BACKGROUND: The number of patients with pacemakers has been increasing and a large number of them will present with chest pain or symptoms suggesting angina pectoris. Myocardial ischemia and presence of coronary artery disease are difficult to detect and assess by noninvasive methods in patients with a pacemaker; the electrocardiogram (ECG) at rest and during exercise is usually very difficult to analyze in terms of ischemia or even presence of an acute myocardial infarction. HYPOTHESIS: To detect significant coronary stenosis in patients with previously implanted pacemakers, we tested a new stress echocardiography method using incremental ventricular pacing by already implanted pacemakers.
METHODS: We studied prospectively 25 consecutive patients who underwent stress echocardiography with increasing ventricular pacing up to either 85% of the age-predicted maximal heart rate or chest pain. Positive tests were defined by new hypokinesia or worsening of a preexisting alteration in wall motion in at least two adjacent territories. All patients underwent coronary angiograms to define the presence and severity of coronary stenoses.
RESULTS: Among the 25 tests, 11 (44%) were stopped for chest pain. 1 (4%) for moderate discomfort, 1 (4%) for a drop in blood pressure, and the target pacing rate was achieved in the tests of the remaining 12 patients (48%). There were no complications. Thirteen patients had significant stenoses. In 10 cases, stress echocardiography was a true positive test with respect to coronary angiography. There were 11 true negative, 1 false positive, and 3 false negative tests. The sensitivity was 77%, specificity was 90%, the positive predictive value was 91%, and the negative predictive value 79%. The accuracy was 84%.
CONCLUSIONS: This new stress echocardiography method appears feasible, easy, safe, and effective for detection of significant coronary stenoses in patients with pacemakers.

Entities:  

Mesh:

Year:  2000        PMID: 11097132      PMCID: PMC6655253          DOI: 10.1002/clc.4960231111

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  18 in total

1.  Simultaneous transesophageal atrial pacing and transesophageal two-dimensional echocardiography: a new method of stress echocardiography.

Authors:  H Lambertz; A Kreis; H Trümper; P Hanrath
Journal:  J Am Coll Cardiol       Date:  1990-11       Impact factor: 24.094

2.  Simultaneous transesophageal two-dimensional echocardiography and atrial pacing for detecting coronary artery disease.

Authors:  O Kamp; C C De Cock; A J Küpper; J P Roos; C A Visser
Journal:  Am J Cardiol       Date:  1992-06-01       Impact factor: 2.778

Review 3.  Recommendations for quantitation of the left ventricle by two-dimensional echocardiography. American Society of Echocardiography Committee on Standards, Subcommittee on Quantitation of Two-Dimensional Echocardiograms.

Authors:  N B Schiller; P M Shah; M Crawford; A DeMaria; R Devereux; H Feigenbaum; H Gutgesell; N Reichek; D Sahn; I Schnittger
Journal:  J Am Soc Echocardiogr       Date:  1989 Sep-Oct       Impact factor: 5.251

4.  Electrocardiographic changes subsequent to artificial ventricular depolarization.

Authors:  K Chatterjee; A Harris; G Davies; A Leatham
Journal:  Br Heart J       Date:  1969-11

5.  Detection of coronary artery disease by two-dimensional echocardiography and transesophageal atrial pacing.

Authors:  S Iliceto; M Sorino; G D'Ambrosio; A Papa; S Favale; G Biasco; P Rizzon
Journal:  J Am Coll Cardiol       Date:  1985-05       Impact factor: 24.094

6.  Echocardiographic detection of coronary artery disease during dobutamine infusion.

Authors:  S G Sawada; D S Segar; T Ryan; S E Brown; A M Dohan; R Williams; N S Fineberg; W F Armstrong; H Feigenbaum
Journal:  Circulation       Date:  1991-05       Impact factor: 29.690

7.  Thallium-201 scintigraphy in complete left bundle branch block.

Authors:  H O Hirzel; M Senn; K Nuesch; C Buettner; A Pfeiffer; O M Hess; H P Krayenbuehl
Journal:  Am J Cardiol       Date:  1984-03-01       Impact factor: 2.778

8.  Regional myocardial perfusion and glucose metabolism in experimental left bundle branch block.

Authors:  S Ono; R Nohara; H Kambara; K Okuda; C Kawai
Journal:  Circulation       Date:  1992-03       Impact factor: 29.690

9.  Exercise echocardiography: a clinically practical addition in the evaluation of coronary artery disease.

Authors:  W S Robertson; H Feigenbaum; W F Armstrong; J C Dillon; J O'Donnell; P W McHenry
Journal:  J Am Coll Cardiol       Date:  1983-12       Impact factor: 24.094

10.  Stress echocardiography with transesophageal atrial pacing: preliminary report of a new method for detection of ischemic wall motion abnormalities.

Authors:  P D Chapman; T P Doyle; P J Troup; C M Gross; L S Wann
Journal:  Circulation       Date:  1984-09       Impact factor: 29.690

View more
  1 in total

Review 1.  Pacing stress echocardiography.

Authors:  Suzana Gligorova; Marco Agrusta
Journal:  Cardiovasc Ultrasound       Date:  2005-12-09       Impact factor: 2.062

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.