Waheeda Nargis1, Akm Shafiqur Rahman2, Borhan Uddin Ahamed3, Md Zakir Hossain2. 1. Department of Biochemistry, Uttara Adhunik Medical College Hospital, Uttara, Dhaka, Bangladesh. 2. Department of Anaesthesiology and Intensive Care Unit, Uttara Adhunik Medical College Hospital, Uttara, Dhaka, Bangladesh. 3. Department of Forensic Medicine, Dhaka Community Medical College and Hospital, Moghbazar, Dhaka, Bangladesh.
Abstract
BACKGROUND: Measurement of serum or plasma bicarbonate (HCO3 (-)) from a sample of peripheral venous blood is routinely practiced in hospital patient managements. HCO3 (-) status can also be obtained by a simple calculation during blood gas analysis requiring arterial blood as sample which is cumbersome for both patient and doctor. This study compared the measured bicarbonate levels with calculated arterial values in intensive care unit (ICU) patients to determine whether traditionally measured venous HCO3 (-) and calculated HCO3 (-) values [from arterial blood gas (ABG) analyzers] can be used interchangeably. MATERIALS AND METHODS: This prospective study was carried out at a tertiary care teaching hospital in Dhaka, the capital of Bangladesh. A total of 56 adult patients with diverse medical conditions, presenting at the ICU of the health centre were enrolled in this study when deemed by the treating physician to have an ABG analysis. Arterial and venous samples were taken as close in time as possible for gas analysis and routine blood tests. RESULTS: The HCO3 (-) levels from ABG and traditionally measured serum showed acceptably narrow 95% limits of agreement using the Bland-Altman method. CONCLUSIONS: More widely prescribed venous HCO3 (-) measurements can also be a useful substitute for an expensive ABG analyzer in resource-constrained health care sectors when required. However, accuracy of venous blood in assessment of additional ABG parameters is yet to be discovered.
BACKGROUND: Measurement of serum or plasma bicarbonate (HCO3 (-)) from a sample of peripheral venous blood is routinely practiced in hospital patient managements. HCO3 (-) status can also be obtained by a simple calculation during blood gas analysis requiring arterial blood as sample which is cumbersome for both patient and doctor. This study compared the measured bicarbonate levels with calculated arterial values in intensive care unit (ICU) patients to determine whether traditionally measured venous HCO3 (-) and calculated HCO3 (-) values [from arterial blood gas (ABG) analyzers] can be used interchangeably. MATERIALS AND METHODS: This prospective study was carried out at a tertiary care teaching hospital in Dhaka, the capital of Bangladesh. A total of 56 adult patients with diverse medical conditions, presenting at the ICU of the health centre were enrolled in this study when deemed by the treating physician to have an ABG analysis. Arterial and venous samples were taken as close in time as possible for gas analysis and routine blood tests. RESULTS: The HCO3 (-) levels from ABG and traditionally measured serum showed acceptably narrow 95% limits of agreement using the Bland-Altman method. CONCLUSIONS: More widely prescribed venous HCO3 (-) measurements can also be a useful substitute for an expensive ABG analyzer in resource-constrained health care sectors when required. However, accuracy of venous blood in assessment of additional ABG parameters is yet to be discovered.
Entities:
Keywords:
Blood gas analyzer; calculated bicarbonate; measured bicarbonate
The measurement of bicarbonate level in blood is extremely common and often provides vitally important data used in the care of critically illpatients. The bicarbonate level in blood can be directly measured or derived from calculations using the Henderson-Hasselbalch equation; mostly adopted by the blood gas analyzers.1 Arterial blood gas (ABG) analysis is commonly performed for clinical evaluation, but the procedure has certain limitations in the form of reduced patient acceptability (because the procedure can be painful) and the potential to cause complications such as arterial injury, thrombosis with distal ischemia, hemorrhage, aneurysm formation, median nerve damage and, rarely, reflex sympathetic dystrophy.2,3 Thus, venous blood analysis is a relatively safer procedure requiring fewer punctures and hence, reducing the risk of needle stick injury to the health care worker.Although various studies suggest that a venous sample is relatively accurate for acid–base assessment,3456 VBG analysis has not gained much acceptance as a substitute for ABG analysis. On the contrary, Bicarbonate ions make up ~ 95% of the total carbon dioxide of the plasma, and hence both of them have been used interchangeably. Most blood gas analyzers use the Henderson-Hasselbach equation to calculate bicarbonate values based on the assumption that the dissociation constant (pK) and solubility coefficient (α) are invariant. However, pK is affected by changes in pH, ionic strength and temperature, while the values of α varies with the composition of the solution such as the presence of increased salts, proteins or lipids.1 Therefore, the calculated bicarbonate values; irrespective of the type of sample, may have significant error under certain circumstances, making its reliability questionable. Previous studies using different statistical methods to assess the agreement between measured and calculated bicarbonate have shown conflicting results, with some studies showing good agreement,78910 while some studies showed otherwise.111213So far this study is concerned; the feasibility of such expensive procedure in a middle income country like ours is intended to be reconsidered. However, nearly all medical centers in Bangladesh have the traditional laboratory methods to determine the concentrations of this key analyte in routine venous blood. But arterial blood gas analyzers are still limited to a small number of referral hospitals. This study was aimed to investigate the extent of agreement between arterial and venous bicarbonate levels for a group of intensive care unit (ICU) patients to determine whether measured serum HCO3− from traditional peripheral venous blood and calculated HCO3− values (from ABG analyzers) can be used interchangeably.
MATERIALS AND METHODS
This prospective study was conducted in the intensive care unit of a tertiary care teaching hospital of central Bangladesh, between July and December 2011. A total of 56 patients deemed by their treating doctor to require blood gas analysis either during initial evaluations or as part of their clinical care in ICU were included in this study. After having the study explained, a verbal consent was obtained from the patient or the relative. The study obtained 63 matched sample pairs from these 56 patients; 34 males, 27 females, mean age 53.7 ranging from 29 to 70 years. The presenting diagnosis of the patients was: acute respiratory distress (30.1%), acute coronary syndrome (26.9%), acute or chronic renal failure (20.6%), post-operative sepsis (17.4%) and other suspected metabolic derangements (4.7%). Samples for arterial blood gas analysis were obtained either by a registered nurse or a doctor. The peripheral venous samples collected at the time of intravenous line insertion or as per recommendation used for other relevant blood tests were used. Multiple samples from the same patient at different dates were also accepted.Calculated HCO3− concentrations were obtained from arterial blood gas samples using a Roche OPTI CCA op39090 instrument, which calculates the HCO3− concentration from the arterial pH and CO2 partial pressure. The HCO3− concentration was directly measured in the venous serum or plasma by means of ion-selective electrodes (ISE). The samples were analyzed as quickly as possible using the blood gas analyzer; routinely practiced in the ICU. Data were analyzed using SPSS version 12 and MedCal for Windows. Pearson correlation and Bias plot (Bland and Altman)14 method was used to compare the three measurements.
RESULTS
The HCO3− levels measured in venous serum showed higher values in contrast to the calculated ones (ranges 21.6 - 37.1 vs 19.8-35.3). The statistical analysis with chi square test resulted in insignificant differences (Chi-square = 0.381, p= 0.826) among the sample group frequencies based on their differences in values [Table 1].
Table 1
Distribution of differences between measured (venous serum or plasma) and calculated (arterial blood gas) HCO3– values
Distribution of differences between measured (venous serum or plasma) and calculated (arterial blood gas) HCO3– valuesThe values shared highly significant correlation (P < 0.001; r = 0.987) when compared with Pearson correlation test and Bland and Altman method [Table 2]. The bias plot [Figure 1] demonstrates acceptable but poor degree of agreement for the values of bicarbonate (mmol/L) measured in venous serum vs calculated from Arterial Blood Gas analysis.
Table 2
Comparison between the measured (venous serum) and calculated (arterial blood gas) HCO3– values
Figure 1
Bias plot comparing measured venous serum vs calculatedarterial blood gas HCO3- (Cal ABG: calculated arterial blood gas HCO3 −)
Comparison between the measured (venous serum) and calculated (arterial blood gas) HCO3– valuesBias plot comparing measured venous serum vs calculatedarterial blood gas HCO3- (Cal ABG: calculated arterial blood gasHCO3 −)
DISCUSSION
An important part of the assessment of the clinical status and progress of critically illpatients is timely assessment of acid–base status which is invariably practiced in arterial blood when it comes to blood gas analyzer. However, it may not always be practical to obtain arterial samples, particularly in the early stages of resuscitation. Besides, such sophisticated device demands careful handling and adequate maintenance raising the cast of analysis. An ample evidence is available on agreement in ABG and VBG values but it does not reduce the cast load over a patient.3451516 So, this study was intended to focus on the unseen burdens over patient as well as the health care institutes and to look for a relatively easy, price-worthy and more specific measures.The agreement or discrepancy between measured and calculated bicarbonate and whether both can be used interchangeably has long been discussed since the 80's,910171819 without any concrete conclusion. With the advancement in the methodology, this issue seemed to have resurfaced, with a couple of articles published since 2008.6782021The correlation coefficient between the measured and calculated bicarbonate revealed a good correlation (r = 0.987, P < 0.001). However, the use of correlation coefficient alone to assess the agreement between two methods may not be appropriate, as correlation depends on the range of values in the sample; a wide range of values like ours will yield a high correlation coefficient. Therefore, values which seem to be in poor agreement can produce high correlations provided the range is wide enough. Now, using Bland-Altman analysis, this study exhibited excellent agreement in the bicarbonate values. According to the Bland-Altman analysis significant degree of agreement was found between measured serum versus ABGbicarbonate estimates,78910 respectively, unlike a few.111213 However, such comparisons of measured and calculated HCO3− values can be used to detect errors in the collection or analysis of blood samples.
CONCLUSIONS
According to the study results traditionally measured venous bicarbonate can be a convenient substitute for calculated arterial bicarbonate in critically ill ICUpatients. However, more accurate assessments will require ABG for additional parameters. Besides, the present study design did not involve collection of data on patient demographics, severity of illness, and requirement for inotropic support or prognosis. Detailed data, allowing analyses that include these potentially important variables might be focused in future large scale studies to add more to it.