Agreement between two plasma bicarbonate assays in critically ill patients.

D A Story1, S Poustie.   

Abstract

Previous studies have suggested that measurement of plasma bicarbonate concentration using the Henderson-Hasselbalch equation may be unreliable, particularly in critically ill patients. We examined the agreement between two plasma bicarbonate concentration assays in critically ill patients. Data were collected from records of routine daily blood samples. Paired samples were taken at the same time from arterial lines. A Bland-Altman analysis was used to compare two bicarbonate assays in clinical use. The first used the Henderson-Hasselbalch equation for blood-gas machine calculations. The second used a spectrophotometric enzymatic technique. Comparing the enzymatic method to the calculated method (enzymatic minus calculated) the bias was -1.6 mmol/l (95% CI: -1.2 to -2.0 mmol/l). The limits of agreement were -5.85 mmol/l to 2.65 mmol/l. This study found poor agreement between the two bicarbonate assays. This poor agreement is clinically important but the causes are unclear. We suggest further investigation of the reliability of bicarbonate assays.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10969366     DOI: 10.1177/0310057X0002800407

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  4 in total

1.  Comparison of values of traditionally measured venous bicarbonate with calculated arterial bicarbonate in intensive care unit patients of a hospital in a third-world country.

Authors:  Waheeda Nargis; Akm Shafiqur Rahman; Borhan Uddin Ahamed; Md Zakir Hossain
Journal:  Niger Med J       Date:  2014-07

2.  Effectiveness of the Stewart Method in the Evaluation of Blood Gas Parameters.

Authors:  Mustafa Gezer; Fatih Bulucu; Kadir Ozturk; Selim Kilic; Umit Kaldirim; Yusuf Emrah Eyi
Journal:  Turk J Emerg Med       Date:  2016-03-02

3.  Pseudo-anion gap acidosis.

Authors:  Sankar D Navaneethan; Robert Mooney; James Sloand
Journal:  NDT Plus       Date:  2008-01-04

4.  Comparison of bicarbonate values from venous blood gas and chemistry panels measured at the time of diagnosis and resolution of diabetes ketoacidosis.

Authors:  Poorani Nallam Goundan; Devina Luhur Willard; Ayse Sahin-Efe; Shu-Ling Liang Fan; Sara Michelle Alexanian
Journal:  J Clin Transl Endocrinol       Date:  2019-08-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.