Subsidence resulting from simulated postoperative neck movements: an in vitro investigation with a new cervical fusion cage.

H J Wilke1, A Kettler, C Goetz, L Claes.   

Abstract

STUDY
DESIGN: A biomechanical in vitro subsidence test of different cervical interbody fusion devices was performed using a new testing protocol that simulates physiologic conditions.
OBJECTIVES: To investigate the effect of simulated postoperative neck movements on the subsidence of the new WING cervical interbody fusion cage in comparison with two other cages and bone cement. SUMMARY OF BACKGROUND DATA: Cervical interbody fusion cages sometimes cause complications because of subsidence into the adjacent vertebrae with collapse of the intervertebral space. Complications such as cage dislocation or nonunion with instability also have been reported. To prevent such complications, the new WING cervical interbody fusion cage (Medinorm AG, Quierschied, Germany) has been developed. Its area of contact with the adjacent vertebrae is supposed to be large enough to resist excessive subsidence and small enough to prevent stress protection of the tissue growing in the cage.
METHODS: In this study, 24 human cervical spine specimens were tested after stabilization with either a WING, BAK/C, AcroMed I/F cage or bone cement. Then, in a new testing protocol, 700 pure-moment loading cycles (+/-2 Nm) were applied in randomized directions (lateral bending, flexion-extension, and axial rotation alone or in combination with each other) to simulate the patient's neck movements during the first few postoperative days. Measurements of the subsidence depth (total height loss) in combination with flexibility tests (+/-2.5 Nm) were performed before cyclic loading and after 50, 100, 200, 300, 500, and 700 loading cycles.
RESULTS: Cyclic loading caused subsidence in all four device groups, most distinct with BAK/C-cages (1.63 mm after 700 loading cycles) followed by the new WING (0.90 mm) and the AcroMed (0.82 mm) cages. No statistically significant difference could be found among the three cage designs. However, all three cage types showed a significantly higher subsidence depth than bone cement (0.48 mm;P = 0.023 between each of the three cage-types and bone cement). A moderate correlation between bone mineral density and subsidence depth could be found only in the BAK/C group (r2 = 0.495). A large subsidence depth after 700 loading cycles was associated with a large flexibility increase in the WING (r2 = 0.786) and AcroMed groups (r2 = 0.21), but with a small flexibility increase in the BAK/C group (r2 = 0.58).
CONCLUSIONS: Postoperative neck movements caused subsidence in all cervical interbody implant types. The new WING cage and the AcroMed cage seemed to have a better resistance against subsidence than the BAK/C cage. However, all three cage types had a significantly higher subsidence tendency than bone cement.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11064521     DOI: 10.1097/00007632-200011010-00008

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  25 in total

1.  Variation of endplate thickness in the cervical spine.

Authors:  T Pitzen; B Schmitz; T Georg; D Barbier; T Beuter; W I Steudel; W Reith
Journal:  Eur Spine J       Date:  2004-01-17       Impact factor: 3.134

Review 2.  A systematic review of the use of expandable cages in the cervical spine.

Authors:  Benjamin D Elder; Sheng-Fu Lo; Thomas A Kosztowski; C Rory Goodwin; Ioan A Lina; John E Locke; Timothy F Witham
Journal:  Neurosurg Rev       Date:  2015-07-28       Impact factor: 3.042

3.  [HAC-titanium as an implant for interbody fusion in spinal canal stenosis of the cervical spine. Six-year clinical trial].

Authors:  D Daentzer; S Asamoto; D-K Böker
Journal:  Orthopade       Date:  2005-03       Impact factor: 1.087

4.  Biomechanical study of a hat type cervical intervertebral fusion cage.

Authors:  Yu-Tong Gu; Lian-Shun Jia; Tong-Yi Chen
Journal:  Int Orthop       Date:  2006-06-09       Impact factor: 3.075

5.  Factors affecting sagittal malalignment due to cage subsidence in standalone cage assisted anterior cervical fusion.

Authors:  Pavel Barsa; Petr Suchomel
Journal:  Eur Spine J       Date:  2007-01-13       Impact factor: 3.134

6.  Lumbar interbody fusion: a parametric investigation of a novel cage design with and without posterior instrumentation.

Authors:  Fabio Galbusera; Hendrik Schmidt; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2011-09-15       Impact factor: 3.134

7.  Subsidence of Cylindrical Cage (AMSLUtrade mark Cage) : Postoperative 1 Year Follow-up of the Cervical Anterior Interbody Fusion.

Authors:  Young Il Joung; Seong Hoon Oh; Yong Ko; Hyeong Joong Yi; Seung Ku Lee
Journal:  J Korean Neurosurg Soc       Date:  2007-11-20

8.  Radiologic Assessment of Subsidence in Stand-Alone Cervical Polyetheretherketone (PEEK) Cage.

Authors:  Sung-Kon Ha; Jung-Yul Park; Se-Hoon Kim; Dong-Jun Lim; Sang-Dae Kim; Sang-Kook Lee
Journal:  J Korean Neurosurg Soc       Date:  2008-12-31

9.  Subsidence of stand-alone cervical cages in anterior interbody fusion: warning.

Authors:  Erol Gercek; Vincent Arlet; Josee Delisle; Dante Marchesi
Journal:  Eur Spine J       Date:  2003-06-21       Impact factor: 3.134

10.  Anterior cervical fusion with interbody cage containing beta-tricalcium phosphate augmented with plate fixation: a prospective randomized study with 2-year follow-up.

Authors:  Li-Yang Dai; Lei-Sheng Jiang
Journal:  Eur Spine J       Date:  2008-02-27       Impact factor: 3.134

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.