T Grafton; J D Watson; N L Sicotte; J C Mazziotta
Journal:  J Comput Assist Tomogr       Date:  1998 Jan-Feb       Impact factor: 1.826

3.  An fMRI study of the human cortical motor system response to increasing functional demands.

Authors:  B E Wexler; R K Fulbright; C M Lacadie; P Skudlarski; M B Kelz; R T Constable; J C Gore
Journal:  Magn Reson Imaging       Date:  1997       Impact factor: 2.546

4.  Magnetoencephalography may help to improve functional MRI brain mapping.

Authors:  R Beisteiner; M Erdler; C Teichtmeister; M Diemling; E Moser; V Edward; L Deecke
Journal:  Eur J Neurosci       Date:  1997-05       Impact factor: 3.386

5.  Comparing localization of conventional functional magnetic resonance imaging and magnetoencephalography.

Authors:  R Beisteiner; G Gomiscek; M Erdler; C Teichtmeister; E Moser; L Deecke
Journal:  Eur J Neurosci       Date:  1995-05-01       Impact factor: 3.386

6.  A stereotactic method of anatomical localization for positron emission tomography.

Authors:  P T Fox; J S Perlmutter; M E Raichle
Journal:  J Comput Assist Tomogr       Date:  1985 Jan-Feb       Impact factor: 1.826

7.  A new headholder for PET, CT, and NMR imaging.

Authors:  K J Kearfott; D A Rottenberg; R J Knowles
Journal:  J Comput Assist Tomogr       Date:  1984-12       Impact factor: 1.826

8.  Head holder for PET, CT, and MR studies.

Authors:  V Bettinardi; R Scardaoni; M C Gilardi; G Rizzo; D Perani; E Paulesu; G Striano; F Triulzi; F Fazio
Journal:  J Comput Assist Tomogr       Date:  1991 Sep-Oct       Impact factor: 1.826

9.  A method of correlating and merging cerebral morphology and function by a special head holder.

Authors:  H J Kaiser; O Sabri; G Wagenknecht; B Lege; D Hellwig; U Büll
Journal:  Nuklearmedizin       Date:  1994-06       Impact factor: 1.379

10.  Functional analysis of human MT and related visual cortical areas using magnetic resonance imaging.

Authors:  R B Tootell; J B Reppas; K K Kwong; R Malach; R T Born; T J Brady; B R Rosen; J W Belliveau
Journal:  J Neurosci       Date:  1995-04       Impact factor: 6.167

View more
  25 in total

Review 1.  A review of the automated detection of change in serial imaging studies of the brain.

Authors:  Julia Patriarche; Bradley Erickson
Journal:  J Digit Imaging       Date:  2004-06-29       Impact factor: 4.056

Review 2.  [Methodological principles for optimising functional MR experiments].

Authors:  T Wüstenberg; F L Giesel; H Strasburger
Journal:  Radiologe       Date:  2005-02       Impact factor: 0.635

Review 3.  Pitfalls in FMRI.

Authors:  Sven Haller; Andreas J Bartsch
Journal:  Eur Radiol       Date:  2009-06-06       Impact factor: 5.315

Review 4.  Motion artifacts in MRI: A complex problem with many partial solutions.

Authors:  Maxim Zaitsev; Julian Maclaren; Michael Herbst
Journal:  J Magn Reson Imaging       Date:  2015-01-28       Impact factor: 4.813

5.  Presurgical motor, somatosensory and language fMRI: Technical feasibility and limitations in 491 patients over 13 years.

Authors:  Anthony J Tyndall; Julia Reinhardt; Volker Tronnier; Luigi Mariani; Christoph Stippich
Journal:  Eur Radiol       Date:  2016-05-19       Impact factor: 5.315

6.  Reducing Inter-Site Variability for Fluctuation Amplitude Metrics in Multisite Resting State BOLD-fMRI Data.

Authors:  Xinbo Wang; Qing Wang; Peiwen Zhang; Shufang Qian; Shiyu Liu; Dong-Qiang Liu
Journal:  Neuroinformatics       Date:  2021-01

7.  How much are clinical fMRI reports influenced by standard postprocessing methods? An investigation of normalization and region of interest effects in the medial temporal lobe.

Authors:  Roland Beisteiner; Nicolaus Klinger; Ilse Höllinger; Jakob Rath; Susanne Gruber; Thomas Steinkellner; Thomas Foki; Alexander Geissler
Journal:  Hum Brain Mapp       Date:  2010-03-04       Impact factor: 5.038

8.  Comparison of fMRI coregistration results between human experts and software solutions in patients and healthy subjects.

Authors:  Andreas Gartus; Alexander Geissler; Thomas Foki; Amir Reza Tahamtan; Gerald Pahs; Markus Barth; Katja Pinker; Siegfried Trattnig; Roland Beisteiner
Journal:  Eur Radiol       Date:  2006-10-12       Impact factor: 5.315

9.  Evaluation of preoperative high magnetic field motor functional MRI (3 Tesla) in glioma patients by navigated electrocortical stimulation and postoperative outcome.

Authors:  K Roessler; M Donat; R Lanzenberger; K Novak; A Geissler; A Gartus; A R Tahamtan; D Milakara; T Czech; M Barth; E Knosp; R Beisteiner
Journal:  J Neurol Neurosurg Psychiatry       Date:  2005-08       Impact factor: 10.154

10.  Standardizing the intrinsic brain: towards robust measurement of inter-individual variation in 1000 functional connectomes.

Authors:  Chao-Gan Yan; R Cameron Craddock; Xi-Nian Zuo; Yu-Feng Zang; Michael P Milham
Journal:  Neuroimage       Date:  2013-04-28       Impact factor: 6.556

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.