Literature DB >> 11061457

Serum titanium level for diagnosis of a failed, metal-backed patellar component.

S S Leopold1, R A Berger, L Patterson, A K Skipor, R M Urban, J J Jacobs.   

Abstract

A case is presented in which an elevated serum titanium level was used to make the diagnosis of a failed metal-backed patellar component. The preoperative serum titanium level was 536.8 ppb, which was 98 times higher than the patient's previous level (taken 1 year earlier, when he was asymptomatic) and 2 orders of magnitude higher than the expected level with a well-functioning implant of this type. Revision surgery confirmed that the polyethylene portion of the patellar component had worn through, leaving the titanium portion of the patellar implant to articulate with the femoral component. Wear-through was not evident on preoperative radiographs or clinical examination. As knowledge about the expected ranges for serum metal ion levels after total joint arthroplasty continues to increase, the diagnostic utility of serum metal ion testing in the evaluation of joint arthroplasty function will continue to improve.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11061457     DOI: 10.1054/arth.2000.6632

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience.

Authors:  Patrick Boyer; Jean-Yves Lazennec; Joel Poupon; Marc-Antoine Rousseau; Philippe Ravaud; Yves Catonné
Journal:  Int Orthop       Date:  2008-10-28       Impact factor: 3.075

2.  Metallic ion release after knee prosthesis implantation: a prospective study.

Authors:  Adrien Lons; Sophie Putman; Gilles Pasquier; Henri Migaud; Elodie Drumez; Julien Girard
Journal:  Int Orthop       Date:  2017-06-14       Impact factor: 3.075

3.  The role of citrate, lactate and transferrin in determining titanium release from surgical devices into human serum.

Authors:  Justin P Curtin; Minji Wang; Tianfan Cheng; Lijian Jin; Hongzhe Sun
Journal:  J Biol Inorg Chem       Date:  2018-04-05       Impact factor: 3.358

4.  Isolated all-polyethylene patellar revisions for metal-backed patellar failure.

Authors:  Ryan M Garcia; Matthew J Kraay; Victor M Goldberg
Journal:  Clin Orthop Relat Res       Date:  2008-08-08       Impact factor: 4.176

Review 5.  Corrosion degradation and prevention by surface modification of biometallic materials.

Authors:  Raghuvir Singh; Narendra B Dahotre
Journal:  J Mater Sci Mater Med       Date:  2006-12-02       Impact factor: 4.727

6.  A Comparison of Biocompatibility of a Titanium Alloy Fabricated by Electron Beam Melting and Selective Laser Melting.

Authors:  Hong Wang; Bingjing Zhao; Changkui Liu; Chao Wang; Xinying Tan; Min Hu
Journal:  PLoS One       Date:  2016-07-08       Impact factor: 3.240

7.  In vitro testing of femoral impaction grafting with porous titanium particles: a pilot study.

Authors:  René Aquarius; Luc Walschot; Pieter Buma; Berend Willem Schreurs; Nico Verdonschot
Journal:  Clin Orthop Relat Res       Date:  2009-01-13       Impact factor: 4.176

  7 in total