Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT.

G El Fakhri1, I Buvat, H Benali, A Todd-Pokropek, R Di Paola.   

Abstract

UNLABELLED: We determined the relative effect of corrections for scatter, depth-dependent collimator response, attenuation, and finite spatial resolution on various image characteristics in cardiac SPECT.
METHODS: Monte Carlo simulations and real acquisition of a 99mTc cardiac phantom were performed under comparable conditions. Simulated and acquired data were reconstructed using several correction schemes that combined different methods for scatter correction (3 methods), depth-dependent collimator response correction (frequency-distance principle), attenuation correction (nonuniform Chang correction or within an iterative reconstruction algorithm), and finite spatial resolution correction (use of recovery coefficients). Five criteia were considered to assess the effect of the processing schemes: bull's-eye map (BEM) uniformity, contrast between the left ventricle (LV) wall and the LV cavity, spatial resolution, signal-to-noise ratio (SNR), and percent errors with respect to the known LV wall and liver activities.
RESULTS: Similar results were obtained for the simulated and acquired data. Scatter correction significantly improved contrast and absolute quantitation but did not have noticeable effects on BEM uniformity or on spatial resolution and reduced the SNR. Correction for the depth-dependent collimator response improved spatial resolution from 13.3 to 9.5 mm in the LV region, improved absolute quantitation and contrast, but reduced the SNR. Correcting for attenuation was essential for restoring BEM uniformity (78% and 89% without and with attenuation correction, respectively [ideal value being 100%]) and accurate absolute activity quantitation (errors in estimated LV wall and liver activity decreased from 90% without attenuation correction to approximately20% with attenuation correction only). Although accurate absolute activity quantitation was achieved in the liver using scatter and attenuation corrections only, correction for finite spatial resolution was needed to estimate LV wall activity within 10%.
CONCLUSION: The respective effects of corrections for scatter, depth-dependent collimator response, attenuation, and finite spatial resolution on different image features in cardiac SPECT were quantified for a specific acquisition configuration. These results give indications regarding the improvements to be expected when using a specific processing scheme involving some or all corrections.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10945534

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  16 in total

1.  Assessment of geometrical distortion and activity distribution after attenuation correction: A SPECT phantom study.

Authors:  Sara Lappi; Simonetta Lazzari; Graziella Sarti; Pierluigi Pieri
Journal:  J Nucl Cardiol       Date:  2002 Sep-Oct       Impact factor: 5.952

Review 2.  Scatter modelling and compensation in emission tomography.

Authors:  Habib Zaidi; Kenneth F Koral
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-03-31       Impact factor: 9.236

3.  Cross-camera comparison of SPECT measurements of a 3-D anthropomorphic basal ganglia phantom.

Authors:  Walter Koch; Perry E Radau; Wolfgang Münzing; Klaus Tatsch
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-01-25       Impact factor: 9.236

Review 4.  Advances in technical aspects of myocardial perfusion SPECT imaging.

Authors:  Piotr J Slomka; James A Patton; Daniel S Berman; Guido Germano
Journal:  J Nucl Cardiol       Date:  2009-02-26       Impact factor: 5.952

5.  Assessment of the sources of error affecting the quantitative accuracy of SPECT imaging in small animals.

Authors:  Andrew B Hwang; Benjamin L Franc; Grant T Gullberg; Bruce H Hasegawa
Journal:  Phys Med Biol       Date:  2008-04-09       Impact factor: 3.609

6.  Comparison of attenuation, dual-energy-window, and model-based scatter correction of low-count SPECT to 82Rb PET/CT quantified myocardial perfusion scores.

Authors:  R Glenn Wells; Karen Soueidan; Rachel Timmins; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2013-06-05       Impact factor: 5.952

7.  Quantitative SPECT/CT: SPECT joins PET as a quantitative imaging modality.

Authors:  Dale L Bailey; Kathy P Willowson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-09-14       Impact factor: 9.236

Review 8.  Brain single-photon emission CT physics principles.

Authors:  R Accorsi
Journal:  AJNR Am J Neuroradiol       Date:  2008-06-26       Impact factor: 3.825

9.  Correction for respiration artefacts in myocardial perfusion SPECT is more effective when reconstructions supporting collimator detector response compensation are applied.

Authors:  Gil Kovalski; Zohar Keidar; Alex Frenkel; Ora Israel; Haim Azhari
Journal:  J Nucl Cardiol       Date:  2009 Nov-Dec       Impact factor: 5.952

10.  Repeatability of Radiotracer Uptake in Normal Abdominal Organs with ¹¹¹In-Pentetreotide Quantitative SPECT/CT.

Authors:  Steven P Rowe; Esther Vicente; Nadège Anizan; Hao Wang; Jeffrey P Leal; Martin A Lodge; Eric C Frey; Richard L Wahl
Journal:  J Nucl Med       Date:  2015-05-14       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.